1、Designation: D 4199 82 (Reapproved 2003)Standard Test Methods forAutoclavability of Membrane Filters1This standard is issued under the fixed designation D 4199; the number immediately following the designation indicates the year oforiginal adoption or, in the case of revision, the year of last revis
2、ion. A number in parentheses indicates the year of last reapproval. Asuperscript epsilon (e) indicates an editorial change since the last revision or reapproval.1. Scope1.1 These test methods deal with tests to determine theautoclavability of membrane filters.1.2 These test methods are applicable to
3、 all membranefilters. Either Test Method A, Test Method B, or both, may beutilized to assess the autoclavability of membrane filters.SectionTest Method ADimensional Stability 8.1Test Method BWetting Characteristics 8.21.3 This standard does not purport to address all of thesafety concerns, if any, a
4、ssociated with its use. It is theresponsibility of the user of this standard to establish appro-priate safety and health practices and determine the applica-bility of regulatory limitations prior to use.2. Referenced Documents2.1 ASTM Standards:2D 1129 Terminology Relating to WaterD 1193 Specificati
5、on for Reagent Water3. Terminology3.1 DefinitionsFor definitions of terms used in these testmethods, refer to Terminology D 1129.4. Summary of Test Methods4.1 Test Method A Autoclavability is evaluated by mea-suring the diameter of the membrane filter, across three axes,prior to autoclaving and afte
6、r autoclaving, to determine dimen-sional stability.4.2 Test Method B Wettability is determined by placingsamples of autoclaved and nonautoclaved membrane filters inwater and measuring the time required for the filters to becomecompletely wet.5. Significance and Use5.1 These test methods provide info
7、rmation on the dimen-sional stability and wetting characteristics of autoclaved mem-brane filters. Excessive dimensional changes caused by auto-claving a constrained membrane filter could crack the filter,rendering it useless. Increased hydrophobicity, due to autoclav-ing would result in an increase
8、d time to filter a liquid sample.6. Apparatus6.1 Calipers, accurate to 0.1 mm.6.2 Holder, suitable for membrane filters.6.3 Autoclave.6.4 Petri Dish.6.5 Forceps, nonserrated.6.6 Stopwatch, accurate to 0.1 s.7. Purity of Reagents7.1 Water used in these test methods should be reagentgrade Type IV B of
9、 Specification D 1193 or higher quality.8. Procedure8.1 Test Method ADimensional Stability:8.1.1 Measure the diameter of nonautoclaved filters, acrossthree axes, with the calipers. Use three disks for each test.8.1.2 Wrap filters in suitable autoclave packaging and auto-clave at 121C for 10 min.8.1.
10、3 Remove the packaged filter disks from the autoclaveafter the slow exhaust cycle and allow to equilibrate to ambienttemperature.8.1.4 Measure the diameter of the disks as in 8.1.1.8.2 Test Method BWetting Characteristics:8.2.1 Partially fill a petri dish with water containing 1 %methylene blue dye.
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
10000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ASTMD419919822003STANDARDTESTMETHODSFORAUTOCLAVABILITYOFMEMBRANEFILTERS 过滤器 高压 消毒 试验 方法 PDF

链接地址:http://www.mydoc123.com/p-516530.html