ASTM F2624-2007 Standard Test Method for Static Dynamic and Wear Assessment of Extra-Discal Spinal Motion Preserving Implants《防脊椎盘突出植入物静态、动态及磨损评定的标准试验方法》.pdf
《ASTM F2624-2007 Standard Test Method for Static Dynamic and Wear Assessment of Extra-Discal Spinal Motion Preserving Implants《防脊椎盘突出植入物静态、动态及磨损评定的标准试验方法》.pdf》由会员分享,可在线阅读,更多相关《ASTM F2624-2007 Standard Test Method for Static Dynamic and Wear Assessment of Extra-Discal Spinal Motion Preserving Implants《防脊椎盘突出植入物静态、动态及磨损评定的标准试验方法》.pdf(11页珍藏版)》请在麦多课文档分享上搜索。
1、Designation: F 2624 07Standard Test Method forStatic, Dynamic, and Wear Assessment of Extra-DiscalSpinal Motion Preserving Implants1This standard is issued under the fixed designation F 2624; the number immediately following the designation indicates the year oforiginal adoption or, in the case of r
2、evision, the year of last revision. A number in parentheses indicates the year of last reapproval. Asuperscript epsilon (e) indicates an editorial change since the last revision or reapproval.1. Scope1.1 This test method is intended to provide test methods forthe static, dynamic, and wear testing of
3、 extra-discal motionpreserving implants. These implants are intended to augmentspinal stability without significant tissue removal while allow-ing motion of the functional spinal unit(s).1.2 Wear is assessed using a weight loss method and adimensional analysis for determining wear of components used
4、in extra-discal spinal motion preserving procedures, usingtesting medium as defined in this test method (6.1).1.3 This test method is not intended to address any potentialfailure mode as it relates to the fixation of the device to its bonyinterfaces.1.4 It is the intent of this test method to enable
5、 comparisonof motion preserving, extra-discal implants with regard tokinematic, functional, and wear characteristics when testedunder the specified conditions. It must be recognized, however,that there are many possible variations in the in vivo condi-tions. A single laboratory simulation with a fix
6、ed set ofparameters may not be universally representative.1.5 This test method is not intended to address facetarthroplasty devices.1.6 This test method prescribes the use of pure angularrotations for assessing the mechanical characteristics of extra-discal motion preserving implants. This test meth
7、od does not,however, prescribe methods for assessing the mechanicalcharacteristics of the device in translation (for example,anterior/posterior translation), though this type of linear motionmay be clinically relevant.1.7 In order that the data be reproducible and comparablewithin and between labora
8、tories, it is essential that uniformprocedures are established. This test method is intended tofacilitate uniform testing methods and data reporting forextra-discal motion preserving implants.1.8 Without a substantial clinical retrieval history of spinal,motion preserving extra-discal implants, actu
9、al loading profilesand patterns cannot be delineated at the time of the writing ofthis test method. It therefore follows that the motion profilesspecified by this test method do not necessarily accuratelyreproduce those occurring in vivo. Rather this method providesuseful boundary/endpoint condition
10、s for evaluating implantdesigns in a functional manner.1.9 This test method is not intended to be a performancestandard. It is the responsibility of the user of this test methodto characterize the safety and effectiveness of the device underevaluation.1.10 The values stated in SI units are to be reg
11、arded as thestandard. The values given in parentheses are for informationonly.1.11 The values stated in SI units are to be regarded as thestandard with the exception of angular measurements, whichmay be reported in either degrees or radians.1.12 This standard does not purport to address all of thesa
12、fety concerns, if any, associated with its use. It is theresponsibility of the user of this standard to establish appro-priate safety and health practices and determine the applica-bility of regulatory limitations prior to use.2. Referenced Documents2.1 ASTM Standards:2F 561 Practice for Retrieval a
13、nd Analysis of Medical De-vices, and Associated Tissues and FluidsF 1714 Guide for Gravimetric Wear Assessment of Pros-thetic Hip-Designs in Simulator DevicesF 1717 Test Methods for Spinal Implant Constructs in aVertebrectomy ModelF 1877 Practice for Characterization of ParticlesF 2003 Practice for
14、Accelerated Aging of Ultra-High Mo-lecular Weight Polyethylene after Gamma Irradiation inAirF 2423 Guide for Functional, Kinematic, and Wear Assess-ment of Total Disc Prostheses3. Terminology3.1 All terminology is consistent with the referenced stan-dards, unless otherwise stated.1This test method i
15、s under the jurisdiction of ASTM Committee F04 on Medicaland Surgical Materials and Devices and is the direct responsibility of SubcommitteeF04.25 on Spinal Devices.Current edition approved Dec. 15, 2007. Published March 2008.2For referenced ASTM standards, visit the ASTM website, www.astm.org, orco
16、ntact ASTM Customer Service at serviceastm.org. For Annual Book of ASTMStandards volume information, refer to the standards Document Summary page onthe ASTM website.1Copyright ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959, United States.3.1.1 center of rota
17、tion (COR)the point about which thesimulated vertebral bodies rotate in performing the range ofmotion (ROM) specified in this test method.3.1.2 coordinate system/axesthree orthogonal axes aredefined following a right-handed Cartesian coordinate system.The XY-plane is to bisect the sagittal plane bet
18、ween superiorand inferior surfaces that are intended to simulate the adjacentvertebral end plates. The positive Z-axis is to be directedsuperiorly. Force components parallel to the XY-plane are shearcomponents of loading. The compressive axial force is definedto be the component in the negative Z-di
19、rection. Torsional loadis defined to be the component of moment about the Z-axis.3.1.3 degradationloss of material or function or materialproperties due to causes other than that associated with wear.3.1.4 extra-discal motion preserving device or implantanon-biologic structure, which lies entirely o
20、utside the interver-tebral disc space and is intended to at least partially support themotion/load between adjacent vertebral bodies. In this testmethod, this definition does not include facet arthroplastydevices.3.1.5 fluid absorptionfluid absorbed by the device mate-rial during testing or while im
21、planted in vivo.3.1.6 functional failurepermanent deformation or wearthat renders the extra-discal motion preserving implant assem-bly ineffective or unable to adequately resist load/motion orany secondary effects that result in a reduction of clinicallyrelevant motions or the motions intended by th
22、e design of thedevice.3.1.7 functional spinal unit (FSU)two adjacent vertebrae,including the intervertebral disc, and all adjoining ligamentsbetween them, specifically excluding all other connectivetissues such as muscles (Ref (1).33.1.8 interval net volumetric wear rateVRiduring cycleinterval i (mm
23、3/million cycles):VRi5WRir(1)where:r = mass density (for example, units of g/mm3) of the wearmaterial.3.1.9 interval net wear rateWRiduring cycle interval i(g/million cycles):WRi5NWiNWi1!# of cycles in interval i!3 106(2)Note, for i =1,NWi1=0.3.1.10 kinematic profilethe relative motion between adja-
24、cent vertebral bodies that the extra-discal motion preservingdevice is subjected to while being tested.3.1.11 mechanical failurefailure associated with a defectin the material (for example, fatigue crack) or of the bondingbetween materials that may or may not produce functionalfailure.3.1.12 net vol
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
5000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ASTMF26242007STANDARDTESTMETHODFORSTATICDYNAMICANDWEARASSESSMENTOFEXTRADISCALSPINALMOTIONPRESERVINGIMPLANTS

链接地址:http://www.mydoc123.com/p-539126.html