ASTM F1397-1993(2012) Standard Test Method for Determination of Moisture Contribution by Gas Distribution System Components《测定气体分配系统组件水量分布的标准试验方法》.pdf
《ASTM F1397-1993(2012) Standard Test Method for Determination of Moisture Contribution by Gas Distribution System Components《测定气体分配系统组件水量分布的标准试验方法》.pdf》由会员分享,可在线阅读,更多相关《ASTM F1397-1993(2012) Standard Test Method for Determination of Moisture Contribution by Gas Distribution System Components《测定气体分配系统组件水量分布的标准试验方法》.pdf(11页珍藏版)》请在麦多课文档分享上搜索。
1、Designation: F1397 93 (Reapproved 2012)Standard Test Method forDetermination of Moisture Contribution by Gas DistributionSystem Components1This standard is issued under the fixed designation F1397; the number immediately following the designation indicates the year oforiginal adoption or, in the cas
2、e of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. Asuperscript epsilon () indicates an editorial change since the last revision or reapproval.INTRODUCTIONSemiconductor clean rooms are serviced by high-purity gas distribution systems. This test m
3、ethodpresents a procedure that may be applied for the evaluation of one or more components considered foruse in such systems.1. Scope1.1 This test method covers testing components for totalmoisture contribution to a gas distribution system at ambienttemperature. In addition, the test method allows t
4、esting atelevated ambient temperatures as high as 70C and of thecomponent moisture capacity and recovery.1.2 This test method applies to in-line components contain-ing electronics grade materials such as those used in semicon-ductor gas distribution systems.1.3 Limitations:1.3.1 This test method is
5、limited by the sensitivity of currentinstrumentation, as well as by the response time of theinstrumentation. This test method is not intended to be used fortest components larger than 12.7-mm (12-in.) outside diameternominal size. This test method could be applied to largercomponents; however, the s
6、tated volumetric flow rate may notprovide adequate mixing to ensure a representative sample.Higher flow rates may improve the mixing but excessivelydilute the sample.1.3.2 This test method is written with the assumption thatthe operator understands the use of the apparatus at a levelequivalent to si
7、x months of experience.1.4 The values stated in SI units are to be regarded as thestandard. The inch-pound units given in parentheses are forinformation only.1.5 This standard does not purport to address all of thesafety concerns, if any, associated with its use. It is theresponsibility of the user
8、of this standard to establish appro-priate safety and health practices and determine the applica-bility of regulatory limitations prior to use. Specific hazardstatements are given in Section 5.2. Terminology2.1 Definitions:2.1.1 baselinethe instrument response under steady stateconditions.2.1.2 glov
9、e bagan enclosure that contains a controlledatmosphere. A glove box could also be used for this testmethod.2.1.3 heat trace heating of a component, spool piece, ortest stand by a uniform and complete wrapping of the item withresistant heat tape.2.1.4 minimum detection limit (MDL) of the instrumentth
10、elowest instrument response detectable and readable by theinstrument and at least two times the amplitude of the noise.2.1.5 response timethe time required for the system toreach steady state after a change in concentration.2.1.6 spool piecea null component, consisting of astraight piece of electrop
11、olished tubing and appropriate fittings,used in place of the test component to establish the baseline.2.1.7 standard conditions101.3 kPa, 0.0C (14.73 psia,32F).2.1.8 test componentany device being tested, such as avalve, regulator, or filter.2.1.9 test standthe physical test system used to measureim
12、purity levels.2.1.10 V-1, V-2inlet and outlet valves of bypass loop,respectively.2.1.11 V-3, V-4inlet and outlet valves of test loop, respec-tively.1This test method is under the jurisdiction of ASTM Committee F01 onElectronics and is the direct responsibility of Subcommittee F01.10 on Contamina-tio
13、n Control.Current edition approved July 1, 2012. Published August 2012. Originallyapproved in 1992. Last previous edition approved in 2005 as F1397 93(2005).DOI: 10.1520/F1397-93R12.Copyright ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959. United States12.1.
14、12 zero gasa purified gas that has an impurityconcentration below the MDL of the analytical instrument.This gas is to be used for both instrument calibration andcomponent testing.2.2 Abbreviations:2.2.1 MFCmass flow controller.2.2.2 ppbvparts per billion by volume assuming ideal gasbehavior, equival
15、ent to nmole/mole (such as nL/L). The sameas molar parts per billion (ppb).2.2.3 ppbwparts per billion by weight (such as ng/g).2.2.4 ppmvparts per million by volume assuming idealgas behavior, equivalent to mole/mole (such as L/L). Thesame as molar parts per million (ppm).2.2.5 ppmwparts per millio
16、n by weight (such as g/g).2.3 Symbols:2.3.1 P1The inlet pressure measured upstream of thepurifier and filter in the test apparatus.2.3.2 P2The outlet pressure measured downstream of theanalyzer in the test apparatus.2.3.3 Q1the bypass sample flow not going through theanalytical system.2.3.4 Q2the to
17、tal sample flow through the analyticalsystem.2.3.5 Qsthe flow through the spool piece or component.2.3.6 Tathe temperature of the air discharged by theanalyzers cooling exhaust.2.3.7 Tsthe temperature of the spool piece or component.2.3.7.1 DiscussionThe thermocouple must be in contactwith the outsi
18、de wall of the component or spool piece.3. Significance and Use3.1 The purpose of this test method is to define a procedurefor testing components being considered for installation into ahigh-purity gas distribution system. Application of this testmethod is expected to yield comparable data among com
19、po-nents tested for the purposes of qualification for this installa-tion.4. Apparatus4.1 Materials:4.1.1 Nitrogen or Argon, clean, dry, as specified in 8.4.4.1.2 Spool Piece, that can be installed in place of the testcomponent is required. This piece is to be a straight section of316L electropolishe
20、d stainless steel tubing with no restrictions.The length of the spool piece shall be 200 mm (0.8 in.). Thespool piece should have the same end connections as the testcomponent.4.1.3 Tubing, used downstream of the purifier shall be 316Lelectropolished stainless steel seamless tubing. The diameter oft
21、he sample line to the analyzer shall not be larger than 6.4 mm(14 in.). The length of the sample line from the tee (installedupstream of the pressure gage P2) to the analyzer shall not bemore than 600 mm (2.4 in.) to minimize the effect (adsorption/desorption) of the sample line on the result. The s
22、ample lineshall have no more than two mechanical joints.4.1.3.1 Components With Stub EndsUse compression fit-tings with nylon or teflon ferrules to connect the spool pieceand test component to the test loop. Keep the purged glove bagaround each component for the duration of the test. In the caseof l
23、ong pieces of electropolished tubing, use two glove bags,one at each end.4.1.4 Valves, must be diaphragm or bellows type and ca-pable of unimpaired operation at 94C (200F). The use ofall-welded, all-metal valves is preferred.4.2 Instrumentation:4.2.1 Moisture AnalyzerMoisture analyzers (such aselect
24、rolytic, piezo-electric, chilled mirror, or opto-electronic)are used to measure moisture levels. The analyzer is to beplaced downstream of the test component. Accurate baselinereadings must be obtained prior to and subsequent to each ofthe tests. Excessive deviations in baseline levels (620 ppbv)bef
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
5000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ASTMF139719932012STANDARDTESTMETHODFORDETERMINATIONOFMOISTURECONTRIBUTIONBYGASDISTRIBUTIONSYSTEMCOMPONENTS

链接地址:http://www.mydoc123.com/p-534807.html