《ASTM F2183-2002 Standard Test Method for Small Punch Testing of Ultra-High Molecular Weight Polyethylene Used in Surgical Implants《外科植入物用超高分子重量聚乙烯的小型打孔器检测的标准试验方法》.pdf》由会员分享,可在线阅读,更多相关《ASTM F2183-2002 Standard Test Method for Small Punch Testing of Ultra-High Molecular Weight Polyethylene Used in Surgical Implants《外科植入物用超高分子重量聚乙烯的小型打孔器检测的标准试验方法》.pdf(4页珍藏版)》请在麦多课文档分享上搜索。
1、Designation: F 2183 02Standard Test Method forSmall Punch Testing of Ultra-High Molecular WeightPolyethylene Used in Surgical Implants1This standard is issued under the fixed designation F 2183; the number immediately following the designation indicates the year oforiginal adoption or, in the case o
2、f revision, the year of last revision. A number in parentheses indicates the year of last reapproval. Asuperscript epsilon (e) indicates an editorial change since the last revision or reapproval.1. Scope1.1 This test method covers the determination of mechanicalbehavior of ultra-high molecular weigh
3、t polyethylene (UHM-WPE) by small punch testing of miniature disk specimens (0.5mm in thickness and 6.4 mm in diameter). The test method hasbeen established for characterizing UHMWPE surgical mate-rials after ram extrusion or compression molding (1,2)2; forevaluating as-manufactured implants after r
4、adiation crosslink-ing and sterilization (3,4); as well as for testing of implants thathave been retrieved (explanted) from the human body (5,6).1.2 The parameters of the small punch test, namely the peakload, ultimate displacement, ultimate load, and work to failure,provide metrics of the yielding,
5、 ultimate strength, ductility, andtoughness of UHMWPE under multiaxial loading conditions.Because the mechanical behavior of UHMWPE is differentwhen loaded under uniaxial and multiaxial loading conditions(3), the small punch test provides a complementary mechanicaltesting technique to the uniaxial t
6、ensile testing specified formedical grade UHMWPE by Specification F 648.1.3 In addition to its use as a research tool in implantretrieval analysis, the small punch test can be used as alaboratory screening test to evaluate new UHMWPE materials,such as those created by gamma or electron beam irradiat
7、ion(1). The test method is also well suited for characterization ofUHMWPE before and after accelerated aging (for example,Guide F 2003), and in that regard it can provide ranking of themechanical degradation of different UHMWPE samples afteroxidative degradation (4,7).1.4 The small punch test has be
8、en applied to other polymers,including polymethyl methacrylate (PMMA) bone cement,polyacetal, and high density polyethylene (HDPE) (8,9). How-ever, the small punch testing of polymers other than UHMWPEis beyond the scope of this standard.1.5 This standard does not purport to address all of thesafety
9、 concerns, if any, associated with its use. It is theresponsibility of the user of this standard to establish appro-priate safety and health practices and determine the applica-bility of regulatory limitations prior to use.2. Referenced Documents2.1 ASTM Standards:D 695 Method for Compressive Proper
10、ties of Rigid Plas-tics3D 883 Terminology Relating to Plastics3E 4 Practices for Force Verification of Testing Machines4E 83 Practice for Verification and Classification of Exten-someters4F 648 Specification of Ultra-High Molecular Weight Poly-ethylene Powder and Fabricated Forms for Surgical Im-pla
11、nts5F 1714 Guide for Gravimetric Wear Assessment of Pros-thetic Hip Designs in Simulator Devices5F 1715 Guide for Wear Assessment of Prosthetic KneeDesigns in Simulator Devices5F 2003 Guide for Accelerated Aging of Ultra-High Molecu-lar Weight Polyethylene5F 2102 Guide for Evaluating the Extent of O
12、xidation inUltra-High-Molecular-Weight Polyethylene FabricatedForms Intended for Surgical Implants53. Terminology3.1 DefinitionsThe features of a typical small punch testload versus displacement curve for UHMWPE are illustrated inFig. 1.3.1.1 peak loadan initial local maximum in the loadversus displ
13、acement curve (Fig. 1). In certain radiationcrosslinked UHMWPE materials, the load versus displacementcurve increases monotonically and a shoulder, rather than aninitial peak load, may be observed.3.1.2 small punch testa test wherein the specimen is ofminiature size relative to conventional mechanic
14、al test speci-mens, is disk-shaped, and is loaded axisymmetrically inbending by a hemispherical-head punch.3.1.3 ultimate displacementthe displacement at rupture(failure) of the specimen (Fig. 1).3.1.4 ultimate loadthe load at rupture (failure) of thespecimen (Fig. 1).1This test method is under the
15、jurisdiction of ASTM Committee F04 on Medicaland Surgical Materials and Devicesand is the direct responsibility of SubcommitteeF04.15on Materials Test Methods.Current edition approved April 10, 2002. Published June 2002.2The boldface numbers in parentheses refer to the list of references at the end
16、ofthis standard.3Annual Book of ASTM Standards, Vol 08.01.4Annual Book of ASTM Standards, Vol 03.01.5Annual Book of ASTM Standards, Vol 13.01.1Copyright ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959, United States.3.1.5 work to failurethe area under the loa
17、d versus dis-placement curve (Fig. 1).4. Significance and Use4.1 Miniature specimen testing techniques are used to char-acterize the mechanical behavior of UHMWPE stock materialsand surgical implants after manufacture, sterilization, shelfaging, radiation crosslinking, thermal treatment, and implant
18、a-tion (1). Furthermore, experimental UHMWPE materials canbe evaluated after accelerated aging and hip or knee wearsimulation. Consequently, the small punch test makes it pos-sible to examine relationships between wear performance andmechanical behavior of UHMWPE. This test method can alsobe used to
19、 rank the mechanical behavior of UHMWPE relativeto a reference control material (such as the NIST Ultra-HighMolecular Weight Polyethylene Reference Material #8456).4.2 Small punch testing results may vary with specimenpreparation and with the speed and environment of testing.Consequently, where prec
20、ise comparative results are desired,these factors must be carefully controlled.5. Apparatus5.1 Small Punch Test Apparatus6A system consisting of ahemispherical head punch, a die, and a guide for the punch, asshown in Fig. 2. The parts shall be fabricated from a hardenedsteel.5.1.1 GuideThe function
21、of the guide is to align the punchrelative to the specimen, which rests in a disk-shaped recess.The inner diameter of the guide bore shall be 0.1010 +0.0002/-0.0000 in. (2.565 +0.005/-0.000 mm), and the specimen recessshall be 0.0200 +0.0004/-0.0000 in. (0.508 +0.010/-0.000 mm)in depth and 0.2520 6
22、0.0005 in. (6.401 6 0.013 mm) indiameter.5.1.2 DieThe function of the die is to constrain the sampleduring testing. The inner diameter of the die bore shall be0.1500 6 0.0005 in. (3.810 6 0.013 mm).5.1.3 PunchThe hemispherical head punch shall have adiameter of 0.1000 in. (2.540 mm), with a toleranc
23、e of+0.0000/-0.0002 in. (+0.000/-0.005 mm).5.2 Testing MachineAny suitable testing machine as de-scribed in Method D 695, consisting of a drive mechanism anda load indicator. The load indicator shall have a full range of250 N (56.2 lbs). The accuracy of the machine shall be verifiedat least once per
24、 year, as specified by Method D 695 andPractice E 4.5.3 CompressometerThis instrument, described in Sec-tion 5.2 from Method D 695, can be used to determine thedistance between the die and the punch during the test. If theactuator displacement of the testing machine can be shown todetermine punch di
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
5000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ASTMF21832002STANDARDTESTMETHODFORSMALLPUNCHTESTINGOFULTRAHIGHMOLECULARWEIGHTPOLYETHYLENEUSEDINSURGICALIMPLANTS

链接地址:http://www.mydoc123.com/p-537986.html