DOD A-A-54043A-1991 OTOSCOPE AND OPHTHALMOSCOPE SET《耳镜及检眼镜组件》.pdf
《DOD A-A-54043A-1991 OTOSCOPE AND OPHTHALMOSCOPE SET《耳镜及检眼镜组件》.pdf》由会员分享,可在线阅读,更多相关《DOD A-A-54043A-1991 OTOSCOPE AND OPHTHALMOSCOPE SET《耳镜及检眼镜组件》.pdf(7页珍藏版)》请在麦多课文档分享上搜索。
1、A-A-54043A 29 July 1991 Superseding 30 Nov 1989 - - A-A-54043 COMMERCIAL ITEM DESCRIPTION OTOSCOPE AND OPHTHALMOSCOPE SET The General Services Administration has authorized the use of this Commercial Item Description. This Commercial Item Description covers a battery operated otoscope and ophthalmos
2、cope set with carrying case designed and constructed for use in the examination of the internal ear and fundus of the eye and in operating procedures of the ear. This set shall conform to the following salient characteristics: Salient Characteristics: The Set shall consist of the following component
3、s: Otoscope Head with Lamp Ophthalmoscope Head with Lamp Battepy Handle without batteries Three (3) aural specula; large, medium and small One (1) nasal speculum Two (2) replacement lamps for Otoscope Head Two (2) replacement lamps for Ophthalmoscope Head Carrying Case Instruction Manuals All lenses
4、 and filters shall be made of optical grade plastic or glass. OTOSCOPE HEAD (Open-Head construction). GENERAL Open-head style from metal (casting or machining), polished and chrome plated, or painted dull black (non-gloss). Speculum holder may swivel slightly to allow greater accessibility for instr
5、uments into the aural cavity. Attachment of a speculum to the holder shall be by friction or by positive clamping. A replaceable Lamp for illuminating the interior of the distal end of each speculum shall be located on a pedestal along the normal viewing axis at the proximal (large) opening of the s
6、peculum. AMSC N/A FSC 6515 DISTRLBUTION STATEMENT A: Approved for public release; distribution is unlimited. Provided by IHSNot for ResaleNo reproduction or networking permitted without license from IHS-,-,-. i A-A-54043A b5 9999974 0062257 O c .B li c-3 A-A-54043A Attachment to the battery handle s
7、hall be via a machined-metal quick-disconnect bayonet connector. This connector shall be polished and chrome plated. LAMP The Otoscope Lamp shall screw into the pedestal. It shall provide axial illumination at the distal end of the large aural speculum of not less than 210 foot candles when operated
8、 at not more than 2.70 volts. Minimum lamp life shall be 5-hours. The lamp shall be pre-centered, pre-focused, single contact and contain a closed coil filament, SPECULA The Specula shall fit snugly to the Otoscope Head to allcaw easy attachment to or removal from the Otoscope. The aural (3, 4, and
9、5 mm dia.) and nasal (9 mm dia.) specula shall be the reusable type. They shall not become distorted, cracked, or discolored when subjected to boiling and alcohol immersion. The distal ends of the specula shall not have sharp edges. The interior surfaces of the specula shall have a non-reflecting fi
10、nish to prevent glare to the operator. Materials of construction shall be nylon or polypropylene. MAGN I F I ER Each Otoscope Head shall include a magnifier lens which can be rotated into, or out of the line of sight. The magnifier shall be capable of being moved out of the line of vision to permit
11、oblique observation without magnification and insertion of surgical instruments. The magnifying lens shall furnish a magnified view of at least two diopters. The frame shallbe secured to the lamp pedestal on a stud which is parallel to the line of sight between the operator and the pedestal. The fra
12、me shall be retained in a positive manner (journal stud which screws into the pedestal, or stud with lip which is machined or cast into the body). The . magnifier assembly shall hold its position without any adjustments. This may be accomplished with spring-loading. 2 Provided by IHSNot for ResaleNo
13、 reproduction or networking permitted without license from IHS-,-,-Y I A-A-54043A 65 = 9999974 0062258 2 = “t- 4 l-p-p- A-A-54043A OPHTHALMOSCOPE HEAD. ENERAL The Ophthalmoscope Head shall have a mirror reflecting system. It shall not be necessary to manipulate any part of the Head to obtain optimum
14、 illumination, The axis of the filament image shall be fixed and essentially parallel to the upper edge of the mirror. The illuminated spot produced by the Ophthalmoscope shall be uniform and free from distortions. Attachment to the battery handle shall be via a machined-metal quick-disconnect bayon
15、et connector. This connector shall be polished and chrome plated, DIOPTER WHEEL The Ophthalmoscope Head shall have a Diopter Wheel of at least 23 lenses plus one zero aperture, mounted upon a single dial. The minimum lens range shall be from minus 35-diopters to plus 20-diopters. As an alternate, a
16、range of minus 20 to plus 40-diopters is acceptable. Each lens, when it is in the selected posi.tion, shall be identified with its numerical power, through a magnified window with internal light. APERTURE/FILTER WHEEL * The Aperture and Filter system shall contain a minimum of five different positio
17、ns: pinhole, normal aperture, clear slit, red-free filter, and either a grid or color conversion filter. Each aperture and filter shall be capable of being used with any lens selected. HOUS I Na The Ophthalmoscope Hoiising shall be plastic or metal, and shall have a dull (non-gloss) finish. It shall
18、 be designed to assure proper positioning of the Lamp with respect to the optical system. 3 Provided by IHSNot for ResaleNo reproduction or networking permitted without license from IHS-,-,-j A-A-54043A 65 9999974 00b2259 4 I _ A-A-54043A LAMP The Lamp base shall have a radially protruding locator p
19、in for properly positioning the Lamp in the Housing. Axial illumination provided by the Ophthalmoscope Head at a point 7-314 inches from the upper edge of the mirror shall be not less than 10 foot candles when the lamp is operated at not more than 2.70 volts. The Ophthalmoscope Lamp shall be rated a
20、t 2.5 volts at a maximum of 0.66 ampere and have a minimum life of 9-hours. The lamp shall be pre-centered, pre-focused, single contact and provide inert gas i 1 1 umi nat i on. BATTERY HANDLE. GENERAL * The Battery Handle shall mate (mechanically and electrically) with each diagnostic head. It shal
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
10000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- DODAA54043A1991OTOSCOPEANDOPHTHALMOSCOPESET 眼镜 组件 PDF

链接地址:http://www.mydoc123.com/p-701709.html