ASTM G130-2006 Standard Test Method for Calibration of Narrow- and Broad-Band Ultraviolet Radiometers Using a Spectroradiometer《使用光谱辐射计校准窄带和宽带紫外辐射计的标准试验方法》.pdf
《ASTM G130-2006 Standard Test Method for Calibration of Narrow- and Broad-Band Ultraviolet Radiometers Using a Spectroradiometer《使用光谱辐射计校准窄带和宽带紫外辐射计的标准试验方法》.pdf》由会员分享,可在线阅读,更多相关《ASTM G130-2006 Standard Test Method for Calibration of Narrow- and Broad-Band Ultraviolet Radiometers Using a Spectroradiometer《使用光谱辐射计校准窄带和宽带紫外辐射计的标准试验方法》.pdf(6页珍藏版)》请在麦多课文档分享上搜索。
1、Designation: G 130 06Standard Test Method forCalibration of Narrow- and Broad-Band UltravioletRadiometers Using a Spectroradiometer1This standard is issued under the fixed designation G 130; the number immediately following the designation indicates the year oforiginal adoption or, in the case of re
2、vision, the year of last revision. A number in parentheses indicates the year of last reapproval. Asuperscript epsilon (e) indicates an editorial change since the last revision or reapproval.INTRODUCTIONAccurate and precise measurements of ultraviolet irradiance are required in the determination of
3、theradiant exposure of both total and selected narrow bands of ultraviolet radiation for the determinationof exposure levels in (1) outdoor weathering of materials, (2) indoor accelerated exposure testing ofmaterials using manufactured light sources, and (3) UV-Aand UV-B ultraviolet radiation in ter
4、ms bothof the assessment of climatic parameters and the changes that may be taking place in the solarultraviolet radiation reaching earth.Although meteorological measurements usually require calibration of pyranometers and radiom-eters oriented with axis vertical, applications associated with materi
5、als testing require an assessmentof the calibration accuracy at orientations with the axis horizontal (usually associated with testing inindoor exposure cabinets) or with the axis at angles typically up to 45 or greater from the horizontal(for outdoor exposure testing). These calibrations also requi
6、re that deviations from the cosine law, tilteffects, and temperature sensitivity be either known and documented for the instrument model ordetermined on individual instruments.This test method requires calibrations traceable to primary reference standards maintained by anational metrological laborat
7、ory that has participated in intercomparisons of standards of spectralirradiance.1. Scope1.1 This test method covers the calibration of ultravioletlight-measuring radiometers possessing either narrow- orbroad-band spectral response distributions using either a scan-ning or a linear-diode-array spect
8、roradiometer as the primaryreference instrument. For transfer of calibration from radiom-eters calibrated by this test method to other instruments, TestMethod E 824 should be used.NOTE 1Special precautions must be taken when a diode-array spec-troradiometer is employed in the calibration of filter r
9、adiometers havingspectral response distributions below 320-nm wavelength. Such precau-tions are described in detail in subsequent sections of this test method.1.2 This test method is limited to calibrations of radiometersagainst light sources that the radiometers will be used tomeasure during field
10、use.NOTE 2For example, an ultraviolet radiometer calibrated againstnatural sunlight cannot be employed to measure the total ultravioletirradiance of a fluorescent ultraviolet lamp.1.3 Calibrations performed using this test method may beagainst natural sunlight, Xenon-arc burners, metal halideburners
11、, tungsten and tungsten-halogen lamps, fluorescentlamps, etc.1.4 Radiometers that may be calibrated by this test methodinclude narrow-, broad-, and wide-band ultraviolet radiom-eters, and narrow-, broad, and wide-band visible-region-onlyradiometers, or radiometers having wavelength response dis-trib
12、utions that fall into both the ultraviolet and visible regions.NOTE 3For purposes of this test method, narrow-band radiometers arethose with Dl # 20 nm, broad-band radiometers are those with 20 nm#Dl # 70 nm, and wide-band radiometers are those with Dl $ 70 nm.1This test method is under the jurisdic
13、tion of ASTM Committee G3 onDurability of Nonmetallic Materials and is the direct responsibility of SubcommitteeG03.09 on Solar and Ultraviolet Radiation Measurement Standards.Current edition approved June 1, 2006. Published July 2006. Originally approvedin 1995. Last previous edition approved in 20
14、02 as G 13095(2002)1Copyright ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959, United States.NOTE 4For purposes of this test method, the ultraviolet region isdefined as the region from 285 to 400-nm wavelength, and the visibleregion is defined as the region f
15、rom 400 to 750-nm wavelength. Theultraviolet region is further defined as being either UV-A with radiation ofwavelengths from 315 to 400 nm, or UV-B with radiation from 285 to315-nm wavelength.1.5 This standard does not purport to address all of thesafety concerns, if any, associated with its use. I
16、t is theresponsibility of the user of this standard to establish appro-priate safety and health practices and determine the applica-bility of regulatory limitations prior to use.2. Referenced Documents2.1 ASTM Standards:2E 772 Terminology Relating to Solar Energy ConversionE 824 Test Method for Tran
17、sfer of Calibration From Refer-ence to Field Radiometers3. Terminology3.1 Definitions:3.1.1 broad-band radiometera relative term generallyapplied to radiometers with interference filters or cut-on/cut-offfilter pairs having a FWHM between 20 and 70 nm and withtolerances in center (peak) wavelength a
18、nd FWHM no greaterthan 62 nm.3.1.2 diode array detectora detector with from 50 to 1000silicon photodiodes affixed side-by-side in a linear array andmounted in the focal plane of the exit slit of a monochromator.3.1.3 full width at half maximum (FWHM)in a bandpassfilter, FWHM is the interval between
19、wavelengths at whichtransmittance is 50 % of the peak, frequently referred to asbandwidth.3.1.4 narrow-band radiometera relative term generallyapplied to radiometers with interference filters with FWHM#20 nm and with tolerances in center (peak) wavelength andFWHM no greater than6 2 nm.3.1.5 scanning
20、 monochromatora monochromator thatuses either a single, or several interchangeable, detector(s)mounted at the exit slit, that is presented with dispersed lightby sweeping the spectrum across the slit to illuminate thedetector with a succession of different very narrow wavelengthlight distributions.
21、The detector may be either a photomulti-plier tube (PMT) or silicon photodiode (visible), or a PMT oran ultraviolet-enhanced silicon photodiode (ultraviolet andvisible), or a lead sulfide cell or other solid state detector (nearinfrared), etc. The dispersed spectrum is swept across themonochromators
22、 exit slit using a mechanical stage that rotateseither a prism or a grating dispersive element, usually under thecontrol of an external microprocessor or computer.3.1.6 spectroradiometera radiometer consisting of amonochromator with special acceptance optics mounted to theentrance aperture and a det
23、ector mounted to the exit aperture,usually provided with electronic or computer encoding ofwavelength and radiometric intensity. The monochromator ofsuch instruments is either of the linear diode (often called diodearray) or the scanning type.3.1.7 wide-band radiometera relative term generally ap-pl
24、ied to radiometers with combinations of cut-off and cut-onfilters with FWHM greater than 70 nm.3.2 For other terms relating to this test method, see Termi-nology E 772.4. Significance and Use4.1 This test method represents the preferable means forcalibrating both narrow-band and broad-band ultraviol
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
5000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ASTMG1302006STANDARDTESTMETHODFORCALIBRATIONOFNARROWANDBROADBANDULTRAVIOLETRADIOMETERSUSINGASPECTRORADIOMETER

链接地址:http://www.mydoc123.com/p-540333.html