ASTM F2790-2010 Standard Practice for Static and Dynamic Characterization of Motion Preserving Lumbar Total Facet Prostheses《活动腰椎全椎间盘假体的静态和动态特性用标准实施规程》.pdf
《ASTM F2790-2010 Standard Practice for Static and Dynamic Characterization of Motion Preserving Lumbar Total Facet Prostheses《活动腰椎全椎间盘假体的静态和动态特性用标准实施规程》.pdf》由会员分享,可在线阅读,更多相关《ASTM F2790-2010 Standard Practice for Static and Dynamic Characterization of Motion Preserving Lumbar Total Facet Prostheses《活动腰椎全椎间盘假体的静态和动态特性用标准实施规程》.pdf(9页珍藏版)》请在麦多课文档分享上搜索。
1、Designation: F2790 10Standard Practice forStatic and Dynamic Characterization of Motion PreservingLumbar Total Facet Prostheses1This standard is issued under the fixed designation F2790; the number immediately following the designation indicates the year oforiginal adoption or, in the case of revisi
2、on, the year of last revision. A number in parentheses indicates the year of last reapproval. Asuperscript epsilon () indicates an editorial change since the last revision or reapproval.1. Scope1.1 This practice provides guidance for the static anddynamic testing of Lumbar Total Facet Prostheses (FP
3、). Theseimplants are intended to allow motion and lend support to oneor more functional spinal unit(s) through replacement of thenatural facets.1.2 These test methods are intended to provide a basis forthe mechanical comparison among past, present, and futurenon-biologic FP. These test methods allow
4、 comparison ofdevices with different methods of application to the lumbarspine. These test methods are intended to enable the user tomechanically compare devices and do not purport to provideperformance standards for them.1.3 These test methods describe static and dynamic tests byspecifying load typ
5、es and specific methods of applying theseloads.1.4 These test methods do not purport to address all clini-cally relevant failure modes for FP, some of which will bedevice specific. For example, these test methods do not addressimplant wear resistance under expected in vivo loads andmotions. In addit
6、ion, the biologic response to wear debris is notaddressed in these test methods.1.5 Requirements are established for measuring displace-ments and evaluating the stiffness of FP.1.6 Some devices may not be testable in all test configura-tions.1.7 The values stated in SI units are to be regarded as th
7、estandard with the exception of angular measurements, whichmay be reported in terms of either degrees or radians.1.8 The values stated in inch-pound units are to be regardedas standard. The values given in parentheses are mathematicalconversions to SI units that are provided for information onlyand
8、are not considered standard.2. Referenced Documents2.1 ASTM Standards:2D638 Test Method for Tensile Properties of PlasticsE4 Practices for Force Verification of Testing MachinesE6 Terminology Relating to Methods of Mechanical TestingE468 Practice for Presentation of Constant Amplitude Fa-tigue Test
9、Results for Metallic MaterialsE739 Practice for Statistical Analysis of Linear or Linear-ized Stress-Life ( S-N) and Strain-Life (e-N) Fatigue DataF1582 Terminology Relating to Spinal Implants3. Terminology3.1 DefinitionsAll functional and kinematic testing termi-nology is consistent with the refere
10、nced standards (includingTeminology E6 and Terminology F1582), unless otherwisestated.3.1.1 coordinate systems/axesGlobal XYZ orthogonalaxes are defined following a right-handed Cartesian coordinatesystem in which the XY plane is parallel to and co-planar withthe superior endplate of the inferior ve
11、rtebral body. Alternativecoordinate systems may be used with justification. The globalaxes are fixed relative to the inferior vertebral body. Lowercase letters, xyz, denote a local moving orthogonal coordinatesystem attached to the superior vertebral body with directionsinitially coincident with tho
12、se of the global XYZ axes, respec-tively. The 3D motion of the superior relative to inferiorvertebra is specified and is to be measured in terms ofsequential Eulerian angular rotations about the xyz axes,respectively (z axial rotation, x lateral bend, and y flexion-extension).3.1.1.1 origincenter of
13、 the global coordinate system thatis located at the posterior medial position on the superiorendplate of the inferior vertebral body.3.1.1.2 X-axispositive X-axis is to be directed anteriorlyrelative to the specimens initial unloaded position.1This practice is under the jurisdiction ofASTM Committee
14、 F04 on Medical andSurgical Materials and Devices and is the direct responsibility of SubcommitteeF04.25 on Spinal Devices.Current edition approved Jan. 1, 2010. Published February 2010. DOI: 10.1520/F279010.2For referenced ASTM standards, visit the ASTM website, www.astm.org, orcontact ASTM Custome
15、r Service at serviceastm.org. For Annual Book of ASTMStandards volume information, refer to the standards Document Summary page onthe ASTM website.1Copyright ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959, United States.3.1.1.3 Y-axispositive Y-axis is direc
16、ted laterally (towardthe left) relative to the specimens initial unloaded position.3.1.1.4 Z-axispositive Z-axis is to be directed superiorlyrelative to the specimens initial unloaded position.3.1.2 failurefunctional failure or substantial mechanicalfailure.3.1.2.1 functional failurepermanent deform
17、ation resultingfrom fracture, plastic deformation, or loosening beyond theultimate displacement or loosening that renders the spinalimplant assembly ineffective or unable to adequately resistload.3.1.2.2 mechanical failurefailure associated with a defectin the material (for example, fatigue crack) o
18、r of the bondingbetween materials that may or may not produce functionalfailure.3.1.3 fatigue lifethe number of cycles, N, that the FP cansustain at a particular load or moment before failure occurs.3.1.4 intended method of applicationa FP may containdifferent types of features to stabilize the impl
19、ant-tissue inter-face such as threads, spikes, and textured surfaces. Each typeof feature has an intended method of application or attachmentto the spine.3.1.5 insertion point of an anchorthe location where theanchor is attached to the test block. The insertion points shownin Fig. 1 are to be adhere
20、d to if possible. In situations where thedesign of the spinal implant assembly or the manufacturerssurgical instructions for installation dictate otherwise, theattachment points may deviate from these dimensions.3.1.6 longitudinal directionthe initial spatial orientationbetween the insertion points
21、in the superior test blocks and theinferior test blocks.3.1.7 maximum run-out load or momentthe maximumload or moment for a given test that can be applied to a FPwhere all of the tested constructs have withstood 10 000 000cycles without failure.3.1.8 mechanical deteriorationdeterioration that is vis
22、ibleto the naked eye and is associated with mechanical damage tothe device under test (for example, initiation of fatigue crack orsurface wear).3.1.9 permanent deformationthe remaining linear or an-gular displacement (axialmm, angulardegrees or radians)relative to the initial unloaded condition of t
23、he FP after theapplied load or moment has been removed.3.1.10 radius of rotationthe distance between the centerof rotation and the functional position (for example, load-bearing contact point) of the FP, for a given motion (that is,flexion/extension, lateral bending, or axial rotation).3.1.11 spinal
24、 implant assemblya complete spinal implantconfiguration as intended for surgical use. A spinal implantassembly may contain anchors, interconnections, and longitu-dinal elements and may contain transverse elements.3.1.12 stiffness (axialN/mm, angularNmm/degree orNmm/radian)the slope of the initial li
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
5000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ASTMF27902010STANDARDPRACTICEFORSTATICANDDYNAMICCHARACTERIZATIONOFMOTIONPRESERVINGLUMBARTOTALFACETPROSTHESES

链接地址:http://www.mydoc123.com/p-539474.html