《ASTM F1026-1986(2008)e1 Standard Specification for General Workmanship and Performance Measurements of Hemostatic Forceps《止血钳的一般工艺和性能测量的标准规范》.pdf》由会员分享,可在线阅读,更多相关《ASTM F1026-1986(2008)e1 Standard Specification for General Workmanship and Performance Measurements of Hemostatic Forceps《止血钳的一般工艺和性能测量的标准规范》.pdf(2页珍藏版)》请在麦多课文档分享上搜索。
1、Designation: F 1026 86 (Reapproved 2008)e1Standard Specification forGeneral Workmanship and Performance Measurements ofHemostatic Forceps1This standard is issued under the fixed designation F 1026; the number immediately following the designation indicates the year oforiginal adoption or, in the cas
2、e of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. Asuperscript epsilon () indicates an editorial change since the last revision or reapproval.This standard has been approved for use by agencies of the Department of Defense.e1NOTEEditorial change
3、s were made throughout in June 2008.1. Scope1.1 This specification covers general workmanship aspectsof hemostatic forceps fabricated from stainless steel andintended for reuse in surgery.2. Referenced Documents2.1 ASTM Standards:2E18 Test Methods for Rockwell Hardness of MetallicMaterialsE92 Test M
4、ethod for Vickers Hardness of Metallic Mate-rialsE 140 Hardness Conversion Tables for Metals RelationshipAmong Brinell Hardness, Vickers Hardness, RockwellHardness, Superficial Hardness, Knoop Hardness, andScleroscope HardnessF 899 Specification for Wrought Stainless Steels for Surgi-cal Instruments
5、F 921 Terminology Relating to Hemostatic Forceps3. Terminology3.1 Definitions applicable to hemostatic forceps shall be inaccordance with Definitions F 921.4. Material4.1 All of the component parts of the instruments shall bemade of martensitic stainless steel of Type 410, 410X, 416,416MOD, 420A, or
6、 420B of Specification F 899.5. Physical Requirements5.1 Heat Treatment and Hardness for Component Parts:5.1.1 The component parts of the instruments shall be heattreated under conditions recommended for the material used.Typical heat treating guidelines and hardness values are givenin Specification
7、 F 899.5.1.2 The hardness of all opposing parts of the sameinstrument shall not vary in hardness by more than 4 units onthe Rockwell Hardness C scale (HRC) or equivalent. (See TestMethods E18.)5.2 PassivationInstruments and instrument componentsshall be passivated after completion of all fabricating
8、 andfinishing operations.6. Performance Requirements6.1 Finger RingsInside surfaces shall be well roundedand polished and shall comply with the requirements in 7.1.6.2 Jaw Serrations and TeethThe serrations and teethshall be of uniform depth and height and well defined toprovide effective gripping.
9、The serrations and teeth shallinterdigitate. The edges of the serrations shall be chamfered.6.3 Box LockThe hemostatic forceps shall be of box lockconstruction, as defined in Definitions F 921. The pin, or otherfastening component, shall be permanently secured. The jointperformance of the box lock s
10、hall be smooth, of equalresistance, and non-binding when opening or closing theforceps to an included angle of 90 6 5 (the lock may belubricated).6.4 ClearanceThe maximum clearance between the maleand female members of the forceps in the lock area shall be 0.4mm (0.015 in.). The clearance is the vis
11、ible gap that existswhen the instrument is viewed from both the front and sideprofile or end view of the instrument (excluding bevel) (seeFig. 1).6.5 RatchetsRatchets shall securely engage at eachratchet position in a smooth and progressive manner.1This specification is under the jurisdiction of AST
12、M Committee F04 onMedical and Surgical Materials and Devices and is the direct responsibility ofSubcommittee F04.33 on Medical/Surgical Instruments.Current edition approved Feb. 1, 2008. Published March 2008. Originallyapproved in 1986. Last previous edition approved in 2002 as F 1026 86 (2002).2For
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
5000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ASTMF102619862008E1STANDARDSPECIFICATIONFORGENERALWORKMANSHIPANDPERFORMANCEMEASUREMENTSOFHEMOSTATICFORCEPS

链接地址:http://www.mydoc123.com/p-533997.html