1、Designation: E 542 01Standard Practice forCalibration of Laboratory Volumetric Apparatus1This standard is issued under the fixed designation E 542; the number immediately following the designation indicates the year oforiginal adoption or, in the case of revision, the year of last revision. A number
2、 in parentheses indicates the year of last reapproval. Asuperscript epsilon (e) indicates an editorial change since the last revision or reapproval.This standard has been approved for use by agencies of the Department of Defense.1. Scope1.1 This practice covers procedures for use in the calibrationo
3、f volumetric ware, in accordance with Specification E694such as is in common use in chemical and clinical laboratories.It is based on the gravimetric determination of the quantity ofwater either contained or delivered, and the conversion of thisvalue to true volume at the standard temperature of 20C
4、 bymeans of suitable equations and standard tables. Calibrationusing mercury is excluded. Calibration may be performedusing alternative gravimetric methodology, provided that it isdemonstrated and documented that the results obtained areequivalent to those obtained using the methodology describedher
5、ein.1.2 This practice is intended to encompass capacity warebetween the limits of 0.1 cm3and 2000 cm3. Typical productsfalling within the purview of this practice are burets graduated“to deliver”, graduated cylinders, volumetric flasks, specificgravity flasks, measuring and dilution pipets, and tran
6、sfer andcapacity pipets.1.3 The procedures are not recommended for calibration ofapparatus with capacities below 0.1 cm3, such as microglass-ware.1.4 This standard does not purport to address all of thesafety concerns, if any, associated with its use. It is theresponsibility of the user of this stan
7、dard to establish appro-priate safety and health practices and determine the applica-bility of regulatory limitations prior to use.2. Referenced Documents2.1 ASTM Standards:E 1 Specification for ASTM Thermometers2E 694 Specification for Volumetric Ware3SI 10 Standard for Use of the International Sys
8、tem of Units(SI) (the Modernized Metric System)33. Summary of Practice3.1 This practice is based upon a determination of thevolume of water either contained in or delivered by the vessel.Procedures are given for cleaning, setting the meniscus,manipulating the apparatus, weighing, and converting thew
9、eight to the appropriate standard volume.4. Significance and Use4.1 The primary purpose of this practice is to provideuniform procedures that may be used to accurately calibrate awide variety of volumetric ware. The techniques are simple inconcept and can provide reliable results, provided the proce
10、-dures are followed faithfully. Accordingly, the practice shouldprovide a means for checking the original calibration ofglassware and similar apparatus and for periodic rechecks asthe need should arise.4.2 Borosilicate volumetric glassware will hold its calibra-tion indefinitely provided that it is
11、not exposed to hydrofluoricacid, hot phosphoric acid, or strong, hot alkalis, and that it isnot heated above 150C when dry. A frosting of the glasssurface (viewed when dry) indicates that chemical attack hasoccured, and recalibration may be in order. As a precaution,however, it is recommended that t
12、he glassware be recalibratedafter ten years of service regardless of its appearance.4.3 Soda-lime volumetric glassware will become frostedwith time because of attack from moisture in the atmosphere aswell as from the chemicals mentioned above. In addition, itshould not be heated above 90C when dry.
13、It is recommended,therefore, that it be recalibrated after five years of serviceunless frosting (viewed when dry) is observed sooner.5. Units of Measurement5.1 CapacityThe basic SI unit for volume is the cubicmetre, m3. Due to its large size, it is rarely used in volumetriccalibration. Rather, the c
14、ubic centimetre, cm3, is used and willbe employed in this practice. The unit, millilitre, mL, may beconsidered as equivalent to the cubic centimetre.5.2 Standard TemperatureVolumetric ware is almost uni-versally calibrated at 20C. The procedures described providefor such a calibration. When it is ne
15、cessary to work at higherambient temperatures, such as is the case in tropical countries,calibration may be required at the International StandardsOrganization (ISO) recommended temperature of 27C. Equa-tions are given to correct the calibrated volume to othertemperatures as this may be required or
16、desired.1This practice is under the jurisdiction of ASTM Committee E41 on LaboratoryApparatus, and is the direct responsibility of Subcommittee E41.01 on Apparatus.Current edition approved Octo. 10, 2001. Published January 2002. Originallypublished as E 542 79. Last previous edition E 542 00.2Annual
17、 Book of ASTM Standards, Vol 14.03.3Annual Book of ASTM Standards, Vol 14.04.1Copyright ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959, United States.6. Apparatus6.1 Balance, having sufficient capacity to weigh the loadedvessel. The sensitivity of the balanc
18、e will be a limiting factorin the accuracy of the measurements. Either a single-pan,direct-reading balance or an equal-arm balance of adequatesensitivity and capacity may be used. In either case the weightsmust be calibrated with adequate accuracy. Ordinarily, weightswith NBS Class S-1 tolerances ar
19、e required. The balance musthave physical dimensions to accept the size of the vesselswhich need to be weighed.6.2 Thermometer, for measuring the temperature of thewater. The accuracy of this calibration will depend upon theaccuracy requirement of the volumetric calibration (see Section14 for tolera
20、nces).6.3 Barometer, capable of providing atmospheric pressuremeasurements, consistent with the tolerances given in Section14. Alternatively, the existing barometric pressure may beobtained from the local weather service.6.4 Distilled or Deionized Water, suitable for general labo-ratory purposes.7.
21、General Considerations7.1 This section describes manipulative techniques requiredto obtain accurate and reproducible volumetric measurements.7.2 Cleaning Procedures:7.2.1 It is important that volumetric ware be thoroughlycleaned before being tested or used. Glass apparatus must besufficiently clean
22、to permit uniform wetting of the surface.When clean, the walls will be uniformly wetted and the waterwill adhere to the glass surface in a continuous film. The cleanwalls of some plastic apparatus, however, will not be wetted.(Follow the instructions of the manufacturer. Do not usematerials which wi
23、ll attack, discolor, or swell the plasticware.)Lack of cleanliness causes irregularities in capacity by distort-ing the water surface. The liquids usually used for cleaningglassware are sodium dichromatic-sulfuric acid cleaning solu-tion (commercially available from laboratory supply houses),nitric
24、acid, fuming sulfuric acid, alcohol, and water. The choiceof cleaning agent to be used depends on the nature of thecontaminant. After cleaning with the cleaning solution andthoroughly rinsing with tap water, the vessel should be rinsedwith distilled water.7.2.2 After cleaning, the vessel should be r
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
5000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ASTME5422001STANDARDPRACTICEFORCALIBRATIONOFLABORATORYVOLUMETRICAPPARATUS 试验室 量器 校准 标准 规程 PDF

链接地址:http://www.mydoc123.com/p-532997.html