ASTM G207-2011 Standard Test Method for Indoor Transfer of Calibration from Reference to Field Pyranometers《从基准校验场地辐射强度计的传输的标准试验方法》.pdf
《ASTM G207-2011 Standard Test Method for Indoor Transfer of Calibration from Reference to Field Pyranometers《从基准校验场地辐射强度计的传输的标准试验方法》.pdf》由会员分享,可在线阅读,更多相关《ASTM G207-2011 Standard Test Method for Indoor Transfer of Calibration from Reference to Field Pyranometers《从基准校验场地辐射强度计的传输的标准试验方法》.pdf(7页珍藏版)》请在麦多课文档分享上搜索。
1、Designation: G207 11Standard Test Method forIndoor Transfer of Calibration from Reference to FieldPyranometers1This standard is issued under the fixed designation G207; the number immediately following the designation indicates the year oforiginal adoption or, in the case of revision, the year of la
2、st revision. A number in parentheses indicates the year of last reapproval. Asuperscript epsilon () indicates an editorial change since the last revision or reapproval.INTRODUCTIONAccurate and precise measurements of total solar and solar ultraviolet irradiance are required in: (1)the determination
3、of the energy incident on surfaces and specimens during exposure outdoors tovarious climatic factors that characterize a test site, ( 2) the determination of solar irradiance andradiant exposure to ascertain the energy available to solar collection devices such as flat-platecollectors, and (3) the a
4、ssessment of the irradiance and radiant exposure in various wavelength bandsfor meteorological, climatic and earth energy-budget purposes. The solar components of principalinterest include total solar radiant exposure (all wavelengths) and various ultraviolet components ofnatural sunlight that may b
5、e of interest, including both total and narrow-band ultraviolet radiantexposure.This test method for indoor transfer of calibration from reference to field instruments is onlyapplicable to pyranometers and radiometers whose field angles closely approach 180 . instrumentswhich therefore may be said t
6、o measure hemispherical radiation, or all radiation incident on a flatsurface. Hemispherical radiation includes both the direct and sky (diffuse) geometrical components ofsunlight, while global solar irradiance refers only to hemispherical irradiance on a horizontal surfacesuch that the field of vie
7、w includes the entire hemispherical sky dome.For the purposes of this test method, the terms pyranometer and radiometer are used interchangeably.1. Scope1.1 The method described in this standard applies to theindoor transfer of calibration from reference to field radiom-eters to be used for measurin
8、g and monitoring outdoor radiantexposure levels.1.2 This test method is applicable to field radiometersregardless of the radiation receptor employed, but is limited toradiometers having approximately 180 (2p Steradian), fieldangles.1.3 The calibration covered by this test method employs theuse of ar
9、tificial light sources (lamps).1.4 Calibrations of field radiometers are performed withsensors horizontal (at 0 tilt from the horizontal to the earth).The essential requirement is that the reference radiometer shallhave been calibrated at horizontal tilt as employed in thetransfer of calibration.1.5
10、 The primary reference instrument shall not be used as afield instrument and its exposure to sunlight shall be limited tooutdoor calibration or intercomparisons.NOTE 1At a laboratory where calibrations are performed regularly itis advisable to maintain a group of two or three reference radiometers t
11、hatare included in every calibration. These serve as controls to detect anyinstability or irregularity in the standard reference instrument.1.6 Reference standard instruments shall be stored in amanner as to not degrade their calibration.1.7 The method of calibration specified for total solarpyranom
12、eters shall be traceable to the World RadiometricReference (WRR) through the calibration methods of thereference standard instruments (Method G167 and Test MethodE816), and the method of calibration specified for narrow- andbroad-band ultraviolet radiometers shall be traceable to theNational Institu
13、te of Standards and Technology (NIST), orother internationally recognized national standards laboratories(Standard G138).1This test method is under the jurisdiction of ASTM Committee G03 onWeathering and Durability and is the direct responsibility of Subcommittee G03.09on Radiometry.Current edition
14、approved July 1, 2011. Published August 2011. DOI: 10.1520/G0207111Copyright ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959, United States.1.8 This standard does not purport to address all of thesafety concerns, if any, associated with its use. It is theresp
15、onsibility of the user of this standard to establish appro-priate safety and health practices and determine the applica-bility of regulatory limitations prior to use.2. Referenced Documents2.1 ASTM Standards:2E772 Terminology Relating to Solar Energy ConversionE816 Test Method for Calibration of Pyr
16、heliometers byComparison to Reference PyrheliometersE824 Test Method for Transfer of Calibration From Refer-ence to Field RadiometersG113 Terminology Relating to Natural andArtificial Weath-ering Tests of Nonmetallic MaterialsG138 Test Method for Calibration of a SpectroradiometerUsing a Standard So
17、urce of IrradianceG167 Test Method for Calibration of a Pyranometer Usinga Pyrheliometer2.2 Other Standards:3ISO 9847 Solar Energy Calibration of Field Pyranometersby Comparison to a Reference Pyranometer3. Terminology3.1 Definitions:3.1.1 See Terminology E772 and G113 for terminologyrelating to thi
18、s test method.4. Summary of Test Method4.1 Mount the reference pyranometer, and the field (or test)radiometers, or pyranometers, on a common calibration tablefor horizontal calibration. Adjust the height of the radiationreceptor of all instruments to a common elevation.4.2 Connect the signal cables
19、from the reference and testsensors to a data acquisition system.4.3 Adjust the data acquisition system to record data at theselected data collection interval.NOTE 2Data collection interval should be function of the timeconstant of the sensor. Sensor time constant is the period of time requiredfor a
20、sensor to reach 1 1/e = 63% of the maximum minus the minimumamplitude of a step change in input stimulus. (e is base of naturallogarithms, 2.718282.). Often, “one over e” (1/e) time constants arereported for radiation sensors, for example “1/e response time = 3seconds”. This represents the time for
21、the sensor signal to reach 37% ofthe full range step change representing the step change in the stimulus.Four times the 1/e time constant can be considered the time for the sensorto fully respond to a step change in stimulus.4.4 Energize the source to be used for the transfer ofcalibration.NOTE 3It
22、is mandatory that the spectral distribution of the source beknown or well characterized. Indoor calibration transfers between narrowband radiometers such as Ultraviolet and Photopic detectors shall beaccomplished using sources with spectral irradiance distributions assimilar as possible to the spect
23、ral distribution of the sources to bemonitored. This will reduce spectral mismatch errors arising fromdifferences in the spectral response of sensors and dissimilar calibrationand test source spectral distributions. In the special case of pyranometersfor solar radiation measurements, as long as the
24、reference radiometer hasa relatively flat and broad (greater than 700 nm passband) spectralresponse (for example, black thermopile), or has been calibrated outdoors,the difference between calibration and source spectral distributions is lessimportant, however should be taken into consideration.4.5 M
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
5000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ASTMG2072011STANDARDTESTMETHODFORINDOORTRANSFEROFCALIBRATIONFROMREFERENCETOFIELDPYRANOMETERS 基准 校验 场地

链接地址:http://www.mydoc123.com/p-540551.html