《ASTM D5773-2007 374 Standard Test Method for Cloud Point of Petroleum Products (Constant Cooling Rate Method)《测定石油产品始凝点的标准试验方法》.pdf》由会员分享,可在线阅读,更多相关《ASTM D5773-2007 374 Standard Test Method for Cloud Point of Petroleum Products (Constant Cooling Rate Method)《测定石油产品始凝点的标准试验方法》.pdf(7页珍藏版)》请在麦多课文档分享上搜索。
1、Designation: D 5773 07Designation: 446/99An American National StandardStandard Test Method forCloud Point of Petroleum Products (Constant Cooling RateMethod)1This standard is issued under the fixed designation D 5773; the number immediately following the designation indicates the year oforiginal ado
2、ption or, in the case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. Asuperscript epsilon () indicates an editorial change since the last revision or reapproval.INTRODUCTIONThis test method describes an alternative procedure for the determinati
3、on of cloud point ofpetroleum products Test Method D 2500/IP 219 using an automatic apparatus. The temperature resultsfrom this test method have been found to be equivalent to Test Method D 2500/IP 219. Whenspecification requires Test Method D 2500/IP 219, do not substitute this test method or any o
4、thermethod without obtaining comparative data and agreement from the specifier.1. Scope*1.1 This test method covers the determination of the cloudpoint of petroleum products and biodiesel fuels that aretransparent in layers 40 mm in thickness by an automaticinstrument using a constant cooling rate.1
5、.2 This test method covers the range of temperatures from60 to +49C with temperature resolution of 0.1C, however,the range of temperatures included in the 1997 interlaboratorycooperative test program only covered the temperature range of56 to +34C.1.3 The values stated in SI units are to be regarded
6、 as thestandard. The values given in parentheses are for informationonly.1.4 This standard does not purport to address all of thesafety concerns, if any, associated with its use. It is theresponsibility of the user of this standard to establish appro-priate safety and health practices and determine
7、the applica-bility of regulatory limitations prior to use.2. Referenced Documents2.1 ASTM Standards:2D 2500 Test Method for Cloud Point of Petroleum ProductsD 4057 Practice for Manual Sampling of Petroleum andPetroleum ProductsD 4177 Practice for Automatic Sampling of Petroleum andPetroleum Products
8、2.2 Energy Institute Standard:3IP 219 Test Method for Cloud Point of Petroleum Products3. Terminology3.1 Definitions:3.1.1 biodiesel, na fuel comprised of mono-alkyl esters oflong-chain fatty acids derived from vegetable oils or animalfats, designated B100.3.1.1.1 DiscussionBiodiesel is typically pr
9、oduced by areaction of vegetable oil or animal fat with an alcohol such asmethanol or ethanol in the presence of a catalyst to yieldmono-esters and glycerin. The fuel typically may contain up to14 different types of fatty acids that are chemically transformedinto fatty acid methyl esters (FAME).3.1.
10、2 biodiesel blend, na blend of biodiesel fuel withpetroleum-based diesel fuel designated BXX, where XX is thevolume percentage of biodiesel.3.1.3 cloud point, nin petroleum products and biodieselfuels, the temperature of a liquid specimen when the smallestobservable cluster of hydrocarbon crystals f
11、irst occurs uponcooling under prescribed conditions.3.1.3.1 DiscussionThe cloud point occurs when the tem-perature of the specimen is low enough to cause hydrocarboncrystals to precipitate. In a homogeneous liquid, the cloud isalways noted first at the location in the specimen where thespecimen temp
12、erature is the lowest. The cloud point is thetemperature at which the crystals first occur, regardless of theirlocation in the specimen, and not after extensive crystallization1This test method is under the jurisdiction of ASTM Committee D02 onPetroleum Products and Lubricants and is the direct resp
13、onsibility of SubcommitteeD02.07 on Flow Properties.Current edition approved Dec. 1, 2007. Published January 2008. Originallyapproved in 1995. Last previous edition approved in 2005 as D 577305.2For referenced ASTM standards, visit the ASTM website, www.astm.org, orcontact ASTM Customer Service at s
14、erviceastm.org. For Annual Book of ASTMStandards volume information, refer to the standards Document Summary page onthe ASTM website.3Available from the Energy Institute, 61 New Cavendish Street, London,England WIM 8AR.1*A Summary of Changes section appears at the end of this standard.Copyright ASTM
15、 International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959, United States.has taken place. The hydrocarbon crystals that precipitate atlower temperatures are typically, but not excluded to, straightchain hydrocarbons commonly called “wax crystals.”3.1.3.2 DiscussionThe purp
16、ose of the cloud point is tomeasure the wax crystals in the specimen; however, traceamounts of water and inorganic compounds may also bepresent.The intent of the cloud point measurement is to capturethe temperature at which the liquid fuel in the specimen beginsto change from a single liquid phase t
17、o a two-phase systemcontaining solid and liquid. It is not the intent of this testmethod to monitor the phase transition of the trace componentssuch as water.3.2 Definitions of Terms Specific to This Standard:3.2.1 automatic cloud point, nthe temperature of a speci-men, when the appearance of the cl
18、oud is determined under theconditions of this test method.3.2.1.1 DiscussionThe cloud point in this test method isdetermined by an automatic instrument using an optical devicefor detection of the crystal formation. The apparatus and theconditions are different from those established for Test MethodD
19、 2500, although according to interlaboratory examination, theresults have been determined to be equivalent to Test MethodD 2500.3.2.2 D 2500/IP 219 equivalent cloud point, nthe tem-perature of a specimen, in integers, calculated by rounding theresults of this test method to the next lower integer.3.
20、2.2.1 DiscussionThis test method produces results with0.1C resolution. Should the user wish to provide results witha similar format to Test Method D 2500, then this calculationcan be performed. Some apparatus can perform this calculationautomatically.3.2.3 Peltier device, na solid state thermoelectr
21、ic deviceconstructed with dissimilar semiconductor materials and con-figured in such a way that it will transfer heat to or away froma test specimen dependent on the direction of electric currentapplied to the device.4. Summary of Test Method4.1 A specimen is cooled by a Peltier device at a constant
22、rate of 1.5 6 0.1C/min while continuously being illuminatedby a light source. The specimen is continuously monitored byan array of optical detectors for the first appearance of a cloudof wax crystals. The detectors are sufficient in number toensure that any solid phase hydrocarbon crystals that may
23、formare detected. The temperature at which the appearance of acloud of wax crystals is first detected in the specimen isrecorded to 0.1C resolution. When the recorded temperature isrounded to the next lower integer temperature, it is designatedas the D 2500/IP 219 equivalent cloud point per Test Met
24、hodD 5773.5. Significance and Use5.1 The cloud point of petroleum products and biodieselfuels is an index of the lowest temperature of their utility forcertain applications. Wax crystals of sufficient quantity canplug filters used in some fuel systems.5.2 Petroleum blending operations require a prec
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
10000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ASTMD57732007374STANDARDTESTMETHODFORCLOUDPOINTOFPETROLEUMPRODUCTSCONSTANTCOOLINGRATEMETHOD 测定 石油 产品

链接地址:http://www.mydoc123.com/p-520649.html