ASTM F2996-2013 Standard Practice for Finite Element Analysis (FEA) of Non-Modular Metallic Orthopaedic Hip Femoral Stems《对非模块化金属骨科髋部股骨柄进行有限元分析(FEA)的标准实施规程》.pdf
《ASTM F2996-2013 Standard Practice for Finite Element Analysis (FEA) of Non-Modular Metallic Orthopaedic Hip Femoral Stems《对非模块化金属骨科髋部股骨柄进行有限元分析(FEA)的标准实施规程》.pdf》由会员分享,可在线阅读,更多相关《ASTM F2996-2013 Standard Practice for Finite Element Analysis (FEA) of Non-Modular Metallic Orthopaedic Hip Femoral Stems《对非模块化金属骨科髋部股骨柄进行有限元分析(FEA)的标准实施规程》.pdf(11页珍藏版)》请在麦多课文档分享上搜索。
1、Designation: F2996 13Standard Practice forFinite Element Analysis (FEA) of Non-Modular MetallicOrthopaedic Hip Femoral Stems1This standard is issued under the fixed designation F2996; the number immediately following the designation indicates the year oforiginal adoption or, in the case of revision,
2、 the year of last revision. A number in parentheses indicates the year of last reapproval. Asuperscript epsilon () indicates an editorial change since the last revision or reapproval.1. Scope1.1 This practice establishes requirements and consider-ations for the numerical simulation of non-modular (t
3、hat is,limited to monolithic stems with only a femoral head / trunniontaper interface) metallic orthopaedic hip stems using FiniteElement Analysis (FEA) techniques for the estimation ofstresses and strains. This standard is only applicable to stressesbelow the yield strength, as provided in the mate
4、rial certifica-tion.1.2 PurposeThis practice establishes requirements andconsiderations for the development of finite element models tobe used in the evaluation of non-modular metallic orthopaedichip stem designs for the purpose of prediction of the staticimplant stresses and strains. This procedure
5、 can be used forworst case assessment within a family of implant sizes toprovide efficiencies in the amount of physical testing to beconducted. Recommended procedures for performing modelchecks and verification are provided to help determine if theanalysis follows recommended guidelines. Finally, th
6、e recom-mended content of an engineering report covering the mechani-cal simulation is presented.1.3 LimitsThis practice is limited in discussion to thestatic structural analysis of non-modular metallic orthopaedichip stems (which excludes the prediction of fatigue strength).1.4 The values stated in
7、 SI units are to be regarded asstandard. No other units of measurement are included in thisstandard.1.5 This standard does not purport to address all of thesafety concerns, if any, associated with its use. It is theresponsibility of the user of this standard to establish appro-priate safety and heal
8、th practices and determine the applica-bility of regulatory limitations prior to use.2. Referenced Documents2.1 ISO Standards:2ISO 7206-4 (2010) Implants for SurgeryPartial and TotalHip Joint ProsthesesPart 4: Determination of EnduranceProperties and Performance of Stemmed Femoral Com-ponents3. Sign
9、ificance and Use3.1 This practice is applicable to the calculation of stressesseen on a femoral hip stem when loaded in a manner describedin ISO 7206-4 (2010). This method can be used to establish theworst case size for a particular implant. When stresses calcu-lated using this practice were compare
10、d to the stresses mea-sured from physical strain gauging techniques performed attwo laboratories using two different methods, the resultscorrelated to within 8 %.4. Geometric Data4.1 Finite element models are based on a geometric repre-sentation of the device being studied. The source of thegeometri
11、c details can be obtained from drawings, solid models,preliminary sketches, or any other source consistent withdefining the model geometry. In building the finite elementmodel, certain geometric details may be omitted from theorthopaedic implant geometry shown in the CAD model if it isdetermined tha
12、t they are not relevant to the intended analysis.Engineering judgment shall be exercised to establish the extentof model simplification and shall be justified.4.2 It is most appropriate to consider the “worst case” stresscondition for the orthopaedic implant being simulated. The“worst case” shall be
13、 determined from all relevant engineeringconsiderations, such as stem geometry, dimensions, and headoffset. If finite element analysis is being used for determiningthe worst case, then the worst case head offset may not beknown. It may be necessary to run several variants of headoffset, in order to
14、determine this.1This practice is under the jurisdiction ofASTM Committee F04 on Medical andSurgical Materials and Devices and is the direct responsibility of SubcommitteeF04.22 on Arthroplasty.Current edition approved July 15, 2013. Published August 2013. DOI: 10.1520/F2996-13.2Available from Intern
15、ational Organization for Standardization (ISO), 1, ch. dela Voie-Creuse, CP 56, CH-1211 Geneva 20, Switzerland, http:/www.iso.org.Copyright ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959. United States15. Material Properties5.1 The required material properti
16、es for input into an FEAmodel for the calculation of strains and displacement aremodulus of elasticity (E) and Poissons ratio (). These valuescan be obtained from material certification data. It should benoted that as ISO 7206-4 (2010) is run under load control, theFEA should also be run under load
17、control. When the FEA isrun under load control, the modulus of elasticity will not affectthe stress calculations under small displacement theory but willaffect displacement and strain. The influence of Poissons ratioon the stress calculations is negligible.5.2 Ensure that material property units are
18、 consistent withgeometric units in the CAD model. SI units are the preferredunits of measure.6. Loading6.1 The loading and orientation of the hip stem shall beguided by the ISO 7206-4 (2010) standard. The areas ofparticular interest are the stresses in the neck region, driverhole region, potting lev
19、el, and other design-specific criticalregions.6.2 The load shall be applied such that the magnitude anddirection are identical to that defined in ISO 7206-4 (2010).The point of load application shall produce a statically equiva-lent bending moment to a load applied through the head centerwith its wo
20、rst case head offset.6.2.1 The load in the model will be applied to the endcircular face of the hip stem trunnion or in a justifiablyequivalent manner. The trunnion may be extended or truncatedto approximate the loading conditions that simulate the worstcase head offset, which may be determined via
21、an iterativeprocess. This approximation should be reported if performed.Alternatively, a rigid couple can be used to tie the load point tothe trunnion end circular face. Refer to Fig. 1.6.2.2 It is recognized that the loading conditions in thispractice are not identical to that of ISO 7206-4 (2010).
22、However, the difference in loading conditions (for example,load applied to surface of head versus face of stem trunnion;potting level differences; use of bone cement which is notmodeled in FEA) does not significantly affect identification ofthe “worst case” stress condition and construct for subsequ
23、entbench testing, which is the primary objective of this practice.6.3 Ensure that load units are consistent with materialproperty units.7. Boundary Conditions7.1 The hip stem will first be cut at a distance from thecenter of the head as described in ISO 7206-4 (2010) with theworst case head/neck off
24、set. This cut represents the level towhich stresses and strains shall be evaluated.Asecond cut shallthen be made 10 mm below the first cut. The hip stem shall beconstrained in all directions on all faces distal to the secondcut. Constraining the stem in this manner ensures that exces-sive erroneous
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
5000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ASTMF29962013STANDARDPRACTICEFORFINITEELEMENTANALYSISFEAOFNONMODULARMETALLICORTHOPAEDICHIPFEMORALSTEMS

链接地址:http://www.mydoc123.com/p-539874.html