ASTM F560-2013e1 Standard Specification for Unalloyed Tantalum for Surgical Implant Applications (UNS R05200 UNS R05400)《外科植入物用非合金钽的标准规范 (UNS R05200 UNS R05400)》.pdf
《ASTM F560-2013e1 Standard Specification for Unalloyed Tantalum for Surgical Implant Applications (UNS R05200 UNS R05400)《外科植入物用非合金钽的标准规范 (UNS R05200 UNS R05400)》.pdf》由会员分享,可在线阅读,更多相关《ASTM F560-2013e1 Standard Specification for Unalloyed Tantalum for Surgical Implant Applications (UNS R05200 UNS R05400)《外科植入物用非合金钽的标准规范 (UNS R05200 UNS R05400)》.pdf(5页珍藏版)》请在麦多课文档分享上搜索。
1、Designation: F560 131Standard Specification forUnalloyed Tantalum for Surgical Implant Applications (UNSR05200, UNS R05400)1This standard is issued under the fixed designation F560; the number immediately following the designation indicates the year of originaladoption or, in the case of revision, t
2、he year of last revision. A number in parentheses indicates the year of last reapproval. A superscriptepsilon () indicates an editorial change since the last revision or reapproval.1NOTEThe designation was corrected editorially in December 2013.1. Scope*1.1 This specification covers the chemical, me
3、chanical, andmetallurgical requirements for unalloyed tantalum plate, sheet,strip, bar, and wire used in the manufacture of surgicalimplants.1.2 The values stated in either SI units or inch-pound unitsare to be regarded separately as standard. The values stated ineach system may not be exact equival
4、ents; therefore, eachsystem shall be used independently of the other. Combiningvalues of the two systems may result in non-conformance withthis standard.2. Referenced Documents2.1 ASTM Standards:2E8/E8M Test Methods for Tension Testing of Metallic Ma-terialsE29 Practice for Using Significant Digits
5、in Test Data toDetermine Conformance with SpecificationsF981 Practice for Assessment of Compatibility of Biomate-rials for Surgical Implants with Respect to Effect ofMaterials on Muscle and BoneIEEE/ASTM SI 10 American National Standard for Use ofthe International System of Units (SI): The Modern Me
6、tricSystem2.2 ISO Standards:3ISO 6892 Metallic Materials Tensile Testing at AmbientTemperatureISO 9001 Quality Management SystemsRequirementsISO 13782 Implants for SurgeryMetallic MaterialsUnalloyed Tantalum for Surgical Implant Applications3. Terminology3.1 Definitions of Terms Specific to This Sta
7、ndard:3.1.1 lotall material produced from the same ingot or asingle powder blend with the same cross section under thesame conditions at essentially the same time.3.1.2 platea flat product more than 4.75 mm 0.1875 in.in thickness.3.1.3 barmaterial 3.15 to 63.5 mm 0.125 to 2.5 in. indiameter in round
8、, hexagonal, or octagonal cross sectionsupplied in straight lengths.3.1.4 sheeta flat product 150 mm 6 in. or more in widthand from 0.13 to 4.75 mm 0.005 to 0.1875 in. in thickness.3.1.5 stripa flat product less than 150 mm 6 in. in widthand from 0.13 to 4.75 mm 0.005 to 0.1875 in. in thickness,may
9、be supplied in coil.3.1.6 wirematerial up to 3.15 mm 0.124 in. in diameterfurnished in coils or on spools or reels.4. Ordering Information4.1 Inquiries and orders under this specification shall in-clude the following information:4.1.1 Quantity (weight or number of pieces),4.1.2 ASTM designation, all
10、oy number, and date of issue,4.1.3 Units to be used for certificationSI or inch-pound.4.1.4 Composition designation (see 5.1)4.1.5 Form (strip, sheet, plate, bar, wire) (see 3.1),4.1.6 Condition (see 5.4),4.1.7 Applicable dimensions, including size, thickness,width, and length (random, exact, multip
11、les), or drawingnumber,4.1.8 Special tests,4.1.9 Special requirements, and4.1.10 Mechanical properties (if applicable for special con-ditions) (see 7.1).1This specification is under the jurisdiction of ASTM Committee F04 onMedical and Surgical Materials and Devices and is the direct responsibility o
12、fSubcommittee F04.12 on Metallurgical Materials.Current edition approved Oct. 1, 2013. Published November 2013. Originallyapproved in 1978. Last previous edition approved in 2008 as F560 08. DOI:10.1520/F0560-13E01.2For referenced ASTM standards, visit the ASTM website, www.astm.org, orcontact ASTM
13、Customer Service at serviceastm.org. For Annual Book of ASTMStandards volume information, refer to the standards Document Summary page onthe ASTM website.3Available from American National Standards Institute (ANSI), 25 W. 43rd St.,4th Floor, New York, NY 10036, http:/www.ansi.org.*A Summary of Chang
14、es section appears at the end of this standardCopyright ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959. United States15. Materials and Manufacture5.1 Material covered by this specification shall be madefrom vacuum-arc melted or electron-beam melted ingots(R0
15、5200) or powder-metallurgy consolidated (R05400) unal-loyed tantalum.5.2 The various tantalum mill products covered by thisspecification are formed with the conventional extrusion,forming, swaging, rolling, and drawing equipment normallyavailable in metalworking plants.5.3 FinishThe mill product may
16、 be furnished as descaledor pickled, abrasive blasted, chemically milled, ground,machined, peeled, polished, or as specified by the purchaser.5.4 Condition:5.4.1 Flat mill products shall be supplied in the cold-worked, cold-worked and stress-relieved or annealed condi-tion.5.4.2 Bar and wire product
17、s shall be supplied in the an-nealed or cold worked condition.6. Chemical Requirements6.1 The material shall conform to the chemical compositionrequirements in Table 1.6.1.1 Requirements for the major and minor elementalconstituents are listed in Table 1. Also listed are importantresidual elements.
18、Analysis for elements not listed in Table 1 isnot required to certify compliance with this specification.6.2 The ingot analysis shall be considered the chemicalanalysis for products supplied under this specification.6.3 When requested by the purchaser at the time ofpurchase, the supplier shall furni
19、sh a report certifying thevalues of carbon, oxygen, nitrogen, and hydrogen as specifiedin Table 2 for each lot of material supplied.7. Mechanical Properties7.1 The material supplied under this specification shallconform to the mechanical property requirements in Tables 3and 4. Mechanical properties
20、for material in conditions otherthan those included in Tables 3 and 4 shall be specified by thepurchaser.7.2 Specimens for tension tests shall be prepared and testedin accordance with Test Methods E8/E8M. Tensile propertiesshall be determined using a strain rate of 0.076 to 0.178mm/mm/min 0.003 to 0
21、.007 in./in./min through yield andthen the crosshead speed may be increased so as to producefracture in approximately one additional minute.7.3 Number of Tests:7.3.1 Bar and WirePerform at least one tension for eachlot. Should any of these test pieces not meet the specifiedrequirements, test two add
22、itional test pieces representative ofthe same lot, in the same manner, for each failed test piece. Thelot shall be considered in compliance only if all additional testpieces meet the specified requirements.7.3.2 Tensile test results for which any specimen fracturesoutside the gage length shall be co
23、nsidered acceptable, if boththe elongation and reduction of area meet the minimumrequirements specified. Refer to Test Method E8/E8M, sections7.11.4 and 7.11.5.7.3.3 If either the elongation or the reduction of area is lessthan the minimum requirement, discard the test and retest.Retest one specimen
24、 for each specimen that did not meet theminimum requirements.7.3.4 Sheet, Strip, and PlatePerform at least one tensiontest for each lot. Tension property requirements apply in boththe longitudinal and transverse directions. Tests in the trans-verse direction need be made only on product from which a
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便