ASTM F2381-2010 Standard Test Method for Evaluating Trans-Vinylene Yield in Irradiated Ultra-High-Molecular-Weight Polyethylene Fabricated Forms Intended for Surgical Implants by I.pdf
《ASTM F2381-2010 Standard Test Method for Evaluating Trans-Vinylene Yield in Irradiated Ultra-High-Molecular-Weight Polyethylene Fabricated Forms Intended for Surgical Implants by I.pdf》由会员分享,可在线阅读,更多相关《ASTM F2381-2010 Standard Test Method for Evaluating Trans-Vinylene Yield in Irradiated Ultra-High-Molecular-Weight Polyethylene Fabricated Forms Intended for Surgical Implants by I.pdf(5页珍藏版)》请在麦多课文档分享上搜索。
1、Designation: F2381 10Standard Test Method forEvaluating Trans-Vinylene Yield in Irradiated Ultra-HighMolecular Weight Polyethylene Fabricated Forms Intendedfor Surgical Implants by Infrared Spectroscopy1This standard is issued under the fixed designation F2381; the number immediately following the d
2、esignation indicates the year oforiginal adoption or, in the case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. Asuperscript epsilon () indicates an editorial change since the last revision or reapproval.1. Scope1.1 This test method describes
3、the measurement of thenumber of trans-vinylene groups in ultra-high molecularweight polyethylene (UHMWPE) intended for use in medicalimplants. The material is analyzed by infrared spectroscopy.1.2 This test method is based on Guide F2102.1.3 The applicability of the infrared method has beendemonstra
4、ted in other literature reports. This particularmethod, using the intensity (area) of the C-H absorptioncentered at 1370 cm-1to normalize for the samples thickness,will be validated by an Interlaboratory Study (ILS) conductedaccording to Practice E691.1.4 The values stated in SI units are to be rega
5、rded asstandard. No other units of measurement are included in thisstandard.1.5 This standard does not purport to address all of thesafety concerns, if any, associated with its use. It is theresponsibility of the user of this standard to establish appro-priate safety and health practices and determi
6、ne the applica-bility of regulatory limitations prior to use.2. Referenced Documents2.1 ASTM Standards:2E691 Practice for Conducting an Interlaboratory Study toDetermine the Precision of a Test MethodE1421 Practice for Describing and Measuring Performanceof Fourier Transform Mid-Infrared (FT-MIR) Sp
7、ectrom-eters: Level Zero and Level One TestsF2102 Guide for Evaluating the Extent of Oxidation inUltra-High-Molecular-Weight Polyethylene FabricatedForms Intended for Surgical Implants3. Terminology3.1 Definitions:3.1.1 trans-vinylene index (TVI)a trans-vinylene index isdefined as the ratio of the a
8、rea of the absorption peak centerednear 965 cm-1to the area of the absorption peak centered near1370 cm-1.3.1.2 depth locator (DL)a measurement of the distancefrom the articular surface, or surface of interest, that a spectrumwas collected and a corresponding TVI calculated.3.1.3 trans-vinylene inde
9、x profilea trans-vinylene indexprofile is defined as the graphical representation of variation ofthe samples trans-vinylene index with distance from itsarticular surface or the surface of interest. This is a plot of TVIversus DL. Typically, the graph will show the profile throughthe entire thickness
10、 of the sample.4. Significance and Use4.1 Published literature shows that the yield of radiolyticreactions that occur during radiation treatment increases withradiation dose level. Measurement of the products of thesereactions can be used as an internal dosimeter.4.2 Trans-vinylene unsaturations are
11、 formed during ioniza-tion treatment by abstraction of a hydrogen molecule, and to alesser extent by the recombination of two adjacent alkyl freeradicals that reside on the same chain.4.3 Previous work generated calibration curves of trans-vinylene absorption area as a function of absorbed radiation
12、dose, yielding a linear relationship for both gamma- andelectron beam-irradiated polyethylene.4.4 This data can be used to determine received dose as afunction of position, assuming a calibration curve (TVI versusradiation dose level) is known for the particular material andradiation conditions used
13、, and can be used to determineuniformity of dose level in irradiated polyethylene.1This test method is under the jurisdiction ofASTM Committee F04 on Medicaland Surgical Materials and Devices and is the direct responsibility of SubcommitteeF04.15 on Material Test Methods.Current edition approved Jun
14、e 1, 2010. Published July 2010. Originally approved2004. Last previous edition approved in 2004 as F238104. DOI: 10.1520/F2381-10.2For referenced ASTM standards, visit the ASTM website, www.astm.org, orcontact ASTM Customer Service at serviceastm.org. For Annual Book of ASTMStandards volume informat
15、ion, refer to the standards Document Summary page onthe ASTM website.1Copyright ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959, United States.5. Apparatus5.1 Infrared Spectrometer:5.1.1 A calibrated infrared spectrometer capable of record-ing a transmission
16、absorption spectrum over a minimum rangeof 900 to about 2000 cm-1using about 200 m-thick films at aresolution of 4 cm-1and an aperture of approximately 200 by200 m for a rectangular aperture, or 200 m diameter for acircular aperture.5.1.1.1 Other modes of collection (that is, reflection, attenu-ated
17、 total reflection (ATR), and so forth) and aperture andsampling step sizes may be used to generate the samplesabsorption spectrum provided they can be demonstrated toproduce equivalent results. Too large an aperture can result ina loss of profile accuracy.5.1.1.2 When a Fourier Transform Infrared (F
18、TIR) spec-trometer is used, a minimum of 32 scans shall be collected perspectrum.5.1.1.3 The FTIR instrument and sample compartmentshould be purged with a moisture- and carbon-dioxide-freeinert gas (for example, nitrogen, helium, or argon) to minimizespectral interference from these components.5.2 S
19、pecimen HolderEquipment, such as an x-y table,capable of accurately positioning the sample under the FTIRaperture with a minimum resolution at the scale of the aperturedimensions.5.3 MicrotomeEquipment capable of producing films ofthickness 200 m or less of a sample perpendicular to thearticular sur
20、face or the surface of interest.6. Sampling, Test Specimens, and Test Units6.1 Using a microtome, or other appropriate device, preparea thin slice of the sample about 200 m thick. If the detectedsignal from the FTIR is too weak with this thickness, a thickersample may be used.6.2 The slice shall typ
21、ically be taken near the center of thesamples articular surface or the surface of interest.6.3 The orientation of the slice shall typically be perpen-dicular to the articular surface or the surface of interest.6.4 Waviness in the infrared spectrum caused by Fourierrippling on the lower wavelengths c
22、an interfere with thetrans-vinylene absorbance at 965 cm-1. This rippling is causedby internal reflection of the infrared beam, and can be avoidedby lightly rubbing the sample film against 400 grit sandpaperuntil the film becomes translucent. This roughening procedureshould be done slowly to avoid h
23、eating the film, and should beperformed until the area under a single peak due to Fourierrippling is less than 10 % of the area under the trans-vinylenepeak.7. Preparation of Apparatus7.1 Prepare the infrared spectrometer for collection of atransmission absorption spectrum from a thin film of theUHM
24、WPE sample according to the manufacturers recommen-dations, Practice E1421, and the conditions described in 5.1.8. Procedure8.1 The test film (slice) shall be first configured in thespectrometer (after an appropriate background spectrum hasbeen collected) such that the aperture is positioned over th
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
5000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ASTMF23812010STANDARDTESTMETHODFOREVALUATINGTRANSVINYLENEYIELDINIRRADIATEDULTRAHIGHMOLECULARWEIGHTPOLYETHYLENEFABRICATEDFORMSINTENDEDFORSURGICALIMPLANTSBYIPDF

链接地址:http://www.mydoc123.com/p-538556.html