ASTM E685-1993(2013) 0678 Standard Practice for Testing Fixed-Wavelength Photometric Detectors Used in Liquid Chromatography《液相色谱法用固定波长光度探测装置试验的标准实施规程》.pdf
《ASTM E685-1993(2013) 0678 Standard Practice for Testing Fixed-Wavelength Photometric Detectors Used in Liquid Chromatography《液相色谱法用固定波长光度探测装置试验的标准实施规程》.pdf》由会员分享,可在线阅读,更多相关《ASTM E685-1993(2013) 0678 Standard Practice for Testing Fixed-Wavelength Photometric Detectors Used in Liquid Chromatography《液相色谱法用固定波长光度探测装置试验的标准实施规程》.pdf(7页珍藏版)》请在麦多课文档分享上搜索。
1、Designation: E685 93 (Reapproved 2013)Standard Practice forTesting Fixed-Wavelength Photometric Detectors Used inLiquid Chromatography1This standard is issued under the fixed designation E685; the number immediately following the designation indicates the year oforiginal adoption or, in the case of
2、revision, the year of last revision. A number in parentheses indicates the year of last reapproval. Asuperscript epsilon () indicates an editorial change since the last revision or reapproval.1. Scope1.1 This practice is intended to serve as a guide for thetesting of the performance of a photometric
3、 detector (PD) usedas the detection component of a liquid-chromatographic (LC)system operating at one or more fixed wavelengths in the range210 to 800 nm. Measurements are made at 254 nm, if possible,and are optional at other wavelengths.1.2 This practice is intended to describe the performance ofth
4、e detector both independently of the chromatographic system(static conditions) and with flowing solvent (dynamic condi-tions).1.3 For general liquid chromatographic procedures, consultRefs (1-9).21.4 For general information concerning the principles,construction, operation, and evaluation of liquid-
5、chromatography detectors, see Refs (10 and 11) in addition tothe sections devoted to detectors in Refs (1-7).1.5 This standard does not purport to address all of thesafety problems, if any, associated with its use. It is theresponsibility of the user of this standard to establish appro-priate safety
6、 and health practices and determine the applica-bility of regulatory limitations prior to use.2. Referenced Documents2.1 ASTM Standards:3E275 Practice for Describing and Measuring Performance ofUltraviolet and Visible SpectrophotometersE682 Practice for Liquid Chromatography Terms and Rela-tionships
7、3. Terminology3.1 Definitions:3.1.1 absorbance calibration, nthe procedure that verifiesthat the absorbance scale is correct within 65%.3.1.2 drift, nthe average slope of the noise envelopeexpressed in absorbance units per hour (AU/h) as measuredover a period of 1 h.3.1.3 dynamic, nunder conditions
8、of a flow rate of 1.0mL/min.3.1.4 linear range, n of a PD, the range of concentrationsof a test substance in a mobile phase over which the responseof the detector is constant to within 5 % as determined from thelinearity plot specified below and illustrated in Fig. 1. Thelinear range should be expre
9、ssed as the ratio of the highestconcentration to the minimum detectable concentration or thelowest linear concentration, whichever is greatest.3.1.5 long-term noise, nthe maximum amplitude in AUfor all random variations of the detector signal of frequenciesbetween 6 and 60 cycles per hour (0.1 and 1
10、.0 cycles per min).3.1.5.1 DiscussionIt represents noise that can be mistakenfor a late-eluting peak. This noise corresponds to the observednoise only and may not always be present.3.1.6 minimum detectability, nof a PD, that concentrationof a specific solute in a specific solvent that results in a d
11、etectorresponse corresponding to twice the static short-term noise.3.1.7 response time (speed of output), nthe detector, thetime required for the detector output to change from 10 to 90 %of the new equilibrium value when the composition of themobile phase is changed in a stepwise manner, within the
12、linearrange of the detector.3.1.7.1 DiscussionBecause the detector volume is verysmall and the transport rate is not diffusion dependent, theresponse time is generally fast enough to be unimportant. It isgenerally comparable to the response time of the recorder anddependent on the response time of t
13、he detector electrometerand on the recorder amplifier. Factors that affect the observedresponse time include the true detector response time, elec-tronic filtering, and system band-broadening.3.1.8 short-term noise, nthe maximum amplitude, peak topeak, in AU for all random variations of the detector
14、 signal ofa frequency greater than one cycle per minute.1This practice is under the jurisdiction of ASTM Committee E13 on MolecularSpectroscopy and Separation Science and is the direct responsibility of Subcom-mittee E13.19 on Separation Science.Current edition approved Jan. 1, 2013. Published Janua
15、ry 13. Originallyapproved in 1979. Last previous edition approved in 2005 as E685 93 (2005).DOI: 10.1520/E0685-93R13.2The boldface numbers in parentheses refer to the list of references at the end ofthis practice.3For referenced ASTM standards, visit the ASTM website, www.astm.org, orcontact ASTM Cu
16、stomer Service at serviceastm.org. For Annual Book of ASTMStandards volume information, refer to the standards Document Summary page onthe ASTM website.Copyright ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959. United States13.1.8.1 DiscussionIt determines th
17、e smallest signal detect-able by a PD, limits the precision attainable in quantitation oftrace-level samples, and sets the lower limit on linearity. Thisnoise corresponds to the observed noise only.3.1.9 static, nunder conditions of no flow.4. Significance and Use4.1 Although it is possible to obser
18、ve and measure each ofthe several characteristics of a detector under different andunique conditions, it is the intent of this practice that acomplete set of detector specifications should be obtainedunder the same operating conditions. It should also be notedthat to completely specify a detectors c
19、apability, its perfor-mance should be measured at several sets of conditions withinthe useful range of the detector. The terms and tests describedin this practice are sufficiently general that they may be usedregardless of the ultimate operating parameters.4.2 Linearity and response time of the reco
20、rder or otherreadout device used should be such that they do not distort orotherwise interfere with the performance of the detector. Thisrequires adjusting the gain, damping, and calibration in accor-dance with the manufacturers directions. If additional elec-tronic filters or amplifiers are used be
21、tween the detector and thefinal readout device, their characteristics should also first beestablished.5. Noise and Drift5.1 Test ConditionsPure, degassed methanol of suitablegrade4shall be used in the sample cell.Air or nitrogen shall beused in the reference cell if there is one. Nitrogen is preferr
22、edwhere the presence of high-voltage equipment makes it likelythat there is ozone in the air. Protect the entire system fromtemperature fluctuations because these will lead to detectabledrift.5.1.1 The detector should be located at the test site andturned on at least 24 h before the start of testing
23、. Insufficientwarm-up may result in drift in excess of the actual value for thedetector.5.2 Methods of Measurement:5.2.1 Connect a suitable device (Note 1) between the pumpand the detector to provide at least 75 kPa (500 psi) backpressure at 1.0 mL/min flow of methanol. Connect a shortlength (about
24、100 mm) of 0.25-mm (0.01-in.) internal-diameterstainless steel tubing to the outlet tube of the detector to retardbubble formation. Connect the recorder to the proper detectoroutput channels.NOTE 1Suggested devices include (a)2to4mof0.1-mm (0.004-in.)internal-diameter stainless steel tubing, (b) abo
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
5000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ASTME685199320130678STANDARDPRACTICEFORTESTINGFIXEDWAVELENGTHPHOTOMETRICDETECTORSUSEDINLIQUIDCHROMATOGRAPHY

链接地址:http://www.mydoc123.com/p-533312.html