ASTM C1168-2001 Standard Practice for Preparation and Dissolution of Plutonium Materials for Analysis《分析用钚材料的制备和溶解的标准实施规程》.pdf
《ASTM C1168-2001 Standard Practice for Preparation and Dissolution of Plutonium Materials for Analysis《分析用钚材料的制备和溶解的标准实施规程》.pdf》由会员分享,可在线阅读,更多相关《ASTM C1168-2001 Standard Practice for Preparation and Dissolution of Plutonium Materials for Analysis《分析用钚材料的制备和溶解的标准实施规程》.pdf(4页珍藏版)》请在麦多课文档分享上搜索。
1、Designation: C 1168 01Standard Practice forPreparation and Dissolution of Plutonium Materials forAnalysis1This standard is issued under the fixed designation C 1168; the number immediately following the designation indicates the year oforiginal adoption or, in the case of revision, the year of last
2、revision. A number in parentheses indicates the year of last reapproval. Asuperscript epsilon (e) indicates an editorial change since the last revision or reapproval.1. Scope1.1 This practice is a compilation of dissolution techniquesfor plutonium materials that are applicable to the test methodsuse
3、d for characterizing these materials. Dissolution treatmentsfor the major plutonium materials assayed for plutonium oranalyzed for other components are listed. Aliquants of thedissolved samples are dispensed on a weight basis when one ofthe analyses must be highly reliable, such as plutonium assay;o
4、therwise they are dispensed on a volume basis.1.2 The treatments, in order of presentation, are as follows:Procedure Title SectionDissolution of Plutonium Metal with Hydrochloric Acid 7.1Dissolution of Plutonium Metal with Sulfuric Acid 7.2Dissolution of Plutonium Oxide and Uranium-Plutonium MixedOx
5、ide by the Sealed-Reflux Technique7.3Dissolution of Plutonium Oxide and Uranium-Plutonium MixedOxides by Sodium Bisulfate Fusion7.4Dissolution of Uranium-Plutonium Mixed Oxides and Low-FiredPlutonium Oxide in Beakers7.51.3 The values stated in SI units are to be regarded asstandard.1.4 This standard
6、 does not purport to address all of thesafety concerns, if any, associated with its use. It is theresponsibility of the user of this standard to establish appro-priate safety and health practices and determine the applica-bility of regulatory limitations prior to use.2. Referenced Documents2.1 ASTM
7、Standards:C 757 Specification for Nuclear-Grade Plutonium DioxidePowder, Sinterable2C 833 Specification for Sintered (Uranium-Plutonium) Di-oxide Pellets2C 1008 Specification for Sintered (Uranium-Plutonium) Di-oxide PelletsFast Reactor Fuel23. Summary of Dissolution Methods3.1 Most plutonium-contai
8、ning samples are dissolved withvarious mineral acids and, except for plutonium metal, withapplied heat. Dissolution-resistant materials are dissolved inheated and sealed containers in which pressurization providesa higher temperature than is attained at ambient pressure.3.2 Another dissolution techn
9、ique is fusion of refractoryplutonium oxide with sodium bisulfate.3.3 The dissolved materials are quantitatively transferred toa tared polyethylene dispensing bottle for subsequent deliveriesof weight aliquants for high-precision analysis methods, suchas assays, or to a volumetric flask for deliveri
10、es of volumealiquants for less precise analysis methods, such as impurityanalyses. Acids, usually 1 M, are used as rinses to effectquantitative transfers and as diluents in the polyethylenedispensing bottles and volumetric flasks. The use of water forthese purposes can, in some cases, result in hydr
11、olysis ofplutonium to produce polymers that, although soluble, arenonreactive in separation treatments or in plutonium assaymethods that have no pretreatments, such as fuming with acid.3.4 Plutonium metal is dissolved with hydrochloric acid orwith sulfuric acid.3.5 Plutonium oxide, calcined at about
12、 1000C or lower, isdissolved with a mixture of hydrochloric, nitric, and hydrof-luoric acids using the sealed-reflux techniques (1).3Morerefractory plutonium oxide is dissolved with a fusion usingsodium bisuflate (2). Low-fired (650C) plutonium oxide canalso be dissolved in a mixture of nitric and h
13、ydrofluoric acidsin beakers. Plutonium oxide fired at temperatures above 650Cmay also be dissolved in beakers using a mixture of nitric andhydrofluoric acids when complete dissolution can be demon-strated.3.6 Uranium-plutonium mixed oxide is dissolved in eitherof three ways: sodium bisulfate fusion,
14、 a heated mixture ofnitric and hydrofluoric acids in a beaker, or a mixture ofhydrochloric, nitric, and hydrofluoric acids by the sealed-refluxtechnique.3.7 Combinations of these dissolution techniques describedin 3.4 to 3.6 are sometimes used on difficult-to-dissolvesamples.3.8 Quantitative transfe
15、rs of samples and subsequent solu-tion are required. Whenever a loss is incurred or even sus-pected, the sample is rejected. Solutions of dissolved samplesare inspected for undissolved particles; if particles are present,1This practice is under the jurisdiction of ASTM Committee C26 on NuclearFuel C
16、ycle and is the direct responsibility of Subcommittee C26.05 on Methods ofTest.Current edition approved Jan. 10, 2001. Published March 2001. Originallypublished as C 116890. Last previous edition C 116890(1995).2Annual Book of ASTM Standards, Vol 12.01.3The boldface numbers in parentheses refer to a
17、 list of references at the end ofthis practice.1Copyright ASTM, 100 Barr Harbor Drive, West Conshohocken, PA 19428-2959, United States.further treatment is necessary to attain complete solubility.When analyzing the dissolved sample for trace impurities,caution should be exercised so the dissolution
18、process does notcause the impurity to be lost or does not significantly increasethe level of impurity being determined.4. Significance and Use4.1 The materials covered are plutonium metal, plutoniumoxide, and uranium-plutonium mixed oxide, including thosethat must meet ASTM product specifications.4.
19、2 Plutonium and uranium mixtures are used as nuclearreactor fuels. For use as a nuclear reactor fuel, the materialmust meet certain criteria for combined uranium and pluto-nium content, effective fissile content, and impurity content asdescribed in Specifications C 757, C 833, and C 1008. Themateria
20、l is assayed for plutonium and uranium to determine ifthe content is correct as specified by the purchaser.4.3 The materials not covered are unique plutonium mate-rials, including alloys, compounds, and scrap materials. Theuser must determine the applicability of this practice to theseother material
21、s. In general, these unique plutonium materialsare dissolved with various acid mixtures or by fusion withvarious fluxes. Many plutonium salts are soluble in hydrochlo-ric acid.5. Apparatus5.1 Balances, for weighing samples and solutions. A bal-ance with a sensitivity of 0.1 mg is necessary; however,
22、 abalance with 0.01 mg sensitivity is more desirable. A calibratedbalance must be used.5.2 Beakers, Test Tubes, and Erlenmeyer FlasksGenerally, borosilicate glass is recommended; however, theanalyst should be sure that safety and sample contaminationfrom the container are considered when choosing ap
23、propriatecontainers.5.3 Furnace, with controller for timed operation. The fur-nace must be capable of maintaining an even temperature of610C up to 700C.5.4 Heating EquipmentHot plates and infrared lamps areused.5.5 Inert Atmosphere Glove-Box Systemcapability ofmaintaining less than 10 ppm of H2O and
24、 of O2is preferred.5.6 Sealed-Reflux Dissolution ApparatusThe example ap-paratus is shown in Fig. 1 and Fig. 2 and is further describedin Ref (1).6. ReagentsNOTE 1Use distilled or demineralized water for all reagents.6.1 Hydrochloric Acid (sp gr 1.18), 12 M.6.2 Hydrochloric Acid,6MAdd 500 mL of 12 M
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
10000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ASTMC11682001STANDARDPRACTICEFORPREPARATIONANDDISSOLUTIONOFPLUTONIUMMATERIALSFORANALYSIS 分析 材料 制备 溶解

链接地址:http://www.mydoc123.com/p-463598.html