ASTM D7059-09(2017) Standard Test Method for Determination of Methanol in Crude Oils by Multidimensional Gas Chromatography.pdf
《ASTM D7059-09(2017) Standard Test Method for Determination of Methanol in Crude Oils by Multidimensional Gas Chromatography.pdf》由会员分享,可在线阅读,更多相关《ASTM D7059-09(2017) Standard Test Method for Determination of Methanol in Crude Oils by Multidimensional Gas Chromatography.pdf(15页珍藏版)》请在麦多课文档分享上搜索。
1、Designation: D7059 09 (Reapproved 2017)Standard Test Method forDetermination of Methanol in Crude Oils byMultidimensional Gas Chromatography1This standard is issued under the fixed designation D7059; the number immediately following the designation indicates the year oforiginal adoption or, in the c
2、ase of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. Asuperscript epsilon () indicates an editorial change since the last revision or reapproval.1. Scope1.1 This test method covers the determination of methanolin crude oils by direct injection mu
3、ltidimensional gas chroma-tography in the concentration range of 15 ppm (m/m) to900 ppm (m/m). The pooled limit of quantification (PLOQ) is15 ppm (m/m).1.2 This test method is applicable only to crude oils con-taining less than or equal to 0.1 % (v/v) water.1.3 This test method has not been tested w
4、ith crude oilsamples that are solid or waxy, or both, at ambient tempera-tures.1.4 The values stated in SI units are to be regarded asstandard. Alternate units, in common usage, are also providedto increase clarity and aid the users of this test method.1.5 This standard does not purport to address a
5、ll of thesafety concerns, if any, associated with its use. It is theresponsibility of the user of this standard to establish appro-priate safety, health, and environmental practices and deter-mine the applicability of regulatory limitations prior to use.1.6 This international standard was developed
6、in accor-dance with internationally recognized principles on standard-ization established in the Decision on Principles for theDevelopment of International Standards, Guides and Recom-mendations issued by the World Trade Organization TechnicalBarriers to Trade (TBT) Committee.2. Referenced Documents
7、2.1 ASTM Standards:2D4006 Test Method for Water in Crude Oil by DistillationD4057 Practice for Manual Sampling of Petroleum andPetroleum ProductsD4307 Practice for Preparation of Liquid Blends for Use asAnalytical StandardsD4928 Test Method for Water in Crude Oils by CoulometricKarl Fischer Titratio
8、nD6596 Practice for Ampulization and Storage of Gasolineand Related Hydrocarbon Materials3. Terminology3.1 Definitions:3.1.1 analytical column, nporous layer open tubular(PLOT) column with a stationary phase selective for oxygen-ates. It is used to resolve methanol from 1-propanol to provideaccurate
9、 quantitative results.3.1.2 cool-on-column injector, nan injection port thatallows controlled injection of the sample at a temperature closeto or lower than the boiling point of the solvent into the gaschromatographic column or a liner within the injection portconnected to the column.3.1.2.1 Discuss
10、ionAfter the injection, the injection port isheated at a fixed rate to a temperature sufficiently high enoughto allow the transfer of sample components of interest from theinjection port to the part of the column located in the gaschromatograph (GC) oven.3.1.3 electronic pressure control, nelectroni
11、c pneumaticcontrol of carrier gas flows. It can be flow or pressureprogrammed to speed up elution of components.3.1.4 low-volume connector, na special union for connect-ing two lengths of tubing 1.6-mm inside diameter and smaller;sometimes referred to as a zero dead-volume union.3.1.5 pre-column, na
12、 polydimethylsiloxane WCOT col-umn used to isolate the methanol and 1-propanol and severallight hydrocarbons from the higher boiling portion of the crudeoil sample for transfer to the analytical column for furtherseparation and quantification.3.1.6 programmable temperature vaporizer (PTV), natempera
13、ture programmable injector similar to a cool-on-column injector except that the sample is injected cool into aglass liner or insert instead of the WCOT (3.1.5) column andthen the temperature is programmed in a manner similar to theon-column injector.3.1.6.1 DiscussionThe liner may be replaced, as1Th
14、is test method is under the jurisdiction of ASTM Committee D02 onPetroleum Products, Liquid Fuels, and Lubricantsand is the direct responsibility ofSubcommittee D02.04.0L on Gas Chromatography Methods.Current edition approved Oct. 1, 2017. Published November 2017. Originallyapproved in 2004. Last pr
15、evious edition approved in 2013 as D7059 09 (2013).DOI: 10.1520/D7059-09R17.2For referenced ASTM standards, visit the ASTM website, www.astm.org, orcontact ASTM Customer Service at serviceastm.org. For Annual Book of ASTMStandards volume information, refer to the standards Document Summary page onth
16、e ASTM website.Copyright ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959. United StatesThis international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for theDevel
17、opment of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.159 necessary, to remove non-volatile materials. This injector maybe operated in low split mode or direct (no splitting) mode.3.1.7 split/splitless injecto
18、r, na heated capillary inlet orsample introduction system that allows controlled splitting ofthe injected sample into two unequal portions, the smaller ofwhich goes to the capillary column, and the greater to a vent.3.1.7.1 DiscussionWhen the vent is closed, the entiresample enters the capillary col
19、umn and the inlet is operated asa splitless injector. When the vent is open, the inlet is operatedin the split mode and only a portion of the sample reaches thecapillary column. The ratio of the split between the capillarycolumn and the vent is calculated as described in 3.1.7.1.3.1.7.1 split ratio,
20、 nin capillary gas chromatography, theratio of the total flow of carrier gas to the sample inlet versusthe flow of the carrier gas to the capillary column, expressedby:split ratio 5 S1C!/C (1)where:S = flow rate at the splitter vent, andC = flow rate at the column outlet.4. Summary of Test Method4.1
21、 An internal standard, 1-propanol, is added to the sample,which is then introduced into a gas chromatograph equippedwith two columns and a flow switching system between thetwo columns. The sample first passes through the polydimeth-ylsiloxane WCOT column that performs a pre-separation of themethanol
22、 and 1-propanol and eliminates unwanted hydrocar-bons. The methanol and 1-propanol are transferred to theanalytical PLOT column for oxygenates. While the methanoland 1-propanol are eluting from the analytical PLOT columnfor oxygenates, auxiliary carrier gas is used to elute higherboiling crude oil h
23、ydrocarbons from the pre-column, either inthe forward or backflush mode, to yield a stable baseline for thenext analysis.5. Significance and Use5.1 Methanol is used in production of crude oil to preventformation of gas hydrates. The presence of residual methanolin crude oils can lead to costly probl
24、ems in refinery operations.6. Apparatus6.1 ChromatographA multidimensional two-WCOT col-umn gas chromatographic system, capable of adequately re-solving methanol and the 1-propanol internal standard and ofeliminating hydrocarbon and other interferences, is requiredfor this analysis. Flow switching b
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
5000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ASTM D7059 09 2017 Standard Test Method for Determination of Methanol in Crude Oils by Multidimensional Gas Chromatography

链接地址:http://www.mydoc123.com/p-286555.html