ASTM F2580-2009 Standard Test Method for Evaluation of Modular Connection of Proximally Fixed Femoral Hip Prosthesis《最接近固定股骨髋假体标准接头的评估用标准试验方法》.pdf
《ASTM F2580-2009 Standard Test Method for Evaluation of Modular Connection of Proximally Fixed Femoral Hip Prosthesis《最接近固定股骨髋假体标准接头的评估用标准试验方法》.pdf》由会员分享,可在线阅读,更多相关《ASTM F2580-2009 Standard Test Method for Evaluation of Modular Connection of Proximally Fixed Femoral Hip Prosthesis《最接近固定股骨髋假体标准接头的评估用标准试验方法》.pdf(5页珍藏版)》请在麦多课文档分享上搜索。
1、Designation: F 2580 09Standard Test Method forEvaluation of Modular Connection of Proximally FixedFemoral Hip Prosthesis1This standard is issued under the fixed designation F 2580; the number immediately following the designation indicates the year oforiginal adoption or, in the case of revision, th
2、e year of last revision. A number in parentheses indicates the year of last reapproval. Asuperscript epsilon () indicates an editorial change since the last revision or reapproval.1. Scope1.1 This test method covers a procedure for the fatiguetesting of metallic femoral hip prostheses used in hip jo
3、intreplacements. This test method covers the procedures for theperformance of fatigue tests on metallic femoral hip stemsusing a cyclic, constant-amplitude force. It applies to hipprostheses that utilize proximal metaphyseal fixation and are ofa modular construct, and it is intended to evaluate the
4、fatigueperformance of the modular connections in the metaphysealfilling (that is, proximal body) region of the stem.1.2 This test method is intended to provide useful, consis-tent, and reproducible information about the fatigue perfor-mance of metallic hip prostheses while held in a proximallyfixate
5、d manner, with the distal end not held by a pottingmedium.1.3 The values stated in SI units are to be regarded asstandard. No other units of measurement are included in thisstandard.1.4 This standard does not purport to address all of thesafety concerns, if any, associated with its use. It is theres
6、ponsibility of the user of this standard to establish appro-priate safety and health practices and determine the applica-bility of regulatory limitations prior to use.2. Referenced Documents2.1 ASTM Standards:2E 467 Practice for Verification of Constant Amplitude Dy-namic Forces in an Axial Fatigue
7、Testing SystemE 468 Practice for Presentation of Constant Amplitude Fa-tigue Test Results for Metallic MaterialsE 1150 Definitions of Terms Relating to Fatigue2.2 ISO Standards:3ISO 72064 Determination of Endurance Properties ofStemmed Femoral Components with Application of Tor-sion3. Terminology3.1
8、 Definitions:3.1.1 R value, nThe R value is the ratio of the minimumload to the maximum load.R 5minimum loadmaximum load3.2 Definitions of Terms Specific to This Standard:3.2.1 extractionremoval of the femoral hip implant fromthe femur during surgery.3.2.2 extractor holea hole in the proximal body o
9、f thestem in which an apparatus is placed to remove the implantfrom the femur.3.2.3 femoral headconvex spherical bearing member forarticulation with the natural acetabulum or prosthetic acetabu-lum.3.2.4 femoral head offsetthe perpendicular distance fromthe centerline of the implant stem to the cent
10、er of the femoralhead.3.2.5 frontal planethe plane that lies in the medial-lateraldirection of the implant. Adduction occurs in this plane.3.2.6 implant centerlinethe axis that runs vertically fromthe proximal body of the implant, down the center of the stemto the distal end.3.2.7 pivot axisthe cent
11、er of rotation of the pivot fixture(and prosthesis potted within it) within the test fixture setup; itslocation is determined by the intersection of the neck and stemcenterlines of the prothesis (Figs. 1 and 2).3.2.8 pivot fixturethe fixture in which the specimen ispotted, and is attached to the mai
12、n test fixture; characterized bytwo pins on the side that serve as the pivot axis.3.2.9 rotational planethe plane that lies perpendicular tothe stem axis of the implant.3.2.10 sagittal planethe plane that lies perpendicular tothe Frontal plane; flexion occurs in this plane.1This test method is under
13、 the jurisdiction of ASTM Committee F04 on Medicaland Surgical Materials and Devices and is the direct responsibility of SubcommitteeF04.22 on Arthroplasty.Current edition approved Sept. 15, 2009. Published September 2009. Originallyapproved in 2007. Last previous edition approved in 2007 as F 2580
14、07.2For referenced ASTM standards, visit the ASTM website, www.astm.org, orcontact ASTM Customer Service at serviceastm.org. For Annual Book of ASTMStandards volume information, refer to the standards Document Summary page onthe ASTM website.3Available from American National Standards Institute (ANS
15、I), 25 W. 43rd St.,4th Floor, New York, NY 10036, http:/www.ansi.org.1Copyright ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959, United States.4. Significance and Use4.1 This test method can be used to describe the effects ofmaterials, manufacturing, and desi
16、gn variables on the fatigueperformance of metallic femoral hip prostheses subject tocyclic loading for large numbers of cycles.4.2 The loading of femoral hip designs in vivo will, ingeneral, differ from the loading defined in this test method. Theresults obtained here cannot be used to directly pred
17、ict in vivoperformance. However, this test method is designed to allowfor comparisons between the fatigue performance of differentmetallic femoral hip designs, when tested under similar con-ditions.4.3 In order for fatigue data on femoral hip prostheses to becomparable, reproducible, and capable of
18、being correlatedamong laboratories, it is essential that uniform procedures beestablished.5. Specimen Selection5.1 The test component selected shall have the same geom-etry as the final product, and shall be in finished condition. Thetest component shall be of the worst-case size and configura-tion
19、(that is, the component that produces the highest stresses)of the implant family to be tested.5.2 The femoral head component selected for load applica-tion shall be of the same design and material as a currentproduct in use, but may be previously tested.5.3 The femoral head selected shall offer the
20、greatest loadoffset from the hip centerline, to represent a worst-casebending scenario during testing.6. Apparatus6.1 The hip implant may be tested in different orientationsto better reproduce specific testing conditions that are beingevaluated. For example: An anatomical orientation of 9flexion, an
21、d 10 adduction (per ISO 7206-4), or vertically inboth planes. The criteria used to determine the orientationshould be reported.6.2 Care shall be taken to ensure that the fixation of theimplant does not produce abnormal stress concentrations thatcould change the failure mode of the part.6.3 A fixed-b
22、earing load applicator shall be used to keep thespecimen aligned in the chosen orientation during testing, aswell as a fixture that allows the stem to bend during testing,such as a u-joint.6.4 The fixture used to hold the implant during testingshould have a reaction bolt that will oppose the loading
23、 on thefemoral head, keeping the implant in equilibrium. The positionof the reaction bolt should be adjustable to accommodate stemsof different lengths and design features.6.5 The fixtures and aligning materials used should be of adesign that positions the implant, when potted, so that: thepoint def
24、ined by the intersection of the neck and stem center-lines is coincident with the pivot axis (Fig. 1), the stem is fixedvertically in both medial/lateral and anterior/posterior direc-tions, the stem is aligned facing forward in the rotational plane(that is, the frontal plane is normal to the pivot a
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
5000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ASTMF25802009STANDARDTESTMETHODFOREVALUATIONOFMODULARCONNECTIONOFPROXIMALLYFIXEDFEMORALHIPPROSTHESIS

链接地址:http://www.mydoc123.com/p-539028.html