1、A-A-54040 = 9999974 0342254 397 D A-A-5 4 04 O 19 March 1993 COMMERCIAL ITEM DESCRIPTION REGULATOR, OXYGEN PRESSURE (Lightweight with Integral Flow Controller) The General Services Administration has authorized the use of this Commercial Item Description. This Commercial Item Description covers an o
2、xygen regulator with two 50 psig outlet ports and an integral flow controller. Salient characteristics: The regulator shall be an oxygen pressure regulator with an oxygen flow controller for medical use. It shall be suitable for the operation of two demand valves or one demand valve and an aspirator
3、. It shall consist of: * regulator * flow controller * gauge for supply pressure * yoke for oxygen cylinder valve The unit shall have a total of three external outlet ports, two factory set 50 psig ports with, at least , 100 LPM flow rate and one 2 LPM to 15 LPM flow controller port. When supplied w
4、ith cylinder pressures between 500 and 2000 psig pressure, all three outlet ports shall be usable simultaneously. The 50 psig outlet ports shall maintain 50 psig plus or minus 10 psig and a minimum flow of 100 LPM at each 50 psig port. AMSC N/A FSC 6680 DISTRIBUTION STATEMENT A: Approved for public
5、release; distribution is unlimited. Provided by IHSNot for ResaleNo reproduction or networking permitted without license from IHS-,-,-A-A-5 4040 9999974 OLY2255 223 D A-A-5 4 04 O The envelope size of the regulator shall not exceed 4 inches by 4 inches by 6 inches. The supply oxygen shall be interna
6、lly filtered, stopping all particles larger than 25 microns. In addition to requirements herein, the regulator and components shall be in accordance with CGA E-7. Also, any materials of construction in contact with oxygen shall be resistant to the action of oxygen and these materials shall be in con
7、formance with requirements of UL 252. The regulator shall provide an internal port to supply oxygen to the flow controller. The flow controller outlet port shall be CGA, V-5, No. 1240 barbed end to accept 1/4 inch ID tubing. The flow controller shall have a selectable flow control which is marked to
8、 provide 2, 4, 6, 10, and 15 LPM flows. There shall be zero (O) leakage at the flow controller outlet port when it is set to zero/off position and the regulator is supplied with 2000 psig. The two 50 pig outlet ports shall be DISS, V-5, No. 1240, 9/16 inch male threads. Either single stage or two st
9、age pressure reduction is acceptable and either piston or diaphragm design is acceptable. The regulator shall have a safety relief valve which shall prevent outlet pressure in excess of 100 psig in the event of regulator failure. A safety relief valve which operates at an outlet pressure above the s
10、et outlet pressure and below 100 psig is acceptable. The gauge shall be 1-1/2 inch diameter with a rubber bumper to protect it from shock. The gauge shall indicate t e Kpa). The gauge shall be marked to indicate a full cylinder tank at tank pressure of approximately 1900 psig. In addition to CGA E-7
11、 requirement that gauges meet the requirements of ANSI B40.1, the gauge shall have a minimum accuracy of grade B as defined in ANSI B40.1, shall have a range of 0-4000 PSIG, and shall be in accordance with UL 404. cylinder pressure and be calibrated in psig and Kg/cm i? (or The yoke shall be integra
12、l with the regulator and be pin- indexed for size D and E medical oxygen cylinder valve in accordance with CGA, V-1, No. 870. The regulator, with all components, shall be packaged in a closed plastic bag to prohibit dust collection. 2 Provided by IHSNot for ResaleNo reproduction or networking permit
13、ted without license from IHS-,-,-A-A-54040 m 9999974 0342256 LbT m e e I I I I I I I I I I I I e A-A-54 O 4 O Two owners manuals shall be included providing, at least, information required by CGA E-7. The complete unit shall conform to applicable requirements of a nationally recognized standard for
14、safety. unit shall be approved as conforming to such a standard by an independent nationally recognized testing laboratory (i.e. UL listed or approved by CSA, ETL, or City of Los Angeles). Workmanship. The regulator shall be free from defects which detract from its appearance or impair its serviceab
15、ility. The complete Unit. Each (EA), One regulator, as specified, constitutes one unit. puality Assurance Provisions. Responsibility for inspection. Unless otherwise specified in the contract or purchase order, the contractor is responsible for the performance of all inspection requirements as speci
16、fied herein. Except as otherwise specified in the contract or purchase order, the contractor may use his own or any facilities suitable for the performance of the inspection requirements specified herein, unless disapproved by the Government. The Government reserves the right to perform any of the i
17、nspections set forth in the specification where such inspections are deemed necessary to assure supplies and services conform to prescribed requirements. Records. Records of examinations and tests performed by or for the contractor shall be maintained by the contractor and made available to the Gove
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
10000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- DODAA540401993REGULATOROXYGENPRESSURELIGHTWEIGHTWITHINTEGRALFLOWCONTROLLER 压力 调节器 完整 流量 控制器 PDF

链接地址:http://www.mydoc123.com/p-701708.html