ASTM D4327-2003 Standard Test Method for Anions in Water by Chemically Suppressed Ion Chromatography《用化学压缩离子色谱法对水中阴离子的标准试验方法》.pdf
《ASTM D4327-2003 Standard Test Method for Anions in Water by Chemically Suppressed Ion Chromatography《用化学压缩离子色谱法对水中阴离子的标准试验方法》.pdf》由会员分享,可在线阅读,更多相关《ASTM D4327-2003 Standard Test Method for Anions in Water by Chemically Suppressed Ion Chromatography《用化学压缩离子色谱法对水中阴离子的标准试验方法》.pdf(9页珍藏版)》请在麦多课文档分享上搜索。
1、Designation: D 4327 03Standard Test Method forAnions in Water by Chemically Suppressed IonChromatography1This standard is issued under the fixed designation D 4327; the number immediately following the designation indicates the year oforiginal adoption or, in the case of revision, the year of last r
2、evision. A number in parentheses indicates the year of last reapproval. Asuperscript epsilon (e) indicates an editorial change since the last revision or reapproval.1. Scope*1.1 This test method2covers the sequential determination offluoride, chloride, nitrite, ortho-phosphate, bromide, nitrate,and
3、sulfate ions in water by chemically suppressed ion chro-matography.NOTE 1Order of elution is dependent upon the column used; see Fig.1.1.2 This test method is applicable to drinking and wastewa-ters. The ranges tested for this test method for each anion wereas follows (measured in mg/L):Fluoride 0.2
4、6 to 8.49Chloride 0.78 to 26.0Nitrite-N 0.36 to 12.0Bromide 0.63 to 21.0Nitrate-N 0.42 to 14.0o-Phosphate 0.69 to 23.1Sulfate 2.85 to 95.01.3 It is the users responsibility to ensure the validity of thistest method for other matrices.1.4 Concentrations as low as 0.01 mg/L were determineddepending up
5、on the anions to be quantitated, in single labora-tory work. Utilizing a 50-L sample volume loop and asensitivity of 3 S/cm full scale, the approximate detectionlimits shown in Table 1 can be achieved. If lower detectionlevels are required, the sensitivity may be improved by using alower scale setti
6、ng (100 L). The analyst must assure optimum instrumentperformance to maintain a stable baseline at more sensitiveconductivity full-scale settings.1.5 The upper limit of this test method is dependent upontotal anion concentration and may be determined experimen-tally as described in Annex A1. These l
7、imits may be extendedby appropriate dilution or by use of a smaller injection volume.1.6 Using alternate separator column and eluents may per-mit additional anions such as formate or citrate to be deter-mined. This is not the subject of this test method.1.7 This standard does not purport to address
8、all of thesafety problems, if any, associated with its use. It is theresponsibility of the user of this standard to establish appro-priate safety and health practices and determine the applica-bility of regulatory limitations prior to use.2. Referenced Documents2.1 ASTM Standards:D 1066 Practice for
9、 Sampling Steam3D 1129 Terminology Relating to Water3D 1193 Specification for Reagent Water3D 2777 Practice for Determination of Precision and Bias ofApplicable Methods of Committee D-19 on Water31This test method is under the jurisdiction of ASTM Committee D19 on Waterand is the direct responsibili
10、ty of Subcommittee D19.05 on Inorganic Constituentsin Water.Current edition approved Jan. 10, 2003. Published January 2003. Originallyapproved in 1984. Last previous edition approved in 1997 as D 4327 97.2The following references may be consulted for additional information:Small, H., Stevens, T. S.,
11、 and Bauman, W. C., “Novel Ion Exchange Chromato-graphic Method Using Conductrimetric Detection,” Analytical Chemistry, Vol 47,1975, p. 1801.Stevens, T. S., Turkelson, V. T., and Alve, W. R., “Determination of Anions inBoiler Blow Down Water with Ion Chromatography,” Analytical Chemistry, Vol 49,197
12、7, p. 1176.Sawicki, E., Mulik, J. D., and Witgenstein, E., Editors, Ion ChromatographicAnalysis of Environmental Pollutants, Ann Arbor Science Publishers, Ann Arbor,MI, 1978.Mulik, J. D., and Sawicki, E., Editors, Ion Chromatographic Analysis ofEnvironmental Pollutants, Vol/No. 2,AnnArbor Science Pu
13、blishers,AnnArbor, MI,1979.Weiss, J., Handbook of Ion Chromatography, Dionex Corp., Sunnyvale, CA,1986.Waters Innovative Methods for Anion Analysis, Waters Chromatography Divisionof Millipore, Method A 107 and A 116, 1990.Haddad, P. R., and Jackson, P. E., Ion Chromatography: Principles andApplicati
14、ons, Elsevier Scientific Publishing Co., 1990.3Annual Book of ASTM Standards, Vol 11.01.FIG. 1 Chromatogram Showing Separation Using the AS4AColumn1*A Summary of Changes section appears at the end of this standard.Copyright ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, P
15、A 19428-2959, United States.D 3370 Practices for Sampling Water from Closed Con-duits3D 5810 Guide for Spiking into Aqueous Samples3D 5847 Practice for the Writing Quality Control Specifica-tions for Standard Test Methods for Water Analysis33. Terminology3.1 DefinitionsFor definitions of terms used
16、in this testmethod, refer to Terminology D 1129.3.2 Definitions of Terms Specific to This Standard:3.2.1 analytical columnsa combination of one or moreguard columns followed by one or more separator columnsused to separate the ions of interest. It should be rememberedthat all of the columns in serie
17、s contribute to the overallcapacity of the analytical column set.3.2.2 chemical suppressor devicea device that is placedbetween the analytical columns and the detector. Its purpose isto inhibit detector response to the ionic constituents in theeluent, so as to lower the detector background and at th
18、e sametime enhance detector response to the ions of interest.3.2.3 eluentthe ionic mobile phase used to transport thesample through the system.3.2.4 guard columna column used before the separatorcolumn to protect it from contaminants, such as particulatematter or irreversibly retained materials.3.2.
19、5 ion chromatographya form of liquid chromatogra-phy in which ionic constituents are separated by ion exchangefollowed by a suitable detection means.3.2.6 resolutionthe ability of an analytical column toseparate constituents under specific test conditions.3.2.7 separator columnthe ion exchange colum
20、n used toseparate the ions of interest according to their retentioncharacteristics prior to their detection.4. Summary of Test Method4.1 An aliquot of sample is injected into an ion chromato-graph. The sample is pumped through two columns and asuppressor device and into a conductivity detector. The
21、ana-lytical column and the guard column are packed with low-capacity anion exchanger. Ions are separated based on theiraffinity for the exchange sites of the resin. The suppressordevice contains a fiber or membrane based cation exchangerthat is continuously regenerated by a flow of dilute sulfuricac
22、id. The suppressor device reduces the background conduc-tivity of the eluent to a low or negligible level by replacing thecations with the hydrogen ion, thereby converting the anions inthe sample to their corresponding acids. The separated anionsin their acid form are measured using an electrical-co
23、nductivitycell. Anions are identified based on their retention timescompared to known standards. Quantitation is accomplished bymeasuring the peak height or area and comparing it to acalibration curve generated from known standards.5. Significance and Use5.1 Ion chromatography provides for both qual
24、itative andquantitative determination of seven common anions, F,Cl,NO2, HPO42,Br,NO3, and SO42, in the milligram perlitre range from a single analytical operation requiring only afew millilitres of sample and taking approximately 10 to 15min for completion.NOTE 2This test method may be used to deter
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
10000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ASTMD43272003STANDARDTESTMETHODFORANIONSINWATERBYCHEMICALLYSUPPRESSEDIONCHROMATOGRAPHY 化学 压缩 离子 色谱 水中

链接地址:http://www.mydoc123.com/p-516830.html