《ASTM G175-2003 Standard Test Method for Evaluating the Ignition Sensitivity and Fault Tolerance of Oxygen Regulators Used for Medical and Emergency Applications《医疗和应急设备用氧气调节器的引燃敏感性.pdf》由会员分享,可在线阅读,更多相关《ASTM G175-2003 Standard Test Method for Evaluating the Ignition Sensitivity and Fault Tolerance of Oxygen Regulators Used for Medical and Emergency Applications《医疗和应急设备用氧气调节器的引燃敏感性.pdf(14页珍藏版)》请在麦多课文档分享上搜索。
1、Designation: G 175 03Standard Test Method forEvaluating the Ignition Sensitivity and Fault Tolerance ofOxygen Regulators Used for Medical and EmergencyApplications1This standard is issued under the fixed designation G 175; the number immediately following the designation indicates the year oforigina
2、l adoption or, in the case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. Asuperscript epsilon (e) indicates an editorial change since the last revision or reapproval.1. Scope1.1 This standard describes a test method for evaluating theignition
3、sensitivity and fault tolerance of oxygen regulatorsused for medical and emergency applications.1.2 For the purpose of this standard, a pressure regulator isa device, also called a pressure-reducing valve, that is intendedfor medical or emergency purposes and that is used to converta medical or emer
4、gency gas pressure from a high, variablepressure to a lower, more constant working pressure 21 CFR868.2700 (a).1.3 This standard applies only to oxygen regulators used formedical and emergency applications that are designed andfitted with CGA 540 inlet connections or CGA 870 pin-indexadapters (CGA V
5、-1).1.4 This standard provides an evaluation tool for determin-ing the fault tolerance of oxygen regulators used for medicaland emergency applications. A fault tolerant regulator isdefined as (1) having a low probability of ignition as evaluatedby rapid pressurization testing, and (2) having a low c
6、onse-quence of ignition as evaluated by forced ignition testing.1.5 This standard is not a design standard; however, it canbe used to aid designers in designing and evaluating the safeperformance and fault tolerance capability of oxygen regula-tors used for medical and emergency applications (G 128)
7、.NOTE 1It is essential that a risk assessment be carried out onbreathing gas systems, especially concerning oxygen compatibility (referto ASTM G 63 and G 94) and toxic product formation due to ignition ordecomposition of nonmetallic materials as weighed against the risk offlammability (refer to ISO
8、15001.2). See Appendix X1 and X2.1 fordetails.1.6 This standard is also used to aid those responsible forpurchasing or using oxygen regulators used for medical andemergency applications in ensuring that selected regulators aretolerant of the ignition mechanisms that are normally active inoxygen syst
9、ems.1.7 This standard does not purport to address the ignitionsensitivity and fault tolerance of an oxygen regulator caused bycontamination during field maintenance or use. Regulatordesigners and manufacturers should provide design safeguardsto minimize the potential for contamination or its consequ
10、ences(G 88).1.8 This standard does not purport to address all of thesafety concerns, if any, associated with its use. It is theresponsibility of the user of this standard to establish appro-priate safety and health practices and determine the applica-bility of regulatory limitations prior to use.2.
11、Referenced Documents2.1 ASTM Standards:G 63 Guide for Evaluating Nonmetallic Materials for Oxy-gen Service2G 88 Guide for Designing Systems for Oxygen Service2G 93 Practice for Cleaning Methods and Cleanliness Levelsfor Material and Equipment Used in Oxygen-EnrichedEnvironments2G 94 Guide for Evalua
12、ting Metals for Oxygen Service2G 128 Guide for Control of Hazards and Risks in OxygenEnriched Systems22.2 ASTM Adjuncts:Manual 36, Safe Use of Oxygen and Oxygen Systems32.3 Compressed Gas Association (CGA) Standards:CGA E-4, Standard for Gas Pressure Regulators4CGA G-4, Oxygen4CGA G-4.1, Cleaning Eq
13、uipment for Oxygen Service4CGA V-1, American National/Compressed Gas AssociationStandard for Compressed Gas Cylinder Valve Outlet andInlet Connections42.4 United States Pharmacopeial Convention Standard:1This test method is under the jurisdiction of ASTM Committee G04 onCompatibility and Sensitivity
14、 of Materials in Oxygen Enriched Atmospheres and isthe direct responsibility of Subcommittee G04.01 on Test Methods.Current edition approved April 10, 2003. Published May 2003. Originallypublished as PS 127 00. Last previous edition PS 127 00.2Annual Book of ASTM Standards, Vol 14.04.3Available from
15、 American Society for Testing and Materials, 100 Barr HarborDr., West Conshohocken, PA 194282959.4Available from Compressed Gas Association (CGA), 1725 Jefferson DavisHwy., Suite 1004, Arlington, VA 22202-4102.1Copyright ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 1
16、9428-2959, United States.USP 24 NF 19, Oxygen monograph52.5 Federal Regulation:21 CFR 868.2700 (a), Pressure regulator62.6 ISO Standards:ISO 10524 Pressure regulators and pressure regulators withflow-metering devices for medical gas systems7ISO 15001 Anaesthetic and respiratory equipment Com-patibil
17、ity with oxygen73. Summary of Test Method3.1 This test method comprises two phases. A regulatormust pass both phases in order to be considered ignitionresistant and fault tolerant.3.2 Phase 1: Oxygen Pressure Shock TestIn this testphase, fault tolerance is evaluated by testing the ignitionresistance
18、 of the regulator design by subjecting the regulator toheat from oxygen pressure shocks. The test is performedaccording to ISO 10524, Section 11.8.1, which is similar toCGA E-4.3.3 Phase 2: Regulator Inlet Promoted Ignition TestInthis test phase, fault tolerance is evaluated by subjecting theregulat
19、or to the forced application of a positive ignition sourceat the regulator inlet to simulate cylinder valve seat ignitionand particle impact events. The ignition source is representa-tive of severe, but realistic, service conditions. The Phase 1component test system is used for Phase 2 to pressure s
20、hock aregulator upstream of its inlet so that an ignition pill is kindledto initiate combustion within the regulator.4. Significance and Use4.1 This test method comprises two phases and is used toevaluate the ignition sensitivity and fault tolerance of oxygenregulators used for medical and emergency
21、 applications.4.2 Phase 1: Oxygen Pressure Shock TestThe objective ofthis test phase is to determine whether the heat from oxygenpressure shocks will result in burnout or visible heat damage tothe internal parts of the regulator. Phase 1 is performedaccording to ISO 10524, Section 11.8.1.4.2.1 The c
22、riteria for an acceptable test are specified in ISO10524, Section 11.8.1.4.2.2 The pass/fail criteria for a regulator are specified inISO 10524, Section 11.8.1.4.3 Phase 2: Regulator Inlet Promoted Ignition TestTheobjective of this test phase is to determine if an ignition eventupstream of the regul
23、ator inlet filter will result in sustainedcombustion and burnout of the regulator.4.3.1 The criterion for an acceptable test is either, (1) failureof the regulator, which is defined as the breach of thepressurized regulator component (burnout) and ejection ofmolten or burning metal or any internal p
24、arts from theregulator, or (2) if the regulator does not fail, consumption ofat least 90 % of the ignition pill as determined by visualinspection or mass determination. Failure of the regulator at theseal ring does not constitute an acceptable test.4.3.2 Momentary (less than 1 s) ejection of flame t
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
5000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ASTMG1752003STANDARDTESTMETHODFOREVALUATINGTHEIGNITIONSENSITIVITYANDFAULTTOLERANCEOFOXYGENREGULATORSUSEDFORMEDICALANDEMERGENCYAPPLICATIONS

链接地址:http://www.mydoc123.com/p-540463.html