《ASTM F2706-2008(2014) Standard Test Methods for Occipital-Cervical and Occipital-Cervical-Thoracic Spinal Implant Constructs in a Vertebrectomy Model《椎骨切除术模型中枕颈和枕颈-胸椎脊骨植入结构的标准试验方法》.pdf》由会员分享,可在线阅读,更多相关《ASTM F2706-2008(2014) Standard Test Methods for Occipital-Cervical and Occipital-Cervical-Thoracic Spinal Implant Constructs in a Vertebrectomy Model《椎骨切除术模型中枕颈和枕颈-胸椎脊骨植入结构的标准试验方法》.pdf(14页珍藏版)》请在麦多课文档分享上搜索。
1、Designation: F2706 08 (Reapproved 2014)Standard Test Methods forOccipital-Cervical and Occipital-Cervical-Thoracic SpinalImplant Constructs in a Vertebrectomy Model1This standard is issued under the fixed designation F2706; the number immediately following the designation indicates the year oforigin
2、al adoption or, in the case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. Asuperscript epsilon () indicates an editorial change since the last revision or reapproval.1. Scope1.1 These test methods cover the materials and methods forthe static
3、and fatigue testing of occipital-cervical and occipital-cervical-thoracic spinal implant assemblies in a vertebrectomymodel. The test materials for most combinations of occipital-cervical and occipital-cervical-thoracic spinal implant compo-nents can be specific depending on the intended location an
4、dintended method of attachment.1.2 These test methods are intended to provide a basis forthe mechanical comparison among past, present, and futureoccipital-cervical and occipital-cervical-thoracic spinal im-plant assemblies. They allow comparison of occipital-cervicaland occipital-cervical-thoracic
5、spinal implant constructs withdifferent methods of application to the spine. These testmethods are not intended to define levels of performance, sincesufficient knowledge is not available to predict the conse-quences of the use of a particular device.1.3 These test methods set out guidelines for loa
6、d types andmethods of applying loads. Methods for three static load typesand two fatigue tests for the comparative evaluation ofoccipital-cervical and occipital-cervical-thoracic spinal im-plant assemblies are defined.1.4 These test methods establish guidelines for measuringdisplacements, determinin
7、g the yield load, and evaluating thestiffness and strength of occipital-cervical or occipital-cervical-thoracic spinal implant assemblies.1.5 It may not be possible to test some occipital-cervical andsome occipital-cervical-thoracic spinal constructs in all testconfigurations.1.6 The values stated i
8、n SI units are to be regarded asstandard. No other units of measurement are included in thisstandard.1.7 This standard does not purport to address all of thesafety concerns, if any, associated with its use. It is theresponsibility of the user of this standard to establish appro-priate safety and hea
9、lth practices and determine the applica-bility of regulatory limitations prior to use.2. Referenced Documents2.1 ASTM Standards:2E4 Practices for Force Verification of Testing MachinesE6 Terminology Relating to Methods of Mechanical TestingE739 Practice for StatisticalAnalysis of Linear or Linearize
10、dStress-Life (S-N) and Strain-Life (-N) Fatigue DataE1823 Terminology Relating to Fatigue and Fracture TestingF1582 Terminology Relating to Spinal ImplantsF1717 Test Methods for Spinal Implant Constructs in aVertebrectomy ModelF2077 Test Methods For Intervertebral Body Fusion Devices3. Terminology3.
11、1 DefinitionsFor definitions of terms relating to thesetest methods, see Terminologies E6, F1582, and E1823.3.2 Definitions of Terms Specific to This Standard:3.2.1 active length of the longitudinal element, nthestraight line distance between the centers of rotation of the testblocks.3.2.2 block mom
12、ent arm, nthe perpendicular to the ap-plied load between the insertion point of an anchor and the axisof the hinge pin.3.2.3 compressive or tensile bending stiffness (N/mm),nthe compressive or tensile bending yield force divided byelastic displacement (see the initial slope of line BC in Fig. 1).3.2
13、.4 compressive or tensile bending ultimate load (N),nthe maximum compressive or tensile force in the X-Z planeapplied to an occipital-cervical or occipital-cervical-thoracicspinal implant assembly (see the force at Point E in Fig. 1). Theultimate load should be a function of the device and not of th
14、eload cell or testing machine.1These test methods are under the jurisdiction of ASTM Committee F04 onMedical and Surgical Materials and Devices and are the direct responsibility ofSubcommittee F04.25 on Spinal Devices.Current edition approved Oct. 1, 2014. Published November 2014. Originallyapproved
15、 in 2008. Last previous edition approved in 2008 as F2706-08. DOI:10.1520/F2706-08R14.2For referenced ASTM standards, visit the ASTM website, www.astm.org, orcontact ASTM Customer Service at serviceastm.org. For Annual Book of ASTMStandards volume information, refer to the standards Document Summary
16、 page onthe ASTM website.Copyright ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959. United States13.2.5 compressive or tensile bending yield load (N), nthecompressive or tensile bending force in the X-Z plane neces-sary to produce a permanent deformation equa
17、l to 0.020 timesthe active length of the longitudinal element (see the force atPoint D in Fig. 1).3.2.6 coordinate system/axes, nthree orthogonal axes aredefined in Figs. 2 and 3. The anterior-posterior axis is X withpositive being anterior. The medial-lateral axis is Y with leftbeing positive when
18、viewed posteriorly. The superior-inferioraxis is Z with superior being positive.3.2.7 displacement at 2 % offset yield (mm), nthe dis-placement of a construct measured via the actuator thatproduces a permanent deformation equal to 0.020 times theactive length of the longitudinal element (distance OA
19、 in Fig.1).3.2.8 elastic angular displacement (degrees), nthe angulardisplacement at 2 % offset yield (see Point A in Fig. 1) minusthe 2 % offset angular displacement (see Point B in Fig. 1) (thatis, the distance between Point A and Point B in Fig. 1).3.2.9 elastic displacement (mm), nthe displaceme
20、nt at2 % offset yield (see Point A in Fig. 1) minus the 2 % offsetdisplacement (see Point B in Fig. 1). (The distance betweenPoint A and Point B in Fig. 1.)3.2.10 failure, npermanent deformation resulting fromfracture, plastic deformation, or loosening beyond the ultimatedisplacement or loosening th
21、at renders the occipital-cervical oroccipital-cervical-thoracic spinal implant assembly ineffectiveor unable to adequately resist load.3.2.11 fatigue life, nthe number of loading cycles, N,ofaspecified character that the occipital-cervical or occipital-cervical-thoracic spinal implant assembly susta
22、ins before fail-ure of a specified nature occurs (see Terminology E1823).3.2.12 hinge pin, nthe cylindrical rod connecting a testblock to a side support. The superior and inferior aspects of thetest construct are each secured with a single 9.6-mm diameterpin.3.2.13 insertion point of an anchor, nthe
23、 location wherethe anchor is attached to the test block. The insertion pointsshown in Figs. 4-7 are to be adhered to, if possible. Insituations where the design of the occipital-cervical oroccipital-cervical-thoracic spinal implant assembly or themanufacturers surgical instructions for installation
24、dictateotherwise, the attachment points may deviate from thesedimensions.3.2.14 intended method of application, n occipital-cervical and occipital-cervical-thoracic spinal implant assem-blies contain different types of anchors. Each type of anchorhas an intended method of application to the spine.3.
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
5000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ASTMF270620082014STANDARDTESTMETHODSFOROCCIPITALCERVICALANDOCCIPITALCERVICALTHORACICSPINALIMPLANTCONSTRUCTSINAVERTEBRECTOMYMODEL

链接地址:http://www.mydoc123.com/p-539302.html