《ASTM E508-2009 516 Standard Test Method for Determination of Calcium and Magnesium in Iron Ores by Flame Atomic Absorption Spectrometry《用火焰原子吸收光谱法测定铁矿石中钙和镁含量的标准试验方法》.pdf》由会员分享,可在线阅读,更多相关《ASTM E508-2009 516 Standard Test Method for Determination of Calcium and Magnesium in Iron Ores by Flame Atomic Absorption Spectrometry《用火焰原子吸收光谱法测定铁矿石中钙和镁含量的标准试验方法》.pdf(4页珍藏版)》请在麦多课文档分享上搜索。
1、Designation: E 508 09Standard Test Method forDetermination of Calcium and Magnesium in Iron Ores byFlame Atomic Absorption Spectrometry1This standard is issued under the fixed designation E 508; the number immediately following the designation indicates the year oforiginal adoption or, in the case o
2、f revision, the year of last revision. A number in parentheses indicates the year of last reapproval. Asuperscript epsilon () indicates an editorial change since the last revision or reapproval.1. Scope1.1 This test method covers the determination of calciumand magnesium in iron ores, concentrates,
3、and agglomerates inthe mass fraction (%) range from 0.05 % to 5 % of calcium and0.05 % to 3 % of magnesium.1.2 The values stated in SI units are to be regarded asstandard. No other units of measurement are included in thisstandard.1.3 This standard does not purport to address all of thesafety concer
4、ns, if any, associated with its use. It is theresponsibility of the user of this standard to establish appro-priate safety and health practices and determine the applica-bility of regulatory limitations prior to use.2. Referenced Documents2.1 ASTM Standards:2D 1193 Specification for Reagent WaterE50
5、 Practices for Apparatus, Reagents, and Safety Consid-erations for Chemical Analysis of Metals, Ores, andRelated MaterialsE 135 Terminology Relating to Analytical Chemistry forMetals, Ores, and Related MaterialsE 276 Test Method for Particle Size or Screen Analysis atNo. 4 (4.75-mm) Sieve and Finer
6、for Metal-Bearing Oresand Related MaterialsE 663 Practice for Flame Atomic Absorption Analysis3E 882 Guide for Accountability and Quality Control in theChemical Analysis Laboratory3. Terminology3.1 DefinitionsFor definitions of terms used in this testmethod, refer to Terminology E 135.4. Summary of
7、Test Method4.1 The sample is dissolved in HCl with the addition of asmall amount of HNO3. After evaporation and dehydration,any insoluble residue is filtered, treated for the recovery ofcalcium and magnesium, and added to the main solution. Thesolution is aspirated into the flame of the atomic absor
8、ptionspectrometer. For calcium, a nitrous oxide-acetylene burner isused, for magnesium, an air-acetylene burner is preferred (Note1). The absorbed energy at the resonance line of the spectrumof calcium or magnesium, or both, is measured by atomicabsorption spectrometry and compared with that of the
9、cali-bration solutions of the two elements.NOTE 1If a nitrous oxide burner is used for the determination ofmagnesium, the sensitivity is decreased by a factor of about three.If an air-acetylene burner is used for the determination of calcium,several elements, if present in a mass fraction (%) of 1.0
10、 or greater,interfere with the calcium determination. Also, lanthanum chloride(LaCl3) must be employed and the sensitivity is decreased by a factor ofabout two.5. Significance and Use5.1 This test method is intended as a referee method forcompliance with compositional specifications for impuritycont
11、ent. It is assumed that all who use this procedure will betrained analysts capable of performing common laboratorypractices skillfully and safely. It is expected that work will beperformed in a properly equipped laboratory and that properwaste disposal procedures will be followed. Follow appropriate
12、quality control practices such as those described in GuideE 882.1This test method is under the jurisdiction of ASTM Committee E01 onAnalytical Chemistry for Metals, Ores, and Related Materials and is the directresponsibility of Subcommittee E01.02 on Ores, Concentrates, and Related Metal-lurgical Ma
13、terials.Current edition approved May 1, 2009. Published June 2009. Originallyapproved in 1973. Last previous edition approved in 2008 as E 508 08.2For referenced ASTM standards, visit the ASTM website, www.astm.org, orcontact ASTM Customer Service at serviceastm.org. For Annual Book of ASTMStandards
14、 volume information, refer to the standards Document Summary page onthe ASTM website.3Withdrawn. The last approved version of this historical standard is referencedon www.astm.org.1Copyright ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959, United States.6. In
15、terferences6.1 None of the elements normally found in iron oresinterfere with this test method (Note 1).7. Apparatus7.1 Atomic Absorption Spectrometer, meeting the followingcriteria:7.1.1 Minimum SensitivityThe absorbance of the highestcalibration solution (see 8.4) must be at least 0.3.7.1.2 Curve-
16、LinearityThe difference between the read-ings of the highest calibration solutions must be more than 1.4times the difference between the readings for the zero solutionsand the lowest calibration solution.7.1.3 Minimum StabilityThe coefficient of variation of anumber of measurements of the highest ca
17、libration solutionand of the zero calibration solution must be less than 1.5 % and0.5 % respectively.NOTE 2Use of digital readout devices or printers, or both, is advis-able.NOTE 3Instrument parameters will vary with each instrument. Thefollowing parameters were successfully used in several laborato
18、ries andthey can be used as guidelines. In the case of calcium, solutions wereaspirated into a nitrous oxide-acetylene flame of a premix burner. Formagnesium an air-acetylene burner was used.Ca MgHollow cathode lamp, mA 15 15Wavelength, nm 422.7 285.2Nitrous oxide flow rate, L/min 13.8 .Air flow rat
19、e, L/min . 22Acetylene flow rate, L/min 6.6 4.28. Reagents and Materials8.1 Purity of ReagentsReagent grade chemicals shall beused in all tests. Unless otherwise indicated, it is intended thatall reagents conform to the specifications of the Committee onAnalytical Reagents of the American Chemical S
20、ociety wheresuch specifications are available.4Other grades may be used,provided it is first ascertained that the reagent is of sufficientlyhigh purity to permit its use without lessening the accuracy ofthe determination.8.2 Purity of WaterUnless otherwise indicated, referencesto water shall be unde
21、rstood to mean reagent water as definedby Type I of Specification D 1193.8.3 Calcium, Standard Solution (1 mL = 25 g Ca)Dissolve 1.2487 g of anhydrous calcium carbonate (CaCO3)in100 mL of dilute HCl (1 + 3). When solution is complete, cooland dilute to 1 L in a volumetric flask. Transfer 10 mL of th
22、issolution to a 200-mL volumetric flask. Dilute to volume andmix.8.4 Calcium and Magnesium Calibration SolutionsUsingpipets, transfer, (5.0, 10.0, 20.0, 40.0, and 50.0)-mL portions ofthe standard calcium and magnesium solutions to 200-mLvolumetric flasks (Note 4). Dilute to 90 mL and add 6 mL ofHCl,
23、 60 mL of background solution, and 40 mL of LaCl3solution. Dilute to volume and mix.8.5 Iron, Background Solution (10 g/L)Dissolve 10 g ofhigh-purity iron in 50 mL of HCl, and oxidize by the dropwiseaddition of HNO3. Evaporate to a syrupy consistency. Add20 mL of HCl and dilute to 200 mL with water.
24、 Add 20 g ofsodium chloride dissolved in 200 mL water and dilute to 1 Lwith water.8.6 Lanthanum Chloride Solution (50 g/L)Dissolve 50 gof LaCl3in 50 mL of HCl and 300 mL of hot water. Cool anddilute to 1 L.8.7 Magnesium, Standard Solution (1 mL = 17.5 g Mg)Dissolve 0.3500 g of high-purity magnesium
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
5000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ASTME5082009516STANDARDTESTMETHODFORDETERMINATIONOFCALCIUMANDMAGNESIUMINIRONORESBYFLAMEATOMICABSORPTIONSPECTROMETRY

链接地址:http://www.mydoc123.com/p-532914.html