1、w- 3 A-A-54444 65 = 9999974 0062166 8 r A-A-54 4 4 4 19 July 1991 a e AMSC N/A FSC 6530 DISTRIBUTION STATEMENT A: distribution is unlimited. Approved for public release; COMMERCIAL ITEM DESCRIPTION VENTILATOR, VOLUME, PORTABLE The General Services Administration has authorized the use of this Commer
2、cial Item Description This Commercial Item Description covers a self-contained, portable volume ventilator which operates on 11-30 volts DC and also on 110/230 volts AC, 50-400 Hz when used with the separate AC power supply furnished. Salient Characteristics: The portable volume ventilator containin
3、g time cycling, pressure limiting, constant flow, and microprocessor (CPU) control shall be suitable for use in hospitals, ambulances, fields, and in military aeromedical aircraft. It shall be suitable for providing controlled ventilations to the non- breathing patient and for assisting ventilations
4、 to the spontaneously breathing patient. The ventilator shall be a self-contained apparatus except for a compressed gas source and shall contain at least the following: a control/display panel; internal battery; battery recharge circuit; EMI/RFI suppression circuitry; valve to select flow of gas; al
5、arms; and chart depicting tidal volume by gas flow/inspiratory time. The apparatus with above mentioned items shall be self-contained and have its own separate carrying case with condensed operating instructions marked directly on the case or on a label attached to the case. the specified Operating
6、and Maintenance Manuals. The case shall be a semi-rigid plastic or corrosion-resistant metal material with integral carrying handle, and shall have a hinged, clear door to protect the operator controls of the control/display panel from inadvertent contact and provide splash protection. These instruc
7、tions are in addition to Provided by IHSNot for ResaleNo reproduction or networking permitted without license from IHS-,-,-A-A-54444 65 9999974 0062Lb7 T = A-A-54 4 4 4 The internal battery shall be a rechargeable 12 volt battery. Two batteries wired to provide 12 volts is acceptable. The ventilator
8、 battery shall be rechargeable from any of the following inputs: 11 to 30 volts DC, positive or negative ground; and nominal. 110/230 volts AC, 50-400 Hz, single phase. A totally depleted internal battery shall be capable of being fully recharged within 24 hours from any of the voltages cited. The v
9、entilator apparatus shall include a patient valve with a removable PEEP valve (attached), all tubing and connectors required for complete ventilator system operation, a separate AC power supply with power cord assembly, and operating/maintenance manuals. All of the above, including the case containi
10、ng the ventilator, shall be contained in a carry-all, semi-rigid carrying case. The apparatus shall have a separate AC power supply. power supply shall contain an Input Voltage selector switch to allow selection of either 110-125 volts, 50-400 Hz, single phase service or nominal 220-250 volts AC, 50
11、-400 Hz, single phase service. electromagnetic interference/radio frequency interference requirements. relieved line cord, at least 6 feet long, consisting of SJT- type wire with a minimum rating of 18/3 AWG, and terminating in a 3 conductor male hospital-grade plug. The output cable of this supply
12、shall be at least 4 feet long and terminate in a suitable, non-shorting type plug to interface with the ventilator apparatus. The power supply output voltage shall be within the range of acceptance of the ventilator apparatus described herein. The AC This supply shall conform with the It shall also
13、contain an appropriately strain- The ventilator shall be operational from the 12 volt internal battery and from external DC sources of 11-30 volts including, vehicle generating system rated 120 amperes at 14 volts; 120 amp-hour heavy duty 12 volt battery, and aircraft 28 volt generating system. It s
14、hall operate from these external voltages irrespective of ground polarity (positive or negative). Additionally, the ventilator apparatus shall be operational from external AC power sources of 110/230 volts, 50-400 Hz, single phase when connected to the separate AC power supply furnished. It shall al
15、so be operational during battery recharge. 2 Provided by IHSNot for ResaleNo reproduction or networking permitted without license from IHS-,-,-0. .i A-A-54444 65 9999974 0062Lb8 L A-A- 5 4 4 4 4 A vehicle power cord assembly shall be provided approximately 6 feet long, having a vehicle cigar lighter
16、 plug on one end and a ventilator plug on the other end. An aircraft DC power cord assembly shall be provided approximately 6 feet long, having one free end (no plug) and a ventilator plug on the other end. Alternately, the vehicle power cord and/or aircraft power cord assemblies may be permanently
17、attached to the ventilator. The ventilator apparatus shall automatically perform a complete functional self-check prior to the commencement of operation. ventilator apparatus to ambient barometric pressure. It shall also provide a means to calibrate the The ventilator shall have a minimum operating
18、temperature range from minus 4OF to plus 115OF, from all power sources cited including the internal rechargeable battery. minimum long-term storage temperature range shall be from 50F to 80F. The The unit shall be provided with a self-contained patient valve. The patient valve shall be reusable, aut
19、oclavable, and transparent. It shall be fabricated from material(s) compatible-for-use in environments of 100 percent oxygen. The patient valve shall include at least: combination 15mm (female)/22mm (male) patient connection port: anti-asphyxiation/demand valve inlet port; exhalation port with remov
20、able PEEP (positive end expiratory pressure) .valve (attached); and pressure transducer adapter. The patient valve shall be configured to allow the ventilator apparatus to detect spontaneous inhalation efforts, and allow the ventilator to trigger at predetermined negative pressure thresholds, in con
21、junction with synchronized assist and synchronized intermittent mandatory ventilation modes of operation. a gas inlet port; The apparatus shall also include a PEEP monitoring circuit, which when activated, enables the user to automatically monitor PEEP performance or manually program the apparatus f
22、or a specific PEEP value. 3 Provided by IHSNot for ResaleNo reproduction or networking permitted without license from IHS-,-,-A-A-54444 65 9999974 00b21b9 3 s A-A- 54 4 4 4 The ventilator shall be provided with connecting tubing (with connectors) for attachment between the output connector of ventil
23、ator apparatus and gas inlet of patient valve; between the transducer input of ventilator apparatus and pressure transducer port of patient valve: and between gas input connector of ventilator apparatus and gas source of 50 psi. The unit shall have an input to accept 100 percent oxygen, medical-grad
24、e air, or blended air/oxygen mixtures from an external blender with a nominal rating of 50 psi. The ventilator shall have a gas-efficient patient circuit and not require purging of exhaled gas accumulations. The ventilator shall have a flow rate variable from O to at least 100 liters per minute (LPM
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
10000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- DODAA544441991VENTILATORVOLUMEPORTABLE 便携式 大量 通风设备 PDF

链接地址:http://www.mydoc123.com/p-701742.html