《ASTM F2624-2012(2016) Standard Test Method for Static Dynamic and Wear Assessment of Extra-Discal Single Level Spinal Constructs《盘外单节段脊柱结构静态 动态和磨损评估的标准试验方法》.pdf》由会员分享,可在线阅读,更多相关《ASTM F2624-2012(2016) Standard Test Method for Static Dynamic and Wear Assessment of Extra-Discal Single Level Spinal Constructs《盘外单节段脊柱结构静态 动态和磨损评估的标准试验方法》.pdf(16页珍藏版)》请在麦多课文档分享上搜索。
1、Designation: F2624 12 (Reapproved 2016)Standard Test Method forStatic, Dynamic, and Wear Assessment of Extra-DiscalSingle Level Spinal Constructs1This standard is issued under the fixed designation F2624; the number immediately following the designation indicates the year oforiginal adoption or, in
2、the case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. Asuperscript epsilon () indicates an editorial change since the last revision or reapproval.1. Scope1.1 This test method describes methods to assess the staticand dynamic properties of sin
3、gle level spinal constructs.1.2 An option for assessing wear using a weight loss methodand a dimensional analysis is given. This method, describedherein, is used for the analysis of devices intended for motionpreservation, using testing medium as defined in this standard(6.1).1.3 This test method is
4、 not intended to address any potentialfailure mode as it relates to the fixation of the device to its bonyinterfaces.1.4 It is the intent of this test method to enable single levelextra-discal spinal constructs with regard to kinematic,functional, and wear characteristics when tested under thespecif
5、ied conditions.1.5 This test method is not intended to address facetarthroplasty devices.1.6 In order that the data be reproducible and comparablewithin and between laboratories, it is essential that uniformprocedures be established. This test method is intended tofacilitate uniform testing methods
6、and data reporting.1.7 The motion profiles specified by this test method do notnecessarily accurately reproduce those occurring in vivo.Rather this method provides useful boundary/endpoint condi-tions for evaluating implant designs in a functional manner.1.8 This test method is not intended to be a
7、performancestandard. It is the responsibility of the user of this test methodto characterize the safety and effectiveness of the device underevaluation.1.9 Multiple test methods are included in this standard.However, it must be noted that the user is not obligated to testusing all of the described m
8、ethods. Instead, the user shouldonly select test methods that are appropriate for a particulardevice design. In most instances, only a subset of the hereindescribed test methods will be required.1.10 The values stated in SI units are to be regarded as thestandard with the exception of angular measur
9、ements, whichmay be reported in either degrees or radians. No other units ofmeasurement are included in this standard.1.11 This test method does not purport to address all of thesafety concerns, if any, associated with its use. It is theresponsibility of the user of this test method to establishappr
10、opriate safety and health practices and to determine theapplicability of regulatory limitations prior to use.2. Referenced Documents2.1 ASTM Standards:2E2309 Practices for Verification of Displacement MeasuringSystems and Devices Used in Material Testing MachinesF561 Practice for Retrieval and Analy
11、sis of MedicalDevices, and Associated Tissues and FluidsF1714 Guide for Gravimetric WearAssessment of ProstheticHip Designs in Simulator DevicesF1717 Test Methods for Spinal Implant Constructs in aVertebrectomy ModelF1877 Practice for Characterization of ParticlesF2003 Practice for Accelerated Aging
12、 of Ultra-High Mo-lecular Weight Polyethylene after Gamma Irradiation inAirF2423 Guide for Functional, Kinematic, and Wear Assess-ment of Total Disc Prostheses3. Terminology3.1 All terminology is consistent with the referencedstandards, unless otherwise stated.3.2 Definitions:3.2.1 center of rotatio
13、n (COR)the point about which thesimulated vertebral bodies rotate in performing the range ofmotion (ROM) specified in this test method.1This test method is under the jurisdiction ofASTM Committee F04 on Medicaland Surgical Materials and Devices and is the direct responsibility of SubcommitteeF04.25
14、on Spinal Devices.Current edition approved Dec. 1, 2016. Published December 2016. Originallyapproved in 2007. Last previous edition approved in 2012 as F2624 12. DOI:10.1520/F2624-12R16.2For referenced ASTM standards, visit the ASTM website, www.astm.org, orcontact ASTM Customer Service at serviceas
15、tm.org. For Annual Book of ASTMStandards volume information, refer to the standards Document Summary page onthe ASTM website.Copyright ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959. United StatesThis international standard was developed in accordance with i
16、nternationally recognized principles on standardization established in the Decision on Principles for theDevelopment of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.13.2.2 compressive bending stiffness (N/mm)th
17、e compres-sive bending yield force divided by elastic displacement (seethe initial slope of line BC in Fig. 1).3.2.3 compressive bending ultimate load (N)the maxi-mum compressive force in the X-Z plane applied to a spinalimplant assembly (see the force at Point E in Fig. 1). Theultimate load should
18、be a function of the device and not of theload cell or testing machine.3.2.4 compressive bending yield load (N)the compressivebending force in the X-Z plane necessary to produce apermanent deformation equal to 0.020 times the active lengthof the longitudinal element (see the force at Point D in Fig.
19、 1).3.2.5 coordinate system/axesthree orthogonal axes aredefined following a right-handed Cartesian coordinate system.The XY plane is to bisect the sagittal plane between superiorand inferior surfaces that are intended to simulate the adjacentvertebral end plates. The positive Z axis is to be direct
20、edsuperiorly. Force components parallel to the XY plane are shearcomponents of loading. The compressive axial force is definedto be the component in the negative Z direction. Torsional loadis defined to be the component of moment about the Z-axis.3.2.5.1 originthe center of the coordinate system is
21、lo-cated at the center of rotation of the testing fixture.3.2.5.2 X-Axisthe positive X-Axis is a global fixed axisrelative to the testing machines stationary base and is to bedirected anteriorly relative to the specimens initial unloadedposition.3.2.5.3 Y-Axisthe positive Y-Axis is a global fixed ax
22、isrelative to the testing machines stationary base and is directedlaterally relative to the specimens initial unloaded position.3.2.5.4 Z-Axisthe positive Z-Axis is a global fixed axisrelative to the testing machines stationary base and is to bedirected superiorly relative to the specimens initial u
23、nloadedposition.3.2.6 degradationloss of material or function or materialproperties due to causes other than that associated with wear.3.2.7 elastic displacement (mm or degrees)the displace-ment at 2 % offset yield (see Point A in Fig. 1) minus the 2 %offset displacement (see Point B in Fig. 1). (Th
24、e distancebetween Point A and Point B in Fig. 1.)3.2.8 fluid absorptionfluid absorbed by the device mate-rial during testing or while implanted in vivo.3.2.9 functional failurepermanent deformation or wearthat renders the implant assembly ineffective or unable toadequately resist load/motion or any
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
5000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ASTMF262420122016STANDARDTESTMETHODFORSTATICDYNAMICANDWEARASSESSMENTOFEXTRADISCALSINGLELEVELSPINALCONSTRUCTS

链接地址:http://www.mydoc123.com/p-539128.html