《ASTM F2391-2005(2011) Standard Test Method for Measuring Package and Seal Integrity Using Helium as the Tracer Gas《用氦做为探漏气体测量包装和密封完整性的标准试验方法》.pdf》由会员分享,可在线阅读,更多相关《ASTM F2391-2005(2011) Standard Test Method for Measuring Package and Seal Integrity Using Helium as the Tracer Gas《用氦做为探漏气体测量包装和密封完整性的标准试验方法》.pdf(8页珍藏版)》请在麦多课文档分享上搜索。
1、Designation: F2391 05 (Reapproved 2011)Standard Test Method forMeasuring Package and Seal Integrity Using Helium as theTracer Gas1This standard is issued under the fixed designation F2391; the number immediately following the designation indicates the year oforiginal adoption or, in the case of revi
2、sion, the year of last revision. A number in parentheses indicates the year of last reapproval. Asuperscript epsilon () indicates an editorial change since the last revision or reapproval.1. Scope1.1 This test method includes several procedures that can beused for the measurement of overall package
3、and seal barrierperformance of a variety of package types and package forms,as well as seal/closure types. The basic elements of this methodinclude:1.1.1 Helium (employed as tracer gas),1.1.2 Helium leak detector (mass spectrometer), and1.1.3 Package/product-specific test fixtures.1.1.4 Most applica
4、tions of helium leak detection are de-structive, in that helium needs to be injected into the packageafter the package has been sealed. The injection site then needsto be sealed/patched externally, which often destroys itssaleability.Alternatively, if helium can be incorporated into theheadspace bef
5、ore sealing, the method can be non-destructivebecause all that needs to be accomplished is to simply detectfor helium escaping the sealed package.1.2 Two procedures are described; however the supportingdata in Section 14 only reflects Procedure B (Vacuum Mode).The alternative, Sniffer Mode, has prov
6、en to be a valuableprocedure for many applications, but may have more variabil-ity due to exactly the manner that the operator conducts the testsuch as whether the package is squeezed, effect of multiplesmall leaks compared to fewer large leaks, background heliumconcentration, package permeability a
7、nd speed at which thescan is conducted. Further testing to quantify this proceduresvariability is anticipated, but not included in this version.1.2.1 Procedure A: Sniffer Modethe package is scannedexternally for helium escaping into the atmosphere or fixture.1.2.2 Procedure B: Vacuum Mode the helium
8、 containingpackage is placed in a closed fixture. After drawing a vacuum,helium escaping into the closed fixture (capture volume) isdetected. Typically, the fixtures are custom made for thespecific package under test.1.3 The sensitivity of the method can range from thedetection of:1.3.1 Large leaks1
9、0-2Pam3/s to 10-5Pam3/s (101cc/sec/atm to 10-4cc/sec/atm).1.3.2 Moderate leaks10-5Pam3/s to 10-7Pam3/s (10-4cc/sec/atm to 10-6cc/sec/atm).1.3.3 Fine leaks10-7Pam3/s to 10-9Pam3/s (10-6cc/sec/atm to 10-8cc/sec/atm).1.3.4 Ultra-Fine leak10-9Pam3/s to 10-11Pam3/s (10-8cc/sec/atm to 10-10cc/sec/atm).NOT
10、E 1Conversion from cc/sec/atm to Pam3/s is achieved by mul-tiplying by 0.1.1.4 The terms large, moderate, fine and ultra-fine are rela-tive terms only and do not imply the acceptability of any leakrate. The individual application dictates the level of integrityneeded. For many packaging applications
11、, only “large leaks”are considered unacceptable and the ability to detect smallerleaks is immaterial. All leak rates referred to in this method arebased on conversion of actual conditions (based on partialpressure of helium) to one atmosphere pressure differential andstandard temperature conditions.
12、1.5 The method may have applicability to any package type:1.5.1 Flexible,1.5.2 Semi-rigid, or1.5.3 Rigid.1.6 The sensitivities reported in the supporting data for thismethod pertain to the detectability of helium emanating fromthe sample and are not a function of the packaging form.1.7 The method is
13、 not applicable to breathable or porouspackaging.1.8 The results obtained can be qualitative, semi-quantitative or quantitative depending on the procedure used.1.9 Test fixture design is not within the scope of this methodexcept to note that different designs will be needed for differentapplications
14、 (which have different package types and packageintegrity requirements). Furthermore, the fixture selection anddesign will be based on where the testing is to be conductedwithin the manufacturing process (in other words, qualitycontrol versus research).1.10 This standard does not purport to address
15、all of thesafety concerns, if any, associated with its use. It is theresponsibility of the user of this standard to establish appro-priate safety and health practices and determine the applica-bility of regulatory limitations prior to use.1This test method is under the jurisdiction of ASTM Committee
16、 F02 on FlexibleBarrier Packaging and is the direct responsibility of Subcommittee F02.40 onPackage Integrity.Current edition approved April 1, 2011. Published April 2011. Originallyapproved in 2005. Last previous edition approved in 2005 as F2391 05. DOI:10.1520/F2391-05R11.1Copyright ASTM Internat
17、ional, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959, United States.2. Referenced Documents2.1 ASTM Standards:2D996 Terminology of Packaging and Distribution Environ-mentsD3078 Test Method for Determination of Leaks in FlexiblePackaging by Bubble EmissionD4991 Test Method for
18、Leakage Testing of Empty RigidContainers by Vacuum MethodE432 Guide for Selection of a Leak Testing MethodE479 Guide for Preparation of a Leak Testing SpecificationE493 Test Methods for Leaks Using the Mass SpectrometerLeak Detector in the Inside-Out Testing ModeE498 Test Methods for Leaks Using the
19、 Mass SpectrometerLeak Detector or Residual Gas Analyzer in the TracerProbe ModeE499 Test Methods for Leaks Using the Mass SpectrometerLeak Detector in the Detector Probe ModeE691 Practice for Conducting an Interlaboratory Study toDetermine the Precision of a Test MethodE1603 Test Methods for Leakag
20、e Measurement Using theMass Spectrometer Leak Detector or Residual Gas Ana-lyzer in the Hood ModeF17 Terminology Relating to Flexible Barrier PackagingF1327 Terminology Relating to Barrier Materials for Medi-cal Packaging32.2 Other Documents:Principal author L. Kirsch, et al - (shown in referenceApp
21、endix X1 as literature references 1, 2, 3 and 5)Principal author L. Nguyen, et al - (shown in referenceappendix I at literature reference 4)Co-authors include C. Moeckly, L. Nguyen, R. Gerth, W.Muangsiri, R. Scheire, D. M. Guazzo, L. Kirsch, G.Schmitt,A. Kirsch, M. Koch, T. Wertli, M. Lehman and G.S
22、chramm.3. Terminology3.1 General Term Definitions For definitions used in thisstandard see Terminology D996, Terminology F17 and Termi-nology F1327.3.2 Specific Term Definitions3.2.1 actual helium leak rate (AHLR)Measured heliumleak rate (MHLR) signal level adjusted to a driving force of100 % concen
23、tration at 101 KPa (1.0 atmosphere), absolute.3.2.2 breathable/porous packagingPackages, in whole orin part, that intentionally allow gases/vapors to flow freely intoand out of the package. (See also Terminology F1327)3.2.3 fine leaksFor the purpose of this test method, leaksthat exhibit gas/vapor l
24、eak rates between 1310-7Pam3/s to10-9Pam3/s (1310-6cc/sec/atmosphere to 1310-8cc/sec/atmosphere).3.2.4 flexible packagingPackages (typically, pouches, sa-chets, and bags) constructed of materials that are readilybendable. (See also Terminology Method F17)3.2.5 impermeable packagingPackages construct
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
5000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ASTMF239120052011STANDARDTESTMETHODFORMEASURINGPACKAGEANDSEALINTEGRITYUSINGHELIUMASTHETRACERGAS 做为 漏气

链接地址:http://www.mydoc123.com/p-538590.html