ASTM E803-1991(2013) 1250 Standard Test Method for Determining the L D&8201 Ratio of Neutron Radiography Beams《测定中子射线照相束L D比率的标准试验方法》.pdf
《ASTM E803-1991(2013) 1250 Standard Test Method for Determining the L D&8201 Ratio of Neutron Radiography Beams《测定中子射线照相束L D比率的标准试验方法》.pdf》由会员分享,可在线阅读,更多相关《ASTM E803-1991(2013) 1250 Standard Test Method for Determining the L D&8201 Ratio of Neutron Radiography Beams《测定中子射线照相束L D比率的标准试验方法》.pdf(7页珍藏版)》请在麦多课文档分享上搜索。
1、Designation: E803 91 (Reapproved 2013)Standard Test Method forDetermining the L/D Ratio of Neutron Radiography Beams1This standard is issued under the fixed designation E803; the number immediately following the designation indicates the year oforiginal adoption or, in the case of revision, the year
2、 of last revision. A number in parentheses indicates the year of last reapproval. Asuperscript epsilon () indicates an editorial change since the last revision or reapproval.1. Scope1.1 This test method defines an empirical technique for themeasurement of the effective collimation ratio, L/D, of neu
3、tronradiography beams. The technique is based upon analysis of aneutron radiographic image and is independent of measure-ments and calculations based on physical dimensions of thecollimator system. The values derived by this technique shouldbe more accurate than those based on physical measurements,
4、particularly for poorly defined apertures.2. Referenced Documents2.1 ASTM Standards:2E748 Practices for Thermal Neutron Radiography of Mate-rialsE1316 Terminology for Nondestructive Examinations3. Summary of Test Method3.1 Determination of neutron beam L/D ratio using the NU(no umbra) technique3is a
5、ccomplished by radiographing theNU device with the neutron beam to be measured andsubsequently analyzing the radiograph by one of three meth-ods. Each of the three methods is based upon the determinationof that point at which the umbral shadow width reaches zero.See Fig. 1. A key feature of the NU t
6、echnique is that L/D canbe determined accurately without the need for expensiveinstrumentation. Neutron radiography practices are discussedin Practices E748 and the terms are defined in TerminologyE1316.4. Significance and Use4.1 The quality of a neutron radiographic image is depen-dent upon many fa
7、ctors. The L/D ratio is one of those factorsand constitutes a numerical definition of the geometry of theneutron beam. The L/D ratio required for a specific neutronradiographic examination is dependent upon the thickness ofthe specimen and the physical characteristics of the particularelement of int
8、erest. Use of this test method allows theradiographer and the user to determine and periodically checkthe effective collimation ratio.5. Apparatus5.1 NU Device (see Fig. 2(a) and (b), and Fig. 3) employsneutron absorbing rods positioned at various distances from theimage plane. In practice this devi
9、ce consists of cadmium andnylon rods located in V-grooves accurately machined in thesurface of an aluminum channel section set at a 45 614 angleto the side support plate. Near the image plane end theV-grooves are machined on 0.283-cm centers. After 21 Vgrooves, counting one on the end, the grooves a
10、re machined on0.707-cm centers to the source end. The 0.64-mm diametercadmium and nylon rods are laid into the V-grooves andsecured with neutron transparent adhesive tape. The aluminumchannel is supported by side plates to maintain the 45 614 angle relative to the image plane. While cadmium rods wit
11、hdiameters other than 0.64 mm may be used, the exact roddiameter must be known and the depth of the V grooves mustbe adjusted accordingly.5.2 A single A unit as shown in Fig. 2(b) is used for L/Dvalues expected to be less than 150.Alternately, a singleAunitused with appropriate spacers may be used t
12、o accommodate awide range of L/D values.6. Procedure6.1 Place the NU device against the cassette with the finelyspaced rods nearest the cassette.6.2 Align the plane of the cassette perpendicular to the axisof the neutron beam.6.3 Expose the single-emulsion film and NU device for atime span that will
13、 produce a nominal background film densityof 2.5 6 0.4.6.4 Process the exposed film in accordance with the manu-facturers recommendations.6.5 Analyze the resultant image in accordance with one ormore of the three methods outlined in Section 7.1This test method is under the jurisdiction of Committee
14、E07 on NondestructiveTesting and is the direct responsibility of Subcommittee E07.05 on Radiology(Neutron) Method.Current edition approved June 1, 2013. Published June 2013. Originallyapproved in 1986. Last previous edition approved in 2008 as E803 91 (2008).DOI: 10.1520/E0803-91R13.2For referenced
15、ASTM standards, visit the ASTM website, www.astm.org, orcontact ASTM Customer Service at serviceastm.org. For Annual Book of ASTMStandards volume information, refer to the standards Document Summary page onthe ASTM website.3Newacheck, R. L., and Underhill, P. E., “The NU Method for Determining L/DRa
16、tio Of Neutron Radiography Facilities,” Aerotest Operations, Inc., Report A.O.77-27, June 1977.Copyright ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959. United States17. Data Analysis7.1 Visual AnalysisAvisual determination of the L/D ratiocan be made direct
17、ly from the neutron radiograph. Whenobserving the individual rod images, the umbral image can berecognized as the “white” line along the center of the rodimage. This “white” line will decrease in width for the rodslocated farther and farther from the film. At some point theumbral images will disappe
18、ar. Beyond this point a less intensewhite line will appear and increase in width with increasing roddistance. Use of a 5 to 10-power magnifier will aid indetermining the point at which the “white” line disappears andthen increases in width with a decreased intensity. Based on thevisual observation,
19、determine the rod with zero umbral widthand then determine its distance (b) from the cassette. The L/Dratio is as follows:L/D 5 b/rod diameter!7.2 Microdensitometric AnalysisThe second data analysismethod is based on a microdensitometric scan across thecadmium rod images beginning with the “0” posit
20、ion rodnearest the film. A typical scan is shown in Fig. 4.Adensitometer aperture of 20 300 m and no horizontalexpansion is suggested for this method. The value of b isobtained from the intersection of a straight line originatingfrom the tip (low film density) of the scan of the “0” rod anda curved
21、line through the tips of the remaining wave forms asshown in Fig. 4. This method gives the best results for L/Dratios up to a few hundred. Higher L/D ratios cannot bedetermined by this method due to the inability to obtain a stablewave form for large values of b.7.3 Alternative Microdensitometric An
22、alysisThis methodalso uses scanning microdensitometric traces for L/D ratiodeterminations and is applicable for both high and low L/Dratios. For this method the recommended microdensitometersettings are: 20 300-m aperture and 50 (or more) chartrecording expansion. These settings will produce individ
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
5000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ASTME803199120131250STANDARDTESTMETHODFORDETERMININGTHELD8201RATIOOFNEUTRONRADIOGRAPHYBEAMS 测定 中子 射线

链接地址:http://www.mydoc123.com/p-533561.html