ASTM F2384-2010 Standard Specification for Wrought Zirconium-2 5Niobium Alloy for Surgical Implant Applications (UNS R60901)《外科植入物用锻制锆-2 5铌合金的标准规格(UNS R60901)》.pdf
《ASTM F2384-2010 Standard Specification for Wrought Zirconium-2 5Niobium Alloy for Surgical Implant Applications (UNS R60901)《外科植入物用锻制锆-2 5铌合金的标准规格(UNS R60901)》.pdf》由会员分享,可在线阅读,更多相关《ASTM F2384-2010 Standard Specification for Wrought Zirconium-2 5Niobium Alloy for Surgical Implant Applications (UNS R60901)《外科植入物用锻制锆-2 5铌合金的标准规格(UNS R60901)》.pdf(4页珍藏版)》请在麦多课文档分享上搜索。
1、Designation: F2384 10Standard Specification forWrought Zirconium-2.5Niobium Alloy for Surgical ImplantApplications (UNS R60901)1This standard is issued under the fixed designation F2384; the number immediately following the designation indicates the year oforiginal adoption or, in the case of revisi
2、on, the year of last revision. A number in parentheses indicates the year of last reapproval. Asuperscript epsilon () indicates an editorial change since the last revision or reapproval.1. Scope1.1 This specification covers the chemical, mechanical, andmetallurgical requirements for wrought zirconiu
3、m-2.5niobiumalloy to be used in the manufacture of surgical implants (1).21.2 The values stated in either SI units or inch-pound unitsare to be regarded separately as standard. The values stated ineach system may not be exact equivalents; therefore, eachsystem shall be used independently of the othe
4、r. Combiningvalues from the two systems may result in non-conformancewith the standard.2. Referenced Documents2.1 ASTM Standards:3B550/B550M Specification for Zirconium and ZirconiumAlloy Bar and WireE8/E8M Test Methods for Tension Testing of MetallicMaterialsE29 Practice for Using Significant Digit
5、s in Test Data toDetermine Conformance with SpecificationsE112 Test Methods for Determining Average Grain SizeE1552 Test Method for Determining Hafnium in Zirconiumand Zirconium Alloys By Direct Current PlasmaAtomicEmission SpectrometryE1941 Test Method for Determination of Carbon in Refrac-tory and
6、 Reactive Metals and Their AlloysE2626 Guide for Spectrometric Analysis of Reactive andRefractory MetalsF67 Specification for Unalloyed Titanium, for SurgicalImplant Applications (UNS R50250, UNS R50400, UNSR50550, UNS R50700)F748 Practice for Selecting Generic Biological Test Meth-ods for Materials
7、 and DevicesSI 10 American National Standard for Use of the Interna-tional System of Units (SI): The Modern Metric System2.2 ISO Standard:4ISO 6892 Metallic Materials Tensile Testing at AmbientTemperatureISO 9001 Quality Management SystemsRequirements3. Terminology3.1 Definitions of Terms Specific t
8、o This Standard:3.1.1 annealed, adjmaterial that exhibits a recrystalizedgrain structure.3.1.2 lot, nthe total number of mill products producedfrom the same melt heat under the same conditions at essen-tially the same time.4. Product Classification4.1 barrounds, flats or shapes from 4.76 to 101.60 m
9、m0.1875 to 4 in. in diameter or thickness (other sizes andshapes by special order).4.2 wirerounds or flats less than 4.76 mm 0.1875 in. indiameter or thickness.5. Ordering Information5.1 Include with inquiries and orders for material under thisspecification the following information:5.1.1 Quantity,5
10、.1.2 ASTM designation and date of issue,5.1.3 Units to be certifiedSI or inch-pound,5.1.4 Grade (if applicable),5.1.5 Form (bar, or wire),5.1.6 Condition (see 6.3),5.1.7 Mechanical properties (if applicable for special con-ditions),5.1.8 Finish (see 6.2),5.1.9 Applicable dimension including size, th
11、ickness, width,or drawing number,5.1.10 Special tests, if any, and1This specification is under the jurisdiction of ASTM Committee F04 onMedical and Surgical Materials and Devices and is the direct responsibility ofSubcommittee F04.12 on Metallurgical Materials.Current edition approved Dec. 1, 2010.
12、Published December 2010. Originallyapproved in 2005. Last previous edition approved in 2005 as F2384 05. DOI:10.1520/F2384-10.2The boldface numbers in parentheses refer to the list of references at the end ofthis standard.3For referenced ASTM standards, visit the ASTM website, www.astm.org, orcontac
13、t ASTM Customer Service at serviceastm.org. For Annual Book of ASTMStandards volume information, refer to the standards Document Summary page onthe ASTM website.4Available from American National Standards Institute (ANSI), 25 W. 43rd St.,4th Floor, New York, NY 10036.1Copyright ASTM International, 1
14、00 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959, United States.5.1.11 Other requirements.6. Materials and Manufacture6.1 Materials covered by this specification shall be pro-duced by multiple vacuum melting in arc furnaces, electronbeam melting, or other melting processes convent
15、ionally usedfor reactive metals.6.2 FinishThe mill product may be furnished to theimplant manufacturer as descaled or pickled, abrasivelyblasted, chemically milled, ground, machined, peeled, pol-ished, or as specified by the purchaser.6.3 ConditionBarstock shall be furnished in the annealedcondition
16、 unless otherwise specified.7. Chemical Requirements7.1 The heat analysis shall conform to the chemical com-position of Table 1. Ingot analysis may be used for reporting allchemical requirements, except hydrogen, oxygen, and nitro-gen. Samples for hydrogen, oxygen and nitrogen shall be takenfrom the
17、 finished mill product. The supplier shall not shipmaterial with chemistry outside the requirements specified inTable 1. Guide E2626 may be used as a guide for chemicalanalysis techniques.7.1.1 Requirements for the major and minor elementalconstituents are listed in Table 1. Also listed are importan
18、tresidual elements. Analysis for elements not listed in Table 1 isnot required to verify compliance with this specification.7.2 Product Analysis:7.2.1 Product analysis tolerances do not broaden the speci-fied heat analysis requirements but cover variations betweenlaboratories in the measurement of c
19、hemical content. Theproduct analysis tolerances shall conform to the producttolerances in Table 2.7.2.2 The product analysis is either for the purpose ofverifying the composition of a heat or manufacturing lot, or todetermine variations in the composition within the heat.7.2.3 Acceptance or rejectio
20、n of a heat or manufacturing lotof material may be made by the purchaser on the basis of thisproduct analysis.7.3 For referee purposes, use Test Method E1552 and E1941or other analytical methods, as agreed upon between thepurchaser and the supplier.7.4 The samples for chemical analysis shall be repr
21、esenta-tive of the material being tested. The utmost care must be usedin sampling zirconium for chemical analysis because of itsaffinity for elements such as oxygen, nitrogen, and hydrogen.In cutting samples for analysis, therefore, the operation shouldbe carried out insofar as possible in a dust-fr
22、ee atmosphere.Cutting tools should be clean and sharp. Samples for analysisshould be stored in suitable containers.8. Mechanical Requirements8.1 The material supplied under this specification shallconform to the mechanical property requirements in Table 3.8.2 Specimens for tension tests shall be mac
23、hined from barin the longitudinal direction and tested in accordance with TestMethods E8/E8M. Tensile properties shall be determined usinga strain rate of 0.003 to 0.007 in./in./min mm/mm/minthrough yield and then the crosshead speed may be increasedso as to produce fracture in approximately one add
24、itionalminute.8.3 Number of TestsPerform a minimum of two tensiontests from each lot (see 3.1.2). Should either of the two testspecimens not meet the specified requirements, test two addi-tional test pieces representative of the same lot in the samemanner. The lot will be considered in compliance on
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
5000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ASTMF23842010STANDARDSPECIFICATIONFORWROUGHTZIRCONIUM25NIOBIUMALLOYFORSURGICALIMPLANTAPPLICATIONSUNSR60901

链接地址:http://www.mydoc123.com/p-538564.html