《ASTM F2183 - 02(2008) Standard Test Method for Small Punch Testing of Ultra-High Molecular Weight Polyethylene Used in Surgical Implants (Withdrawn 2017).pdf》由会员分享,可在线阅读,更多相关《ASTM F2183 - 02(2008) Standard Test Method for Small Punch Testing of Ultra-High Molecular Weight Polyethylene Used in Surgical Implants (Withdrawn 2017).pdf(5页珍藏版)》请在麦多课文档分享上搜索。
1、Designation: F2183 02 (Reapproved 2008)Standard Test Method forSmall Punch Testing of Ultra-High Molecular WeightPolyethylene Used in Surgical Implants1This standard is issued under the fixed designation F2183; the number immediately following the designation indicates the year oforiginal adoption o
2、r, in the case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. Asuperscript epsilon () indicates an editorial change since the last revision or reapproval.1. Scope1.1 This test method covers the determination of mechanicalbehavior of ultra-high
3、molecular weight polyethylene (UHM-WPE) by small punch testing of miniature disk specimens (0.5mm in thickness and 6.4 mm in diameter). The test method hasbeen established for characterizing UHMWPE surgical mate-rials after ram extrusion or compression molding (1,2)2; forevaluating as-manufactured i
4、mplants after radiation crosslink-ing and sterilization (3,4); as well as for testing of implants thathave been retrieved (explanted) from the human body (5,6).1.2 The parameters of the small punch test, namely the peakload, ultimate displacement, ultimate load, and work to failure,provide metrics o
5、f the yielding, ultimate strength, ductility, andtoughness of UHMWPE under multiaxial loading conditions.Because the mechanical behavior of UHMWPE is differentwhen loaded under uniaxial and multiaxial loading conditions(3), the small punch test provides a complementary mechanicaltesting technique to
6、 the uniaxial tensile testing specified formedical grade UHMWPE by Specification F648.1.3 In addition to its use as a research tool in implantretrieval analysis, the small punch test can be used as alaboratory screening test to evaluate new UHMWPE materials,such as those created by gamma or electron
7、 beam irradiation(1). The test method is also well suited for characterization ofUHMWPE before and after accelerated aging (for example,Guide F2003), and in that regard it can provide ranking of themechanical degradation of different UHMWPE samples afteroxidative degradation (4,7).1.4 The small punc
8、h test has been applied to other polymers,including polymethyl methacrylate (PMMA) bone cement,polyacetal, and high density polyethylene (HDPE) (8,9).However, the small punch testing of polymers other thanUHMWPE is beyond the scope of this standard.1.5 The values stated in SI units are to be regarde
9、d asstandard. The units in parentheses are mathematical conver-sions to inch-pound units that are provided for informationonly and are not considered standard.1.6 This standard does not purport to address all of thesafety concerns, if any, associated with its use. It is theresponsibility of the user
10、 of this standard to establish appro-priate safety and health practices and determine the applica-bility of regulatory limitations prior to use.2. Referenced Documents2.1 ASTM Standards:3D695 Test Method for Compressive Properties of RigidPlasticsD883 Terminology Relating to PlasticsE4 Practices for
11、 Force Verification of Testing MachinesE83 Practice for Verification and Classification of Exten-someter SystemsF648 Specification for Ultra-High-Molecular-Weight Poly-ethylene Powder and Fabricated Form for Surgical Im-plantsF1714 Guide for Gravimetric WearAssessment of ProstheticHip Designs in Sim
12、ulator DevicesF1715 Guide for Wear Assessment of Prosthetic Knee De-signs in Simulator Devices (Withdrawn 2006)4F2003 Practice for Accelerated Aging of Ultra-High Mo-lecular Weight Polyethylene after Gamma Irradiation inAirF2102 Guide for Evaluating the Extent of Oxidation inPolyethylene Fabricated
13、Forms Intended for SurgicalImplants3. Terminology3.1 DefinitionsThe features of a typical small punch testload versus displacement curve for UHMWPE are illustrated inFig. 1.1This test method is under the jurisdiction of ASTM Committee F04 on Medicaland Surgical Materials and Devicesand is the direct
14、 responsibility of SubcommitteeF04.15 on Material Test Methods.Current edition approved May 1, 2008. Published June 2008. Originallyapproved in 2002. Last previous edition approved in 2002 as F2083 02. DOI:10.1520/F2183-02R08.2The boldface numbers in parentheses refer to the list of references at th
15、e end ofthis standard.3For referenced ASTM standards, visit the ASTM website, www.astm.org, orcontact ASTM Customer Service at serviceastm.org. For Annual Book of ASTMStandards volume information, refer to the standards Document Summary page onthe ASTM website.4The last approved version of this hist
16、orical standard is referenced onwww.astm.org.Copyright ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959. United StatesNOTICE: This standard has either been superseded and replaced by a new version or withdrawn.Contact ASTM International (www.astm.org) for the
17、latest information13.1.1 peak loadan initial local maximum in the loadversus displacement curve (Fig. 1). In certain radiation cross-linked UHMWPE materials, the load versus displacementcurve increases monotonically and a shoulder, rather than aninitial peak load, may be observed.3.1.2 small punch t
18、esta test wherein the specimen is ofminiature size relative to conventional mechanical testspecimens, is disk-shaped, and is loaded axisymmetrically inbending by a hemispherical-head punch.3.1.3 ultimate displacementthe displacement at rupture(failure) of the specimen (Fig. 1).3.1.4 ultimate loadthe
19、 load at rupture (failure) of thespecimen (Fig. 1).3.1.5 work to failurethe area under the load versus dis-placement curve (Fig. 1).4. Significance and Use4.1 Miniature specimen testing techniques are used to char-acterize the mechanical behavior of UHMWPE stock materialsand surgical implants after
20、manufacture, sterilization, shelfaging, radiation crosslinking, thermal treatment, and implanta-tion (1). Furthermore, experimental UHMWPE materials canbe evaluated after accelerated aging and hip or knee wearsimulation. Consequently, the small punch test makes it pos-sible to examine relationships
21、between wear performance andmechanical behavior of UHMWPE. This test method can alsobe used to rank the mechanical behavior of UHMWPE relativeto a reference control material (such as the NIST Ultra-HighMolecular Weight Polyethylene Reference Material #8456).4.2 Small punch testing results may vary w
22、ith specimenpreparation and with the speed and environment of testing.Consequently, where precise comparative results are desired,these factors must be carefully controlled.5. Apparatus5.1 Small Punch Test Apparatus5A system consisting of ahemispherical head punch, a die, and a guide for the punch,
23、asshown in Fig. 2. The parts shall be fabricated from a hardenedsteel.5.1.1 GuideThe function of the guide is to align the punchrelative to the specimen, which rests in a disk-shaped recess.The inner diameter of the guide bore shall be 0.1010 +0.0002/-0.0000 in. (2.565 +0.005/-0.000 mm), and the spe
24、cimen recessshall be 0.0200 +0.0004/-0.0000 in. (0.508 +0.010/-0.000 mm)in depth and 0.2520 6 0.0005 in. (6.401 6 0.013 mm) indiameter.5.1.2 DieThe function of the die is to constrain the sampleduring testing. The inner diameter of the die bore shall be0.1500 6 0.0005 in. (3.810 6 0.013 mm).5.1.3 Pu
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
5000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ASTM F2183 02 2008 Standard Test Method for Small Punch Testing of Ultra High Molecular Weight Polyethylene Used in Surgical Implants Withdrawn 2017

链接地址:http://www.mydoc123.com/p-287186.html