ASTM F3014-2014 Standard Test Method for Penetration Testing of Needles Used in Surgical Sutures《外科缝合针穿透试验的标准试验方法》.pdf
《ASTM F3014-2014 Standard Test Method for Penetration Testing of Needles Used in Surgical Sutures《外科缝合针穿透试验的标准试验方法》.pdf》由会员分享,可在线阅读,更多相关《ASTM F3014-2014 Standard Test Method for Penetration Testing of Needles Used in Surgical Sutures《外科缝合针穿透试验的标准试验方法》.pdf(4页珍藏版)》请在麦多课文档分享上搜索。
1、Designation: F3014 14Standard Test Method forPenetration Testing of Needles Used in Surgical Sutures1This standard is issued under the fixed designation F3014; the number immediately following the designation indicates the year oforiginal adoption or, in the case of revision, the year of last revisi
2、on. A number in parentheses indicates the year of last reapproval. Asuperscript epsilon () indicates an editorial change since the last revision or reapproval.INTRODUCTIONThe purpose of this test method is to provide a common method for fixturing and measuring thepenetration force of a surgical need
3、le. Most surgical needles are coated to facilitate passage throughtissue. This coating needs to be both lubricious and durable. The lubricity of the coating acts to lowerthe penetration force, while the durability of the coating maintains the needles ability to pass throughtissue multiple times with
4、 minimum increase in penetration force. It has been observed that, when acoating is found to be more lubricious, that coating is often less durable. Likewise, when a coating isfound to be more durable, that coating is often less lubricious. Comparative measurements of lubricityand durability can pro
5、vide an expectation of how the coated needle may perform during actual use.This standard does not presently address the testing medium. Test medium will be addressed in thenext phase of this test method development. However, this test method does require certain mediumcharacteristics to be identifie
6、d.1. Scope1.1 This test method describes the procedure for penetrationtesting sharp and blunt needles used for the placement ofsurgical sutures. It only describes the test procedures (Phase 1)and does not define a test medium at this time. Test media willbe address in the next phase (Phase 2) of thi
7、s test methodsevolution.1.2 This test method applies to straight and curved surgicalneedles.1.3 This test method does not intend to address the value ofthe test with respect to actual conditions of use.1.4 In this test method, it is assumed the needles used forthis test have passed all applicable qu
8、ality standards and haveno physical malformation that would inappropriately influencethe test results.1.5 This test method quantifies the force required for aneedle to pass through a medium and its ability to do somultiple times.1.6 UnitsThe values stated in either SI units or inch-pound units are t
9、o be regarded separately as standard. Thevalues stated in each system may not be exact equivalents;therefore, each system shall be used independently of the other.Combining values from the two systems may result in noncon-formance with the standard.1.7 This standard does not purport to address all o
10、f thesafety concerns, if any, associated with its use. It is theresponsibility of the user of this standard to establish appro-priate safety and health practices and determine the applica-bility of regulatory limitations prior to use.2. Referenced Documents2.1 ASTM Standards:2E122 Practice for Calcu
11、lating Sample Size to Estimate, WithSpecified Precision, the Average for a Characteristic of aLot or ProcessF1840 Terminology for Surgical Suture Needles3. Terminology3.1 DefinitionsTerms used in this test method are inaccordance with Terminology F1840.3.2 Definitions of Terms Specific to This Stand
12、ard:3.2.1 compound curved surgical needle, nneedle havingmore than one radius of curvature associated with the shape ofthe needle.3.2.2 needle test gripping location, narea in front of theattachment zone or tangent portion of the needle where theneedle will be clamped/secured in the test fixture.1Th
13、is test method is under the jurisdiction of ASTM Committee F04 on Medicaland Surgical Materials and Devices and is the direct responsibility of SubcommitteeF04.33 on Medical/Surgical Instruments.Current edition approved March 1, 2014. Published April 2014. DOI: 10.1520/F3014-14.2For referenced ASTM
14、standards, visit the ASTM website, www.astm.org, orcontact ASTM Customer Service at serviceastm.org. For Annual Book of ASTMStandards volume information, refer to the standards Document Summary page onthe ASTM website.Copyright ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocke
15、n, PA 19428-2959. United States13.2.2.1 Discussion3.2.2 can be exempt for exceptionallysmall needles.4. Summary of Test Method4.1 Securely clamp the needle into the clamping fixture in auniform area in front of the attachment zone or tangent portionof the needle, distal to the needle point. This is
16、referred to asthe needle test gripping location, such that the needle move-ment is restricted.4.1.1 The needle shall be held so that the needle passesthrough the medium perpendicular to the test medium.4.2 The clamping fixture is rotated forward to commencepassing the needle through the media accord
17、ing to Table 1 orTable 2, as applicable.4.3 As the needle penetrates and passes through the mediumat 90, data about the force needed to penetrate the medium iscontinuously collected.5. Significance and Use5.1 This test method provides a means of assessing thepenetration force of a surgical needle th
18、rough the chosenmedia.5.2 The needle is passed through a chosen medium at achosen speed and path which applies a force on the load cellthat reflects the needles resistance to penetration as a functionof its shape and coating.6. ApparatusNOTE 1Two methods of testing are possible for curved surgicalne
19、edlesrotational, based on the rotational speed of the sample, andlinear, based on the linear speed of the sample. Either method may beused. The method selected shall be indicated in any use of the datagenerated. The data generated by these methods cannot be compared toeach other. Only like method co
20、mparisons shall be made.6.1 Clamping FixtureA device to clamp the surgicalneedle firmly.6.2 Data Collection SystemThe data shall be collected asa function of force.6.3 Rotational Speed for a Curved Surgical NeedleTheneedle-clamping fixture shall be rotated at a continuous andconstant speed of 4 6 2/
21、s.6.4 Linear Speed for a Straight Surgical NeedleTheneedle-clamping fixture shall move at a continuous and con-stant speed. The minimum speed shall be 10.0 cm/min (4in./min). The maximum speed shall be 30.5 cm/min (12in./min) unless a greater test speed is required as a function ofuse.6.5 Linear Spe
22、ed for a Curved Surgical NeedleTheneedle-clamping fixture shall move at a continuous and con-stant speed. The minimum speed shall be 10.0 cm/min (4in./min). The maximum speed shall be 30.5 cm/min (12in./min).6.5.1 Speed can be set to address a special use of the needle.7. Sampling7.1 There should be
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
5000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ASTMF30142014STANDARDTESTMETHODFORPENETRATIONTESTINGOFNEEDLESUSEDINSURGICALSUTURES 外科 缝合 穿透 试验 标准 方法

链接地址:http://www.mydoc123.com/p-539901.html