1、Designation: D7706 17Standard Practice forRapid Screening of VOC Emissions from Products UsingMicro-Scale Chambers1This standard is issued under the fixed designation D7706; the number immediately following the designation indicates the year oforiginal adoption or, in the case of revision, the year
2、of last revision. A number in parentheses indicates the year of last reapproval. Asuperscript epsilon () indicates an editorial change since the last revision or reapproval.1. Scope1.1 This practice describes a micro-scale chamber apparatusand associated procedures for rapidly screening materials an
3、dproducts for their vapor-phase emissions of volatile organiccompounds (VOCs) including formaldehyde and other carbo-nyl compounds. It is intended to complement, not replacereference methods for measuring chemical emissions forexample, small-scale chamber tests (Guide D5116) and emis-sion cell tests
4、 (Practice D7143).1.2 This practice is suitable for use in and outside oflaboratories, in manufacturing sites and in field locations withaccess to electrical power.1.3 Compatible material/product types that may be tested inthe micro-scale chamber apparatus include rigid materials,dried or cured pain
5、ts and coatings, compressible products, andsmall, irregularly-shaped components such as polymer beads.1.4 This practice describes tests to correlate emission resultsobtained from the micro-scale chamber with results obtainedfrom VOC emission reference methods (for example, GuideD5116, Test Method D6
6、007, Practice D7143, and ISO 16000-9and ISO 16000-10).1.5 The micro-scale chamber apparatus operates at moder-ately elevated temperatures, 30C to 60C, to eliminate theneed for cooling, to reduce test times, boost emission rates, andenhance analytical signals for routine emission screening, andto fac
7、ilitate screening of semi-volatile VOC (SVOC) emissionssuch as emissions of some phthalate esters and other plasticiz-ers.1.6 Gas sample collection and chemical analysis are depen-dent upon the nature of the VOCs targeted and are beyond thescope of this practice. However, the procedures described in
8、Test Method D7339, Practice D6196 and ISO 16000-6 foranalysis of VOCs and in Test Method D5197 and ISO 16000-3for analysis of formaldehyde and other carbonyl compoundsare applicable to this practice.1.7 The values stated in SI units are to be regarded asstandard. No other units of measurement are in
9、cluded in thisstandard.1.8 This standard does not purport to address all of thesafety concerns, if any, associated with its use. It is theresponsibility of the user of this standard to establish appro-priate safety and health practices and determine the applica-bility of regulatory limitations prior
10、 to use.1.9 This international standard was developed in accor-dance with internationally recognized principles on standard-ization established in the Decision on Principles for theDevelopment of International Standards, Guides and Recom-mendations issued by the World Trade Organization TechnicalBar
11、riers to Trade (TBT) Committee.2. Referenced Documents2.1 ASTM Standards:2D1356 Terminology Relating to Sampling and Analysis ofAtmospheresD1914 Practice for Conversion Units and Factors Relating toSampling and Analysis of AtmospheresD5116 Guide for Small-Scale Environmental Chamber De-terminations
12、of Organic Emissions from Indoor Materials/ProductsD5197 Test Method for Determination of Formaldehyde andOther Carbonyl Compounds inAir (Active Sampler Meth-odology)D5337 Practice for Flow Rate Adjustment of Personal Sam-pling PumpsD6007 Test Method for Determining Formaldehyde Concen-trations in A
13、ir from Wood Products Using a Small-ScaleChamberD6196 Practice for Choosing Sorbents, Sampling Param-eters and Thermal Desorption Analytical Conditions forMonitoring Volatile Organic Chemicals in AirD7143 Practice for Emission Cells for the Determination ofVolatile Organic Emissions from Indoor Mate
14、rials/Products1This practice is under the jurisdiction of ASTM Committee D22 on Air Qualityand is the direct responsibility of Subcommittee D22.05 on Indoor Air.Current edition approved May 1, 2017. Published June 2017. Originallyapproved in 2011. Last previous edition approved in 2011 as D7706 11.
15、DOI:10.1520/D7706-17.2For referenced ASTM standards, visit the ASTM website, www.astm.org, orcontact ASTM Customer Service at serviceastm.org. For Annual Book of ASTMStandards volume information, refer to the standards Document Summary page onthe ASTM website.Copyright ASTM International, 100 Barr H
16、arbor Drive, PO Box C700, West Conshohocken, PA 19428-2959. United StatesThis international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for theDevelopment of International Standards, Guides and Recommend
17、ations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.1Mon Apr 30 24 D7339 Test Method for Determination of Volatile OrganicCompounds Emitted from Carpet using a Specific SorbentTube and Thermal Desorption / Gas Chromatography2.2 ISO Standards:3ISO 16000-3 Determi
18、nation of formaldehyde and other car-bonyl compounds Active sampling methodISO 16000-6 Determination of volatile organic compoundsin indoor and test chamber air by active sampling onTenax TA sorbent, thermal desorption and gas-chromatography using MS/FIDISO 16000-9 Indoor airPart 9: Determination of
19、 the emis-sion of volatile organic compounds Emission testchamber methodISO 16000-10 Indoor airPart 10: Determination of theemission of volatile organic compounds Emission testcell method2.3 Other Standard:4U.S. EPA Method TO-17 Determination of volatile organiccompounds in ambient air using active
20、sampling ontosorbent tubes3. Terminology3.1 DefinitionsFor definitions and terms commonly usedfor sampling and analysis of atmospheres, refer to TerminologyD1356. For definitions and terms commonly used when testingmaterials and products for VOC emissions, refer to GuideD5116. For an explanation of
21、general units, symbols andconversion factors, refer to Practice D1914.3.2 Definitions of Terms Specific to This Standard:3.2.1 micro-scale test chamber, nan environmental testchamber ranging in volume from a few milliliters to about 250mL and designed to operate at moderately elevated tempera-tures
22、that is used to measure vapor-phase organic emissionsfrom small specimens of solid materials and products.3.2.2 control level, na user-defined acceptance criterionfor a micro-scale chamber test, for example, presence orabsence of a target compound and/or a concentration oremission rate of a target c
23、ompound, typically used in produc-tion quality control to indicate that the tested product samplelikely will meet the corresponding acceptance criterion for areference test.4. Principles4.1 Micro-scale test chambers operate under the same masstransfer principles as conventional small-scale test cham
24、bersand cells for measuring emissions of VOCs including formal-dehyde and other carbonyl compounds from materials andproducts (see Guide D5116 and Practice D7143).4.2 Clean gas (dry nitrogen or air) is supplied to a micro-scale chamber and passes over the exposed surface of the testspecimen before r
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
5000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ASTM D7706 17 Standard Practice for Rapid Screening of VOC Emissions from Products Using Micro Scale Chambers

链接地址:http://www.mydoc123.com/p-286611.html