《ASTM E879-2001(2007) Standard Specification for Thermistor Sensors for Clinical Laboratory Temperature Measurements《临床实验室温度测量用热敏电阻探测器标准规范》.pdf》由会员分享,可在线阅读,更多相关《ASTM E879-2001(2007) Standard Specification for Thermistor Sensors for Clinical Laboratory Temperature Measurements《临床实验室温度测量用热敏电阻探测器标准规范》.pdf(10页珍藏版)》请在麦多课文档分享上搜索。
1、Designation: E 879 01 (Reapproved 2007)Standard Specification forThermistor Sensors for Clinical Laboratory TemperatureMeasurements1This standard is issued under the fixed designation E 879; the number immediately following the designation indicates the year oforiginal adoption or, in the case of re
2、vision, the year of last revision. A number in parentheses indicates the year of last reapproval. Asuperscript epsilon (e) indicates an editorial change since the last revision or reapproval.1. Scope1.1 This specification covers the general requirements fornegative temperature coefficient thermistor
3、-type sensors in-tended to be used for clinical laboratory temperature measure-ments or control, or both, within the range from 10 to 105 C.1.2 This specification also covers the detailed requirementsfor ASTM designated sensors.1.3 This standard does not purport to address all of thesafety concerns,
4、 if any, associated with its use. It is theresponsibility of the user of this standard to establish appro-priate safety and health practices and determine the applica-bility of regulatory limitations prior to use.2. Referenced Documents2.1 ASTM Standards:2E 344 Terminology Relating to Thermometry an
5、d Hydrom-etryE 563 Practice for Preparation and Use of an Ice-Point Bathas a Reference TemperatureF 29 Specification for Dumet Wire for Glass-to-Metal SealApplications3. Terminology3.1 DefinitionsThe definitions given in TerminologyE 344 shall apply to this specification.3.2 Definitions of Terms Spe
6、cific to This Standard:3.2.1 dissipation constant, d, nthe ratio of the change inenergy dissipated per unit time (power) in a thermistor,DP = P2 P1, to the resultant temperature change of thethermistor, DT = T2 T1.d5DPDT(1)The dimensions of the dissipation constant are W/K.For this specification, T1
7、is in the range from 20 to 38 C andDT = 10 C.3.2.2 dumet, nround, copper-coated 42 % nickel-ironwire intended primarily for sealing to soft glass.Also known asCuNiFe in some communities.3.2.3 insulation resistance, dc, nthe resistance at a speci-fied direct-current voltage between the insulated lead
8、s of athermistor sensor and the metallic enclosure of the sensor, ifsuch an enclosure is present, or else between the sensor leadsand a conductive medium in which the sensor is immersed.3.2.4 qualification test, na series of tests conducted by theprocuring agency or an agent thereof to determine con
9、form-ance of thermistor sensors to the requirements of a specifica-tion, normally for the development of a qualified products listunder the specification.3.2.5 response time, nthe time required for a sensor tochange a specified percentage of the total difference betweenits initial and final temperat
10、ures as determined from zero-power resistances when the sensor is subjected to a stepfunction change in temperature.3.2.6 time constant, nthe 63.2 % response time of asensor that exhibits a single-exponential response.3.2.7 zero-power resistance, nthe dc resistance of a de-vice, at a specified tempe
11、rature, calculated for zero-power.3.2.7.1 DiscussionAccurate zero-power resistance is ob-tained by extrapolating to zero-power the resistance valuesobtained from measurements at three or more levels of powerwith the sensor immersed in a constant temperature medium.For the purpose of this specificati
12、on, this is obtained frommeasurements at a single power level adjusted such that thepower is not greater than one-fifth the product of the dissipa-tion constant specified in Table 1 (see 3.2.1 and 7.3) and theappropriate tolerance requirement of Table 2. When makingstability measurements, the power
13、shall be kept constant.4. Classification4.1 Thermistor sensors covered by this specification shall beclassified with a type designation code that includes theASTMdetailed specification number followed by a descriptive code.See Fig. 1.1This specification is under the jurisdiction of ASTM Committee E2
14、0 onTemperature Measurement and is the direct responsibility of Subcommittee E20.08on Medical Thermometry.Current edition approved May 1, 2007. Published June 2007. Originallyapproved in 1982. Last previous edition approved in 2001 as E 879 01.2For referenced ASTM standards, visit the ASTM website,
15、www.astm.org, orcontact ASTM Customer Service at serviceastm.org. For Annual Book of ASTMStandards volume information, refer to the standards Document Summary page onthe ASTM website.1Copyright ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959, United States.TA
16、BLE 1 Specification for ASTM Clinical Laboratory Thermistor SensorsASTM No. E 879 S B1N E 879 E B1N E 879 S B2N E 879 E B2NDescription Silicone Rubber Coated GlassProbeEpoxy Coated Glass Probe Silicone Rubber Coated GlassProbeEpoxy Coated Glass ProbeMajor Application General Purpose ClinicalLaborato
17、ry TemperatureMeasurementGeneral Purpose ClinicalLaboratory TemperatureMeasurementGeneral Purpose ClinicalLaboratory TemperatureMeasurementGeneral Purpose ClinicalLaboratory TemperatureMeasurementOperating Temperature Range 10 to 60 C 10 to 60 C 10 to 60 C 10 to 60 CAccuracy Class 1 (60.01 C) 1 (60.
18、01 C) 2 (60.02 C) 2 (60.02 C)Temperatures for Accuracy Class 10 to 60 C 10 to 60 C 10 to 60 C 10 to 60 CAccuracies for other TemperaturesWithin Specified TemperatureRangeCalibration Type Non-interchangeable Non-interchangeable Non-interchangeable Non-interchangeableNominal R-T CharacteristicR = expA
19、0+ A1/T + A2/T2+A3/T3 (T in kelvins)T( C) = a0+a1LnR +a2(LnR)2+a3(LnR)31 273.15A04.4495078 4.4495078 4.4495078 4.4495078A13614.7764 3614.7764 3614.7764 3614.7764A288190.906 88190.906 88190.906 88190.906A322328247 22328247 22328247 22328247a00.11766716 3 1020.11766716 3 1020.11766716 3 1020.11766716
20、3 102a10.28173082 3 1030.28173082 3 1030.28173082 3 1030.28173082 3 103a20.23285292 3 1050.23285292 3 1050.23285292 3 1050.23285292 3 105a30.24131652 3 1060.24131652 3 1060.24131652 3 1060.24131652 3 106Type of Immersion Fluid water, air water, oil, air water, air water, oil, airNominal R0at 25 C 2.
21、5 kV 2.5 kV 2.5kV 2.5 kVDissipation Constant 3.5 6 0.9 mW/K 5.0 6 1.2 mW/K 3.56 0.9 mW/K 5.0 6 1.2 mW/K63.2 % Response Time 0.55 6 0.16 s 0.45 6 0.11 s 0.55 6 0.16 s 0.45 6 0.11 sRatio of 95 % to 63.2 % ResponseTimes2.5 6 0.6 2.1 6 0.5 2.5 6 0.6 2.1 6 0.5Design and Construction Fig. 2 Fig. 3 Fig. 2
22、Fig. 3ASTM No. E 879 S A2N E 879 E A2N E 879 G B1N E 879 H B1NDescription Silicone Rubber Coated GlassProbeEpoxy Coated Glass Probe5kV -4-WireNon-interchangeable Sensorin S.S. Housing5kV -2-WireNon-interchangeableSensorin S.S. HousingMajor Application General Purpose ClinicalLaboratory TemperatureMe
23、asurementGeneral Purpose ClinicalLaboratory TemperatureMeasurementGeneral Purpose ClinicalLaboratory TemperatureMeasurementGeneral Purpose ClinicalLaboratory TemperatureMeasurementOperating Temperature Range 10 to 105 C 10 to 105 C 10 to 60 C 10 to 60 CAccuracy Class 2 (60.02 C) 2 (60.02 C) 1 (60.01
24、 C) 1 (60.01 C)Temperatures for Accuracy Class 10 to 105 C 10 to 105 C 10 to 60 C 10 to 60 CAccuracies for other TemperaturesWithin Specified TemperatureRangeCalibration Type Non-interchangeable Non-interchangeable Non-interchangeable Non-interchangeableNominal R-T CharacteristicR = expA0+ A1/T + A2
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
5000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ASTME87920012007STANDARDSPECIFICATIONFORTHERMISTORSENSORSFORCLINICALLABORATORYTEMPERATUREMEASUREMENTS

链接地址:http://www.mydoc123.com/p-533715.html