DIN 51007-1994 Thermal analysis differential thermal analysis principles《热分析(TA) 示差热分析(DTA) 原理》.pdf
《DIN 51007-1994 Thermal analysis differential thermal analysis principles《热分析(TA) 示差热分析(DTA) 原理》.pdf》由会员分享,可在线阅读,更多相关《DIN 51007-1994 Thermal analysis differential thermal analysis principles《热分析(TA) 示差热分析(DTA) 原理》.pdf(11页珍藏版)》请在麦多课文档分享上搜索。
1、DEUTSCHE NORM June 1994 I I General principles of differential thermal analysis DIN 51 007 - ICs 71.040.40 Thermische Analyse (TA); Differenzthermoanalyse (DTA); Grundlagen In keepina with current practice in standards published by the International Organization for Standardization (/SO), a comma ha
2、s been used throughout as the decimal marke,: Contents 1 Scope and field of application . . 1 Page 2 Concepts 3 Principles of methods 4 Apparatus . 5 Procedures 6 Calibration 8 Specimen preparation 9 Preparation of equipment . 1 O Evaluation 11 . Test report . Standards and other documents referred
3、to Explanatory notes . 1 Scope and field of application This standard describes several methods for analyzing solid and liquid materials using differential thermal ana- lysis (DTA) or differential scanning calorimetry (DSC). The methods presented in this standard can be used to examine any physical
4、or chemical process involving thermal effects in order to determine transition tempera- tures or heats, reaction temperature ranges, heats of reaction, the purity of eutectic systems (eutectic purity, for short), specific heat capacities and reaction kinetics. Some of these measurements are performe
5、d under con- trolled atmospheres. DTA or DSC can also be used in conjunction with other measurements. For instance, DTA can be carried out in conjunction with thermogravimetric measurements to determine changes in mass as a function of temperature or time. The principles of thermogravimetric analysi
6、s (t.9.a.) are covered by DIN 51 006, the determination of melting temperatures by DIN 51 004. 2 Concepts See DIN 51 005 for general concepts regarding thermal analysis. 3 Principles of methods 3.1 Differential thermal analysis (DTA) and heat flux differential scanning calorimetry (DSC) A test speci
7、men and a reference specimen are sub- jected to a controlled temperature programme (e.9. heat- ing), and the temperature difference between the two Continued on pages 2 to 11. euth Verlag GmbH, Berlin, has the exclusive right of sale for German Standards (DIM-Normen). DIN 51 007 Engl. Price group 9
8、Sales No. O109 08.96 Page 2 DIN 51 007 Low range Middle range High range Extremely high range specimens is measured as a function of temperature or time. With DTA, this difference is used to determine char- acteristic temperatures; with heat flux DSC, the tempera- ture difference is used to determin
9、e calorific values. - 180 Room temperature Room temperature Room temperature 3.2 Power compensation DCC With this technique, the temperature difference between a test specimen and a reference material is determined by subjecting them to a controlled temperature pro- gramme (e.g. heating) while at th
10、e same time equalizing the temperature of both specimens. The energy input re- quired for this process is measured as a function of tem- perature. Characteristic temperatures and calorific values can be obtained by using this method. 4 Apparatus The following shall be used for DTA and heat flux DSC:
11、 an oven with specimen holders and a temperature re- cording system (e.g. differential thermal analyser or dif- ferential scanning calorimeter). The following shall be used for power-compensation DSC: a calorimeter as above, but with separate ovens, each having its own specimen holders, temperature
12、recording system and compensating device, as well as heaters and thermocouples which maintain the temperature pro- gramme and control the power input. DTA and DSC equipment shall also be capable of gener- ating and maintaining a controlled atmosphere when necessary. The temperature signal of calorim
13、eters shall have a noise not exceeding 10 yW r.m.s*) in low temperature ranges; in higher ranges, this level shall not exceed 1 O0 yW r.m.s.*). See table 1 for suggested temperature ranges. Table 1 : Temperature ranges Minimum temperature, in “C Temperature range Maximum temperature, in “C 700 1 O00
14、 1600 1600 and above 4.1 Ovens Ovens shall be capable of maintaining heating or cooling rates to the accuracies given in subclause 4.2. Rates of 1, 2, 5, 10, 20 or 50“C/min are recommended. Experi- ence has shown that high cooling rates can only be achieved in some temperature ranges. The temperatur
15、e field of the oven shall meet the follow- ing requirements: The vertical plane shall extend out from the surface of the specimen holder to a distance at least 1,5 times the holders length; temperatures along this plane shall not deviate from each other more than I5“C. The horizon- tal plane shall e
16、xtend out from the surface of the holder to a distance at least 1,5 times the holders diameter; temperatures along this plane shall be constant to within + 0,5 “C. NOTE: Changes to the temperature gradient dur- ing the temperature programme can lead to drifts in the baseline, poor resolution and/or
17、erroneous results. If a controlled atmosphere is required, ovens shall be ca- pable of maintaining the required atmospheric conditions throughout their entire temperature ranges. If a gas is to be introduced, this gas should have reached specimen temperature by the time it reaches the specimen. Gas
18、samples (e.g. for mass spectrometric or gas chro- matographic analysis) shall be taken in such a manner that the thermal measurement is not affected. 4.2 Temperature control The temperature control device shall be capable of maintaining a constant heating rate between 5 %/min and 50“C/min to an accu
19、racy of 10 %. During isothermal measurements, the temperature shall not deviate more than 0,5 % from the specified isothermal temperature. For temperatures from - 100 “C to +lo0 “C, the tempera- ture shall be accurate to within * 0,5 “C after calibration. The damping which occurs when switching to a
20、n isothermal programme shall not cause an increase in the preset temperature of more than 20 % of the numeri- cal value of the heating rate (in“C/min); the decay time of the damping shall not exceed five minutes. NOTE: During some measurements, the tempera- ture programme is controlled by the specim
21、en in such a way that a constant rate of reaction or of decomposition is obtained. 4.3 Equipment for producing a controlled atmosphere In some cases, controlled atmospheric conditions are re- quired during measurement. For instance, if the speci- men reacts with air, testing shall be carried out und
22、er an inert gas atmosphere. In this case, elements connecting the gas source and the specimen chamber shall be pro- vided with control valves and a flowmeter. For measurements during which the reaction of the specimen with a gas or gas mixture is to be examined, equipment for preparing the gas is re
23、quired. Elements connecting the gas source and the specimen chamber shall be provided with control valves and a flowmeter. If measurements are to be carried out at a pressure above or below atmospheric pressure, the test chamber must be air-tight and equipment for establishing and maintaining the de
24、sired pressure is necessary. Sealed, pressure-resistant holders (e.g. autoclaves or glass am- poules) shall be used when testing at a pressure lower than that above the specimen. The effect of the gas flow rate on thermal measurements should be kept to a minimum. Any gas introduced into the chamber
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
10000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- DIN510071994THERMALANALYSISDIFFERENTIALTHERMALANALYSISPRINCIPLES 分析 TA 差热分析 DTA 原理 PDF

链接地址:http://www.mydoc123.com/p-658360.html