ASTM D7278-2006 Standard Guide for Prediction of Analyzer Sample System Lag Times《分析仪抽样系统迟延时间预测的标准指南》.pdf
《ASTM D7278-2006 Standard Guide for Prediction of Analyzer Sample System Lag Times《分析仪抽样系统迟延时间预测的标准指南》.pdf》由会员分享,可在线阅读,更多相关《ASTM D7278-2006 Standard Guide for Prediction of Analyzer Sample System Lag Times《分析仪抽样系统迟延时间预测的标准指南》.pdf(9页珍藏版)》请在麦多课文档分享上搜索。
1、Designation: D 7278 06An American National StandardStandard Guide forPrediction of Analyzer Sample System Lag Times1This standard is issued under the fixed designation D 7278; the number immediately following the designation indicates the year oforiginal adoption or, in the case of revision, the yea
2、r of last revision. A number in parentheses indicates the year of last reapproval. Asuperscript epsilon (e) indicates an editorial change since the last revision or reapproval.INTRODUCTIONLag time, as used in this guide, is the time required to transport a representative sample from theprocess tap t
3、o the analyzer. Sample system designs have infinite configurations so this guide gives theuser guidance, based on basic design considerations, when calculating the lag time of on-line sampledelivery systems. Lag time of the analyzer sample system is a required system characteristic whenperforming sy
4、stem validation in Practice D 3764 or D 6122 and in general the proper operation of anyon-line analytical system. The guide lists the components of the system that need to be consideredwhen determining lag time plus a means to judge the type of flow and need for multiple flushes beforeanalysis on an
5、y sample.1. Scope1.1 This guide covers the application of routine calculationsto estimate sample system lag time, in seconds, for gas, liquid,and mixed phase systems.1.2 This guide considers the sources of lag time from theprocess sample tap, tap conditioning, sample transport, pre-analysis conditio
6、ning and analysis.1.3 Lag times are estimated based on a prediction of flowcharacteristics, turbulent, nonturbulent, or laminar, and thecorresponding purge requirements.1.4 Mixed phase systems prevent reliable representativesampling so system lag times should not be used to predictsample representat
7、ion of the stream.1.5 This standard does not purport to address all of thesafety concerns, if any, associated with its use. It is theresponsibility of the user of this standard to establish appro-priate safety and health practices and determine the applica-bility of regulatory limitations prior to u
8、se.2. Referenced Documents2.1 ASTM Standards:2D 3764 Practice for Validation of Process Stream AnalyzerSystemsD 6122 Practice for Validation of Multivariate Process In-frared Spectrophotometers3. Terminology3.1 Definitions:3.1.1 continuous analyzer unit cycle timethe time intervalrequired to replace
9、 the volume of the analyzer measurementcell.3.1.2 intermittent analyzer unit cycle timethe time inter-val between successive updates of the analyzer output.3.1.3 purge volumethe combined volume of the fullanalyzer sampling and conditioning systems.3.1.4 sample system lag timethe time required to tra
10、nsporta representative sample from the process tap to the analyzer.3.1.5 system response timethe sum of the analyzer unitresponse time and the analyzer sample system lag time.3.2 Abbreviations:3.2.1 I.D.Internal Diameter3.2.2 ReReynolds Number4. Summary4.1 The lag time of an analyzer sample system i
11、s estimatedby first determining the flow characteristics. The flow isassigned as turbulent or non-turbulent to assign the number ofpurges required to change out the sample. Based on thehardware employed in the sample system an estimation of thelag time can be calculated.5. Significance and Use5.1 Th
12、e analyzer sample system lag time estimated by thisguide can be used in conjunction with the analyzer output toaid in optimizing control of blender facilities or process units.1This guide is under the jurisdiction of ASTM Committee D02 on PetroleumProducts and Lubricants and is the direct responsibi
13、lity of Subcommittee D02.25 onPerformance Assessment and Validation of Process Stream Analyzer Systems forPetroleum and Petroleum Products.Current edition approved July 1, 2006. Published August 2006.2For referenced ASTM standards, visit the ASTM website, www.astm.org, orcontact ASTM Customer Servic
14、e at serviceastm.org. For Annual Book of ASTMStandards volume information, refer to the standards Document Summary page onthe ASTM website.1Copyright ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959, United States.5.2 The lag time can be used in the tuning of
15、controlprograms to set the proper optimization frequency.5.3 The application of this guide is not for the design of asample system but to estimate the performance of existingsample systems. The principles listed in this guide are there tohelp understand design concepts and allow the application ofth
16、is guide for its intended scope. Additional detailed informa-tion can be found in the references provided in the sectionentitled Additional Reading Material.6. Basic Design Considerations6.1 Acceptable Lag TimeAs a general rule, a one to twominute sample system lag time should be maintained wherepos
17、sible to give acceptable performance. Flow is a keycomponent in the determination of sample system lag time. Inmost systems the desired system lag time is impossible toachieve with maximum allowable sample flow rate to theanalyzer. To improve lag time a fast loop or bypass can be usedto increase sam
18、ple velocities through the system to a point justupstream of the analyzer.Aslipstream is taken from the bypassto feed the analyzer at its optimum flowrate. Excess sample inthe slipstream is vented to atmosphere, to flare or to the processstream dependant upon application and regulatory require-ments
19、.6.2 Physical State of Sample:6.2.1 Liquid SamplesPressure drop properties often gov-ern the design of a liquid system. This is due for the most parton the close relationship between pressure drop and systemflowrate and the fixed pressure differential available from theprocess for sample transport.
20、The sizing of the sample compo-nents is a tradeoff between pressure drop and sample flowrate.High sample flowrates in small sized component systems causehigh-pressure drops and low sample transport times. The sameflowrate in a larger tubing system will yield significant im-provements in pressure dro
21、p through the system, but will alsoincrease the time for sample transport significantly.6.2.2 Vapor SamplesVapor phase sampling is governedless by pressure drop and more by pressure compressionproperties of gases when compared to liquids. In compressiblegases the higher the pressure in a given volum
22、e, the moresample present in that volume. For this reason, and differentfrom liquids, the selection and location of pressure regulatingdevices in the vapor sample system has a great impact on theoverall system design. The optimal location for a high-pressureregulator in a vapor sample is immediately
23、 downstream of thesample tap or high-pressure location thereby limiting thevolume of the system under high pressure. Since the density ofa compressible fluid is a function of the pressure, compressiblefluid flow rate calculations are sometimes done over segmentallengths where average properties adeq
24、uately represent the fluidconditions of the line segment.6.2.3 Liquid to Vapor SamplesA change of phase due tosample vaporization can impact the sample lag time. Thevolume change from the liquid phase to the vapor phase issubstantial. Typical flow rates in gaseous sample lines down-stream of the vap
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
5000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ASTMD72782006STANDARDGUIDEFORPREDICTIONOFANALYZERSAMPLESYSTEMLAGTIMES 分析 抽样 系统 迟延 时间 预测 标准 指南 PDF

链接地址:http://www.mydoc123.com/p-525026.html