ASTM D1545-2007 Standard Test Method for Viscosity of Transparent Liquids by Bubble Time Method《用起泡时间法测定透明液体粘度的标准试验方法》.pdf
《ASTM D1545-2007 Standard Test Method for Viscosity of Transparent Liquids by Bubble Time Method《用起泡时间法测定透明液体粘度的标准试验方法》.pdf》由会员分享,可在线阅读,更多相关《ASTM D1545-2007 Standard Test Method for Viscosity of Transparent Liquids by Bubble Time Method《用起泡时间法测定透明液体粘度的标准试验方法》.pdf(3页珍藏版)》请在麦多课文档分享上搜索。
1、Designation: D 1545 07Standard Test Method forViscosity of Transparent Liquids by Bubble Time Method1This standard is issued under the fixed designation D 1545; the number immediately following the designation indicates the year oforiginal adoption or, in the case of revision, the year of last revis
2、ion. A number in parentheses indicates the year of last reapproval. Asuperscript epsilon (e) indicates an editorial change since the last revision or reapproval.This standard has been approved for use by agencies of the Department of Defense.1. Scope1.1 This test method covers the determination of t
3、he viscos-ity in bubble seconds by timing. The bubble seconds areapproximately equal to stokes for most liquids.1.2 The test method is applicable to transparent liquids thatare free from crystalline or gel particles.1.3 The values stated in SI units are to be regarded asstandard. No other units of m
4、easurement are included in thisstandard.1.4 This standard does not purport to address all of thesafety concerns, if any, associated with its use. It is theresponsibility of the user of this standard to establish appro-priate safety and health practices and determine the applica-bility of regulatory
5、limitations prior to use.2. Terminology2.1 Definitions:2.1.1 viscosity, nthe resistance experienced by one por-tion of a liquid moving over another portion of the liquid. Theabsolute unit of viscosity in the cgs, centimeter-gram-second,system is the poise which is expressed as dyne-seconds persquare
6、 centimetre. Stokes are equal to poises divided bydensity. The absolute SI, International System of Units (metricsystem), viscosity unit is the pascal-second.3. Apparatus3.1 Constant-Temperature BathAny suitable bath capableof maintaining temperature at 25 6 0.1C with water as thebath medium.3.2 Sta
7、ndard Viscosity Tubes of clear glass and with flatbottoms, 10.65 6 0.025 mm in inside diameter; 114 6 1mmin outside length. Plainly legible lines shall be located asfollows (Note 1):27 6 0.5 mm100 6 0.5 mm108 6 0.5 mmThe distance between the first and second lines shall be 73 60.5 mm.NOTE 1All dista
8、nces shall be measured from the outside bottom ofthe tube.3.3 Reference StandardsA series of standard viscositytubes (3.2) filled with transparent liquids having predeterminedviscosities in stokes and bubble seconds and spaced in loga-rithmically even increments of about 26 % (log 1.260 = 0.100),as
9、listed in Table 1. The standards shall be marked alphabeti-cally or numerically as shown in Table 1 without reference tostokes or bubble time (Note 2). Also listed in Table 1, forgeneral and historical reference only, is the long establishedseries of Gardner-Holdt letter standards, in tubes that do
10、nothave the three lines and are shorter than the standard viscositytubes described in 3.2.NOTE 2For convenience, the reference standards can be divided inthree series: light series, 15 tubes marked 0.22 to 8.0; heavy series, 14tubes marked 10 to 200; and very heavy series, 7 tubes, marked 250 to1000
11、.3.4 Timing DeviceStopwatch or electric stop clock ca-pable of being read to a precision of 0.1 s.3.5 Tube Racks, capable of inverting one or more viscositytubes 180 to within 1 of a vertical position while rack andtubes are immersed in the constant temperature bath.3.6 Viscosity Tube Corks, No. 2 s
12、hort.4. Procedure4.1 Fill a standard viscosity tube with the material to betested to approximately level with the 108-mm line.4.2 Transfer the tube to a constant 25C temperature bathwith the cork loosely inserted. Hold at this temperature for 10min.NOTE 3Adequate control of the temperature bath is e
13、ssential. Avariation of 0.1C in the temperature of the bath will cause a 1 % variationin the timed bubble travel.1This test method is under the jurisdiction of ASTM Committee D01 on Paintand Related Coatings, Materials, and Applications and is the direct responsibility ofSubcommittee D01.33 on Polym
14、ers and Resins.Current edition approved July 1, 2007. Published August 2007. Originallyapproved in 1958. Last previous edition approved in 1998 as D 1545 98 whichwas withdrawn March 2007 and reinstated in July 2007.1Copyright ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken,
15、 PA 19428-2959, United States.4.3 At the end of 10 min adjust the level of the liquid so thatthe bottom meniscus will be level with the 100-mm line. Insertthe cork so that the bottom of the cork is on the level with the108-mm line. This will ensure a bubble of suitable and uniformsize.4.4 Insert the
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
10000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ASTMD15452007STANDARDTESTMETHODFORVISCOSITYOFTRANSPARENTLIQUIDSBYBUBBLETIMEMETHOD 起泡 时间 测定 透明 液体 粘度 标准

链接地址:http://www.mydoc123.com/p-510000.html