ASTM D4052-1996(2002)e1 Standard Test Method for Density and Relative Density of Liquids by Digital Density Meter《数字密度计测定液体密度和相对密度的标准试验方法》.pdf
《ASTM D4052-1996(2002)e1 Standard Test Method for Density and Relative Density of Liquids by Digital Density Meter《数字密度计测定液体密度和相对密度的标准试验方法》.pdf》由会员分享,可在线阅读,更多相关《ASTM D4052-1996(2002)e1 Standard Test Method for Density and Relative Density of Liquids by Digital Density Meter《数字密度计测定液体密度和相对密度的标准试验方法》.pdf(5页珍藏版)》请在麦多课文档分享上搜索。
1、Designation: D 4052 96 (Reapproved 2002)e1Designation: 365/84(86)An American National StandardStandard Test Method forDensity and Relative Density of Liquids by Digital DensityMeter1This standard is issued under the fixed designation D 4052; the number immediately following the designation indicates
2、 the year oforiginal adoption or, in the case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. Asuperscript epsilon (e) indicates an editorial change since the last revision or reapproval.This method was adopted as a joint ASTM-IP standard in 198
3、4.This standard has been approved for use by agencies of the Department of Defense.e1NOTEWarning notes were placed in the text editorially in May 2002.1. Scope1.1 This test method covers the determination of the densityor relative density of petroleum distillates and viscous oils thatcan be handled
4、in a normal fashion as liquids at test tempera-tures between 15 and 35C. Its application is restricted toliquids with vapor pressures below 600 mm Hg (80 kPa) andviscosities below about 15 000 cSt (mm2/s) at the temperatureof test.1.2 This test method should not be applied to samples sodark in color
5、 that the absence of air bubbles in the sample cellcannot be established with certainty. For the determination ofdensity in crude oil samples use Test Method D 5002.1.3 The accepted units of measure for density are grams permillilitre or kilograms per cubic metre.1.4 This standard does not purport t
6、o address all of thesafety concerns, if any, associated with its use. It is theresponsibility of the user of this standard to establish appro-priate safety and health practices and determine the applica-bility of regulatory limitations prior to use. For specific hazardstatements, see 7.4, 7.5, and 9
7、.1.2. Referenced Documents2.1 ASTM Standards:D 1193 Specification for Reagent Water2D 1250 Guide for Petroleum Measurement Tables3D 4057 Practice for Manual Sampling of Petroleum andPetroleum Products4D 4177 Practice for Automatic Sampling of Petroleum andPetroleum Products4D 4377 Test Method for Wa
8、ter in Crude Oils by Potentio-metric Karl Fischer Titration4D 5002 Test Method for Density and Relative Density ofCrude Oils by Digital Density Analyzer43. Terminology3.1 Definitions:3.1.1 densitymass per unit volume at a specified tempera-ture.3.1.2 relative densitythe ratio of the density of a mat
9、erialat a stated temperature to the density of water at a statedtemperature.4. Summary of Test Method4.1 A small volume (approximately 0.7 mL) of liquidsample is introduced into an oscillating sample tube and thechange in oscillating frequency caused by the change in themass of the tube is used in c
10、onjunction with calibration data todetermine the density of the sample.5. Significance and Use5.1 Density is a fundamental physical property that can beused in conjunction with other properties to characterize boththe light and heavy fractions of petroleum and petroleumproducts.5.2 Determination of
11、the density or relative density ofpetroleum and its products is necessary for the conversion ofmeasured volumes to volumes at the standard temperature of15C.1This test method is under the jurisdiction of ASTM Committee D02 onPetroleum Products and Lubricants and is the direct responsibility of Subco
12、mmitteeD02.04.0D on Physical and Chemical Methods.Current edition approved Apr. 10, 1996. Published June 1996. Originallypublished as D 4052 81. Last previous edition D 4052 95.2Annual Book of ASTM Standards, Vol 11.01.3Annual Book of ASTM Standards, Vol 05.01.4Annual Book of ASTM Standards, Vol 05.
13、02.1Copyright ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959, United States.6. Apparatus6.1 Digital Density AnalyzerA digital analyzer consistingof a U-shaped, oscillating sample tube and a system forelectronic excitation, frequency counting, and display. Th
14、eanalyzer must accommodate the accurate measurement of thesample temperature during measurement or must control thesample temperature as described in 6.2. The instrument shall becapable of meeting the precision requirements described in thistest method.6.2 Circulating Constant-Temperature Bath, (opt
15、ional) ca-pable of maintaining the temperature of the circulating liquidconstant to 60.05C in the desired range. Temperature controlcan be maintained as part of the density analyzer instrumentpackage.6.3 Syringes, at least 2 mL in volume with a tip or anadapter tip that will fit the opening of the o
16、scillating tube.6.4 Flow-Through or Pressure Adapter, for use as analternative means of introducing the sample into the densityanalyzer either by a pump or by vacuum.6.5 Thermometer, calibrated and graduated to 0.1C, and athermometer holder that can be attached to the instrument forsetting and obser
17、ving the test temperature. In calibrating thethermometer, the ice point and bore connections should beestimated to the nearest 0.05C.7. Reagents and Materials7.1 Purity of ReagentsReagent grade chemicals shall beused in all tests. Unless otherwise indicated, it is intended thatall reagents shall con
18、form to the specifications of the Commit-tee on Analytical Reagents of the American Chemical Society,where such specifications are available.5Other grades may beused, provided it is first ascertained that the reagent is ofsufficiently high purity to permit its use without lessening theaccuracy of th
19、e determination.7.2 Purity of WaterUnless otherwise indicated, referencesto water shall be understood to mean reagent water as definedby Type II of Specification D 1193.7.3 Water, redistilled, freshly boiled and cooled reagentwater for use as a primary calibration standard.7.4 Petroleum Naphtha6for
20、flushing viscous petroleumsamples from the sample tube. (WarningExtremely flam-mable.)7.5 Acetone, for flushing and drying the sample tube.(WarningExtremely flammable.)7.6 Dry Airfor blowing the oscillator tube.8. Sampling, Test Specimens, and Test Units8.1 Sampling is defined as all the steps requi
21、red to obtain analiquot of the contents of any pipe, tank, or other system, andto place the sample into the laboratory test container. Thelaboratory test container and sample volume shall be ofsufficient capacity to mix the sample and obtain a homoge-neous sample for analysis.8.2 Laboratory SampleUs
22、e only representative samplesobtained as specified in Practices D 4057 or D 4177 for this testmethod.8.3 Test SpecimenAportion or volume of sample obtainedfrom the laboratory sample and delivered to the densityanalyzer sample tube. The test specimen is obtained as follows:8.3.1 Mix the sample if req
23、uired to homogenize. The mixingmay be accomplished as described in Practice D 4177 (Section11) or Test Method D 4377 (A.1). Mixing at room temperaturein an open container can result in the loss of volatile material,so mixing in closed, pressurized containers or at sub-ambienttemperatures is recommen
24、ded.8.3.2 Draw the test specimen from a properly mixed labo-ratory sample using an appropriate syringe.Alternatively, if theproper density analyzer attachments and connecting tubes areused then the test specimen can be delivered directly to theanalyzers sample tube from the mixing container.9. Prepa
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
10000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ASTMD405219962002E1STANDARDTESTMETHODFORDENSITYANDRELATIVEDENSITYOFLIQUIDSBYDIGITALDENSITYMETER 数字 密度计

链接地址:http://www.mydoc123.com/p-516157.html