ASTM F2385-2015 Standard Practice for Determining Femoral Head Penetration into Acetabular Components of Total Hip Replacement Using Clinical Radiographs《使用临床射线摄影测定全髋关节置换的股骨头穿透进入髋臼.pdf
《ASTM F2385-2015 Standard Practice for Determining Femoral Head Penetration into Acetabular Components of Total Hip Replacement Using Clinical Radiographs《使用临床射线摄影测定全髋关节置换的股骨头穿透进入髋臼.pdf》由会员分享,可在线阅读,更多相关《ASTM F2385-2015 Standard Practice for Determining Femoral Head Penetration into Acetabular Components of Total Hip Replacement Using Clinical Radiographs《使用临床射线摄影测定全髋关节置换的股骨头穿透进入髋臼.pdf(5页珍藏版)》请在麦多课文档分享上搜索。
1、Designation: F2385 15Standard Practice forDetermining Femoral Head Penetration into AcetabularComponents of Total Hip Replacement Using ClinicalRadiographs1This standard is issued under the fixed designation F2385; the number immediately following the designation indicates the year oforiginal adopti
2、on or, in the case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. Asuperscript epsilon () indicates an editorial change since the last revision or reapproval.1. Scope1.1 This practice provides guidance for the measurement ofthe relative displac
3、ement of the femoral head and acetabularcomponent that result from wear and deformation occurring atthe articular interface of a total hip replacement from sequen-tial clinical radiographs.1.2 This practice is primarily intended for use in evaluatingpatients receiving THRs composed of a polyethylene
4、 acetabu-lar component articulating against a metal or ceramic femoralhead.1.3 So-called hard-on-hard articulations such as metal-on-metal and ceramic-on-ceramic THRs are not intended to bedirectly addressed.1.4 This practice will focus on computer assisted computa-tional methodologies for measuring
5、 relative displacementsover time but not to the exclusion of other methodologies.1.5 This practice describes methods for conducting a radio-graphic wear/creep study utilizing various computational meth-ods and is not intended to promote or endorse a particularmethod.1.6 It is not the intent of this
6、practice to provide detailedinstructions in the use of the various computational methods,which is contained in the respective user manuals.1.7 It is the intent of this practice to enable comparisons ofrelative displacements occurring in groups of patients receivingdifferent formulations of bearing m
7、aterials. It must berecognized, however, that there are many possible variations inthe in vivo conditions. A single clinical study may not beuniversally representative.1.8 This practice is not intended to be a performancestandard. It is the responsibility of the user of this practice tocharacterize
8、the safety and effectiveness of the prosthesis underevaluation.1.9 The values stated in SI units are to be regarded as thestandard, with the exception of angular measurements, whichmay be reported in either degrees or radians. Additionally,pixel density may also be reported in imperial units.1.10 Th
9、e use of this standard may involve the operation ofpotentially hazardous radiographic equipment and does notpurport to address the safety precautions associated withradiography. It is the responsibility of the user of this standardto define and establish appropriate safety practices. Thestandard doe
10、s not determine the applicability of regulatorylimitations prior to operating radiographic equipment.1.11 This standard does not purport to address all of thesafety concerns, if any, associated with its use. It is theresponsibility of the user of this standard to establish appro-priate safety and he
11、alth practices and determine the applica-bility of regulatory limitations prior to use.2. Referenced Documents2.1 ASTM Standards:2E177 Practice for Use of the Terms Precision and Bias inASTM Test MethodsE691 Practice for Conducting an Interlaboratory Study toDetermine the Precision of a Test Method3
12、. Terminology3.1 All radiographic terminology is consistent with thereferenced standards, unless otherwise stated.3.2 Definitions:3.2.1 radiostereometric analysis (RSA)a method devel-oped by Goren Selvik for measuring relative motion betweentwo parts from clinical radiographs (1).3This method utiliz
13、esin vivo tantalum beads, an external reference cage, and twox-ray generators which take two exposures simultaneously.1This practice is under the jurisdiction ofASTM Committee F04 on Medical andSurgical Materials and Devices and is the direct responsibility of SubcommitteeF04.22 on Arthroplasty.Curr
14、ent edition approved Nov. 1, 2015. Published December 2015. Originallyapproved in 2004. Last previous edition approved in 2010 as F2385 04(2010).DOI: 10.1520/F2385-15.2For referenced ASTM standards, visit the ASTM website, www.astm.org, orcontact ASTM Customer Service at serviceastm.org. For Annual
15、Book of ASTMStandards volume information, refer to the standards Document Summary page onthe ASTM website.3The boldface numbers in parentheses refer to the list of references at the end ofthis standard.Copyright ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959
16、. United States1There are several commercially available software/hardwarepackages for RSA analysis.3.2.2 markerstantalum beads 1.0 mm, 0.8 mm, or 0.5 mmin diameter.3.2.2.1 implant markersin vivo markers placed on theimplant in order to define the implant as a rigid body.3.2.2.2 cage markerstantalum
17、 beads held in an externalreference frame used to create a three dimensional coordinatesystem for measuring relative displacements.3.2.2.3 segmenta three dimensional rigid body defined bya minimum of three markers.3.2.3 edge detectionmethod of image analysis used todetermine the two dimensional or t
18、hree dimensional centerpoint of a curved surface. Many computational methods ofedge detection exist.3.2.4 coordinate system/axesthree orthogonal axes aredefined as follows:3.2.4.1 originthe center of the coordinate system is lo-cated at either the geometric center of the acetabular compo-nent segmen
19、t or the center of a circle defined using the edge ofthe acetabular component.3.2.4.2 X-axisthe positive X-axis is to be directed in themedial direction independent of which hip is to be studied.Some software programs correct the sign of this value but theuser must insure that the protocol maintains
20、 the convention,(that is, which way is the patient facing).3.2.4.3 Y-axisthe positive Y-axis is to be fixed in thesuperior direction.3.2.4.4 Z-axisthe positive Z-axis is to be fixed in theposterior direction.3.2.5 radio pairone set of RSA radiographs which weretaken simultaneously.3.3 There are a nu
21、mber of computational methods that canbe used to measure creep/wear of a polyethylene component.Adescription of a few of the commonly used current methods isgiven. This is not meant to be at the exclusion of othermethods.3.3.1 Martell methodas this software is informally re-ferred to in reference to
22、 its developer Dr. John Martell ofChicago University, is a semi-automated computer techniqueusing edge detection and the Hough transformation for thedetermination of polyethylene wear. This technique uses se-quential A/P films for two dimensional analysis and A/P andlateral sequential films for thre
23、e dimensional analysis (2).3.3.2 Polywear methoda software program developed byDr. Peter Devane of the Wellington School of Medicine, NewZealand is a semi-automated computer technique using edgedetection for the determination of polyethylene wear. Thistechnique uses sequential A/P films for two dime
24、nsionalanalysis and A/P and lateral sequential films for three dimen-sional analysis (3).3.3.3 UmRSAan RSA software program developed byBiomedical Innovations AB in Umea Sweden. It utilizessequential radio pairs in order to measure relative displace-ments between two segments or a point relative to
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
5000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ASTMF23852015STANDARDPRACTICEFORDETERMININGFEMORALHEADPENETRATIONINTOACETABULARCOMPONENTSOFTOTALHIPREPLACEMENTUSINGCLINICALRADIOGRAPHS

链接地址:http://www.mydoc123.com/p-538568.html