ASTM E803-1991(2008) 752 Standard Test Method for Determining the L D Ratio of Neutron Radiography Beams《中子射线照相束的长度直径比的标准测定方法》.pdf
《ASTM E803-1991(2008) 752 Standard Test Method for Determining the L D Ratio of Neutron Radiography Beams《中子射线照相束的长度直径比的标准测定方法》.pdf》由会员分享,可在线阅读,更多相关《ASTM E803-1991(2008) 752 Standard Test Method for Determining the L D Ratio of Neutron Radiography Beams《中子射线照相束的长度直径比的标准测定方法》.pdf(7页珍藏版)》请在麦多课文档分享上搜索。
1、Designation: E 803 91 (Reapproved 2008)Standard Test Method forDetermining the L/D Ratio of Neutron Radiography Beams1This standard is issued under the fixed designation E 803; the number immediately following the designation indicates the year oforiginal adoption or, in the case of revision, the ye
2、ar of last revision. A number in parentheses indicates the year of last reapproval. Asuperscript epsilon () indicates an editorial change since the last revision or reapproval.1. Scope1.1 This test method defines an empirical technique for themeasurement of the effective collimation ratio, L/D, of n
3、eutronradiography beams. The technique is based upon analysis of aneutron radiographic image and is independent of measure-ments and calculations based on physical dimensions of thecollimator system. The values derived by this technique shouldbe more accurate than those based on physical measurement
4、s,particularly for poorly defined apertures.2. Referenced Documents2.1 ASTM Standards:2E 748 Practices for Thermal Neutron Radiography of Ma-terialsE 1316 Terminology for Nondestructive Examinations3. Summary of Test Method3.1 Determination of neutron beam L/D ratio using the NU(no umbra) technique3
5、is accomplished by radiographing theNU device with the neutron beam to be measured andsubsequently analyzing the radiograph by one of three meth-ods. Each of the three methods is based upon the determinationof that point at which the umbral shadow width reaches zero.See Fig. 1. A key feature of the
6、NU technique is that L/D canbe determined accurately without the need for expensiveinstrumentation. Neutron radiography practices are discussedin Practices E 748 and the terms are defined in TerminologyE 1316.4. Significance and Use4.1 The quality of a neutron radiographic image is depen-dent upon m
7、any factors. The L/D ratio is one of those factorsand constitutes a numerical definition of the geometry of theneutron beam. The L/D ratio required for a specific neutronradiographic examination is dependent upon the thickness ofthe specimen and the physical characteristics of the particularelement
8、of interest. Use of this test method allows theradiographer and the user to determine and periodically checkthe effective collimation ratio.5. Apparatus5.1 NU Device (see Figs. 2(a), 2(b), and 3) employs neutronabsorbing rods positioned at various distances from the imageplane. In practice this devi
9、ce consists of cadmium and nylonrods located in V-grooves accurately machined in the surface ofan aluminum channel section set at a 45 614 angle to the sidesupport plate. Near the image plane end the V-grooves aremachined on 0.283-cm centers. After 21 V grooves, countingone on the end, the grooves a
10、re machined on 0.707-cm centersto the source end. The 0.64-mm diameter cadmium and nylonrods are laid into the V-grooves and secured with neutrontransparent adhesive tape. The aluminum channel is supportedby side plates to maintain the 45 614 angle relative to theimage plane. While cadmium rods with
11、 diameters other than0.64 mm may be used, the exact rod diameter must be knownand the depth of the V grooves must be adjusted accordingly.5.2 A single A unit as shown in Fig. 2(b) is used for L/Dvalues expected to be less than 150. Alternately, a single A unitused with appropriate spacers may be use
12、d to accommodate awide range of L/D values.6. Procedure6.1 Place the NU device against the cassette with the finelyspaced rods nearest the cassette.6.2 Align the plane of the cassette perpendicular to the axisof the neutron beam.6.3 Expose the single-emulsion film and NU device for atime span that w
13、ill produce a nominal background film densityof 2.5 6 0.4.6.4 Process the exposed film in accordance with the manu-facturers recommendations.6.5 Analyze the resultant image in accordance with one ormore of the three methods outlined in Section 7.1This test method is under the jurisdiction of Committ
14、ee E07 on NondestructiveTesting and is the direct responsibility of Subcommittee E07.05 on Radiology(Neutron) Method.Current edition approved July 1, 2008. Published September 2008. Originallyapproved in 1986. Last previous edition approved in 2002 as E 803 91 (2002).2For referenced ASTM standards,
15、visit the ASTM website, www.astm.org, orcontact ASTM Customer Service at serviceastm.org. For Annual Book of ASTMStandards volume information, refer to the standards Document Summary page onthe ASTM website.3Newacheck, R. L., and Underhill, P. E., “The NU Method for Determining L/DRatio Of Neutron R
16、adiography Facilities,” Aerotest Operations, Inc., Report A.O.77-27, June 1977.1Copyright ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959, United States.2(a) Support Channel Subassembly with Rod SpacingFIG. 1 Diagram of Zero Umbra Image ConfigurationNOTE 1Rod
17、s at “A” positions are 1 cm each side of center line (22 ea.)NOTE 2Rods at “B” positions are 2 cm each side of center line (9 ea.)NOTE 3Rods at “C” positions are 2.5 cm each side of center line (1 ea.)NOTE 4All dimensions from base line to reduce accumulative errorsNOTE 5Rod arrangement shown for si
18、ngle system device. For an add-on device, to form a double system, extend the 11 spaces for 7.78 cm to 19spaces for 13.43 cm and eliminate the close spacing (20 for 5.65 cm)NOTE 6Rods held tightly in position with one layer of transparent tapeE 803 91 (2008)22(b) L/D Apparatus Assembly7. Data Analys
19、is7.1 Visual AnalysisAvisual determination of the L/D ratiocan be made directly from the neutron radiograph. Whenobserving the individual rod images, the umbral image can berecognized as the “white” line along the center of the rodimage. This “white” line will decrease in width for the rodslocated f
20、arther and farther from the film. At some point theumbral images will disappear. Beyond this point a less intensewhite line will appear and increase in width with increasing roddistance. Use of a 5 to 10-power magnifier will aid indetermining the point at which the “white” line disappears andthen in
21、creases in width with a decreased intensity. Based on thevisual observation, determine the rod with zero umbral widthand then determine its distance (b) from the cassette. The L/Dratio is as follows:L/D 5b/rod diameter!7.2 Microdensitometric AnalysisThe second data analysismethod is based on a micro
22、densitometric scan across thecadmium rod images beginning with the “0” position rodnearest the film. A typical scan is shown in Fig. 4.Adensitometer aperture of 20 3 300 m and no horizontalexpansion is suggested for this method. The value of b isobtained from the intersection of a straight line orig
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
5000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ASTME80319912008752STANDARDTESTMETHODFORDETERMININGTHELDRATIOOFNEUTRONRADIOGRAPHYBEAMS 中子 射线 照相 长度 直径

链接地址:http://www.mydoc123.com/p-533560.html