《ASTM D3268-1991(2011) Standard Test Method for Separation and Collection of Particulate and Gaseous Fluorides in the Atmosphere (Sodium Bicarbonate-Coated Glass Tube and Particulat.pdf》由会员分享,可在线阅读,更多相关《ASTM D3268-1991(2011) Standard Test Method for Separation and Collection of Particulate and Gaseous Fluorides in the Atmosphere (Sodium Bicarbonate-Coated Glass Tube and Particulat.pdf(6页珍藏版)》请在麦多课文档分享上搜索。
1、Designation: D3268 91 (Reapproved 2011)Standard Test Method forSeparation and Collection of Particulate and GaseousFluorides in the Atmosphere (Sodium Bicarbonate-CoatedGlass Tube and Particulate Filter Method)1This standard is issued under the fixed designation D3268; the number immediately followi
2、ng the designation indicates the year oforiginal adoption or, in the case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. Asuperscript epsilon () indicates an editorial change since the last revision or reapproval.This standard has been approved
3、 for use by agencies of the Department of Defense.1. Scope1.1 The sodium bicarbonate-coated glass tube and mem-brane filter method provides a means for the separation andcollection of gaseous atmospheric forms of fluoride reactivewith sodium bicarbonate and particulate forms of fluoridewhich are col
4、lected by a filter. The test method is applicable to12-h sampling periods, collecting 1 to 500 g of gaseousfluoride at a 15 L/min (0.5 ft3/min) sampling rate or about 0.1to 50 g/m3. The length of the sampling period can therefore beadjusted so that the amount of fluoride collected will fall withinth
5、is range. The actual lower limit of the test method willdepend upon the sensitivity of the analytical method employedand the quality of reagents used in tube preparation andanalysis. It is recommended that the lower limit of detectionshould be considered as two times the standard deviation of themon
6、thly arithmetic mean blank value. Any values greater thanthe blank by less than this amount should be reported as “blankvalue.”1.2 The values stated in SI units are to be regarded asstandard. The inch-pound units in parentheses may be approxi-mate.1.3 This standard does not purport to address all of
7、 thesafety concerns, if any, associated with its use. It is theresponsibility of the user of this standard to establish appro-priate safety and health practices and determine the applica-bility of regulatory limitations prior to use.2. Referenced Documents2.1 ASTM Standards:2D1193 Specification for
8、Reagent WaterD1356 Terminology Relating to Sampling and Analysis ofAtmospheresD1357 Practice for Planning the Sampling of the AmbientAtmosphereD3266 Test Method for Automated Separation and Collec-tion of Particulate and Acidic Gaseous Fluoride in theAtmosphere (Double Paper Tape Sampler Method)D326
9、7 Test Method for Separation and Collection of Par-ticulate and Water-Soluble Gaseous Fluorides in the At-mosphere (Filter and Impinger Method)D3269 Test Methods for Analysis for Fluoride Content ofthe Atmosphere and Plant Tissues (Manual Procedures)3D3270 Test Methods for Analysis for Fluoride Cont
10、ent ofthe Atmosphere and Plant Tissues (SemiautomatedMethod)3. Terminology3.1 DefinitionsFor definitions of terms used in this testmethod refer to Terminology D1356.4. Summary of Test Method4.1 Gaseous fluorides are removed from the air stream byreaction with sodium bicarbonate coated on the inside
11、wall ofa borosilicate glass tube (Note 1). Particulate fluorides arecollected on a filter following the tube. The fluoride collectedby the tube is eluted with water or buffer and analyzed forfluoride. The particulate matter collected by the filter is elutedwith acid and analyzed for fluoride (1-4)4.
12、 The results arereported as g/m3of gaseous or particulate in air at 25C(77F) and 101.3 kPa (29.92 in. Hg).NOTE 1Some particulate matter will collect on the wall of the sampletube. If this loss is to be evaluated, use test methods such as Test MethodD3266 or Test Method D3267 for comparison since the
13、 filter for collectingparticulate precedes the absorbers for gases Mandl and Weinstein (2)provide some information relative to potential loss of particulate matter.1This test method is under the jurisdiction of ASTM Committee D22 on AirQuality and is the direct responsibility of Subcommittee D22.03
14、on AmbientAtmospheres and Source Emissions.Current edition approved Oct. 1, 2011. Published October 2011. Originallyapproved in 1973. Last previous edition approved in 2005 as D3268 - 91 (2005).DOI: 10.1520/D3268-91R11.2For referenced ASTM standards, visit the ASTM website, www.astm.org, orcontact A
15、STM Customer Service at serviceastm.org. For Annual Book of ASTMStandards volume information, refer to the standards Document Summary page onthe ASTM website.3Withdrawn. The last approved version of this historical standard is referencedon www.astm.org.4The boldface numbers in parentheses refer to r
16、eferences at the end of this testmethod.1Copyright ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959, United States.5. Significance and Use5.1 The sodium bicarbonate coated tube filter method pro-vides a means of separating and collecting atmospheric gas-eous f
17、luoride and particulate fluoride samples.5.2 Since the samples are collected on the dry tube andfilter, the fluoride may be eluted with a small volume of eluant(see Section 10 for specific instructions on fluoride elution).Elution into a small volume and the sensitivity of the analyticalmethods empl
18、oyed allow the analysis of the collected fluorideto fractional parts of a microgram per cubic metre on samplestaken for a 12-h period.6. Interferences6.1 Significant amounts of acid aerosols or gases mightneutralize or acidify the bicarbonate coating and preventquantitative uptake of gaseous fluorid
19、e from the atmosphere. Ifthis potential interference needs to be evaluated, the alkalinityof the water extract may provide relevant information.6.2 The presence of large amounts of aluminum or certainother metals or phosphates can interfere with subsequentanalyses of the tubes or filters by calorime
20、tric or electrometricmethods. This is a problem inherent with any collectionmethod for fluoride.7. Apparatus7.1 Glass Tubing1200-mm (4-ft) lengths of 7-mm insidediameter borosilicate glass tubing, coated with sodium bicar-bonate, in accordance with the requirements outlined in 7.6.7.2 Filter and Hol
21、derConnect the tubing directly to thefilter holder and filter for the collection of particulate matter forparticulate fluoride analysis (see Fig. 2).NOTE 2Use of material other than that recommended in footnote 6 orFig. 2 will result in gaseous fluoride absorption on the material.7.3 Air Sampling Sy
22、stem:7.3.1 The tube and filter are followed by an air samplingsystem which is capable of sampling at a rate of 15 L/min (0.5ft3/min) and measuring the total air sampled on a time ratebasis or with a totalizing meter. See Test Method D3267 forsampling equipment, and the configuration and calibration.
23、FIG. 1 Sodium Bicarbonate-Coated Glass Tube Illustrating Simple Heating DeviceD3268 91 (2011)27.3.2 The system shall be equipped so that pressure andtemperature of the gas at the point of metering also are knownfor correcting sample volumes to standard conditions of 101.3kPa (29.92 in. Hg) at 25C (7
24、7F).7.3.3 Assemble the sampling system so that the inlet of thetube is 4 to 6 m above ground level (see Practice D1357) andprotected from rain in such a manner as not to interfere with thefree passage of aerosol fluorides.7.4 Light Bulb or Cone Heater, 30-W, installed to heat thegases to a temperatu
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
10000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ASTMD326819912011STANDARDTESTMETHODFORSEPARATIONANDCOLLECTIONOFPARTICULATEANDGASEOUSFLUORIDESINTHEATMOSPHERESODIUMBICARBONATECOATEDGLASSTUBEANDPARTICULATPDF

链接地址:http://www.mydoc123.com/p-514455.html