ASTM E671-1998(2004) Standard Specification for Maximum Permissible Thermal Residual Stress in Annealed Glass Laboratory Apparatus《在退火的玻璃实验室器皿中最大容许热残留应力》.pdf
《ASTM E671-1998(2004) Standard Specification for Maximum Permissible Thermal Residual Stress in Annealed Glass Laboratory Apparatus《在退火的玻璃实验室器皿中最大容许热残留应力》.pdf》由会员分享,可在线阅读,更多相关《ASTM E671-1998(2004) Standard Specification for Maximum Permissible Thermal Residual Stress in Annealed Glass Laboratory Apparatus《在退火的玻璃实验室器皿中最大容许热残留应力》.pdf(3页珍藏版)》请在麦多课文档分享上搜索。
1、Designation: E 671 98 (Reapproved 2004)Standard Specification forMaximum Permissible Thermal Residual Stress in AnnealedGlass Laboratory Apparatus1This standard is issued under the fixed designation E 671; the number immediately following the designation indicates the year oforiginal adoption or, in
2、 the case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. Asuperscript epsilon (e) indicates an editorial change since the last revision or reapproval.This standard has been approved for use by agencies of the Department of Defense.1. Scope1.1 T
3、his specification covers a limit for thermal residualstress in reusable annealed glass laboratory apparatus asdetermined by prescribed photoelastic measurement proce-dures.1.2 In broad classification, the laboratory glassware itemscovered by this specification, but not limited to, are:beakers Imhoff
4、 conesbottles, aspirator impingersbottles, dropping jars, batterybottles, gas washing jars, bellbottles, infusion jars, chromatographybottles, milk test jars, cylindricalbottles, reagent joints, ball and socket or standard taperbottles, weighing manometersbulbs, absorption percolatorsbulbs, leveling
5、 pycnometersbulbs, sampling stopcocksburets tubes, centrifugecondensers tubes, chromatographycrystallizing dishes tubes, color comparison (turbidity)culture dishes tubes, combustion (ignition)custom apparatus tubes, connecting and adaptercylinders, graduated tubes, digestionand plain tubes, dryingde
6、siccators tubes, fermentationextraction tubes tubes, thistle (spray traps)flasks vapor trapsfritted ware viscometersfunnels watch glassesgenerators, Kippgrinder, tissue1.3 This specification recognizes that photoelastic measure-ments are proportional to the difference of the principalstresses. The l
7、imit imposed represents a safety factor to covera situation in which one of the principal stresses may be largerthan the apparent stress.1.4 This specification applies only to annealed glasswarethat is intended for sale as such. It excludes glassware that hasbeen thermally tempered, ion-exchanged, o
8、r laminated withglass layers of differing expansion. The intent of this specifi-cation is to limit the residual stresses for safe consumer use inannealed glass, as it leaves the manufacturer.1.5 Stresses introduced by thermal expansion differenceswithin the glassware are covered by this specificatio
9、n. Gradedand glass-to-metal seals are excluded.2. Referenced Documents2.1 ASTM Standards:2E 1157 Specification for Sampling and Testing of ReusableLaboratory GlasswareE 1273 Specification for Color Coding of Reusable Labo-ratory PipetsF 218 Test Method for Analyzing Stress in Glass3. Stress Limit3.1
10、 The stress as measured by the procedure in Section 4and calculated by Eq 2 shall not exceed 5.2 MPa (750 psi),except for combustion, centrifuge, and chromatography tubes,for which a limit of 4.5 MPa (650 psi) applies. Ware exceedingthese limits shall be rejected or reannealed to meet thespecificati
11、on.4. Measurement Procedure4.1 Using a Friedel (Senarmont) polarimeter as described inTest Method F 218, place the glass article to be measured in theviewing field in air. Examine every part of the article with adefinable light path (glass dimension) by rotating the analyzerto compensate for local s
12、tress birefringence. Document thosezones showing the higher values for the retardation or thick-ness ratios by recording analyzer angle, glass thickness (lightpath), and position in ware.4.2 In some orientations, such as sighting perpendicular tothe axis of a thin-walled cylinder, two glass paths mu
13、st beincluded in the measurement. If by rotating the cylinder, theretardation appears to be relatively constant, the measurement1This specification is under the jurisdiction of ASTM Committee E41 onLaboratory Apparatus and is the direct responsibility of Subcommittee E41.01 onApparatus.Current editi
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
5000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ASTME67119982004STANDARDSPECIFICATIONFORMAXIMUMPERMISSIBLETHERMALRESIDUALSTRESSINANNEALEDGLASSLABORATORYAPPARATUS

链接地址:http://www.mydoc123.com/p-533279.html