ASTM E507-2013 red 0195 Standard Test Method for Determination of Aluminum in Iron Ores by Flame Atomic Absorption Spectrometry《采用火焰原子吸收光谱法测定铁矿中铝的标准试验方法》.pdf
《ASTM E507-2013 red 0195 Standard Test Method for Determination of Aluminum in Iron Ores by Flame Atomic Absorption Spectrometry《采用火焰原子吸收光谱法测定铁矿中铝的标准试验方法》.pdf》由会员分享,可在线阅读,更多相关《ASTM E507-2013 red 0195 Standard Test Method for Determination of Aluminum in Iron Ores by Flame Atomic Absorption Spectrometry《采用火焰原子吸收光谱法测定铁矿中铝的标准试验方法》.pdf(4页珍藏版)》请在麦多课文档分享上搜索。
1、Designation: E507 09E507 13Standard Test Method forDetermination of Aluminum in Iron Ores by Flame AtomicAbsorption Spectrometry1This standard is issued under the fixed designation E507; the number immediately following the designation indicates the year oforiginal adoption or, in the case of revisi
2、on, the year of last revision. A number in parentheses indicates the year of last reapproval. Asuperscript epsilon () indicates an editorial change since the last revision or reapproval.1. Scope1.1 This test method covers the determination of aluminum in iron ores, concentrates, and agglomerates in
3、the mass fraction(%) range from 0.1 to 5.1.2 The values stated in SI units are to be regarded as standard. No other units of measurement are included in this standard.1.3 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibilityof
4、 the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatorylimitations prior to use.2. Referenced Documents2.1 ASTM Standards:2D1193 Specification for Reagent WaterE50 Practices forApparatus, Reagents, and Safety Considerations for Ch
5、emicalAnalysis of Metals, Ores, and Related MaterialsE135 Terminology Relating to Analytical Chemistry for Metals, Ores, and Related MaterialsE276 Test Method for Particle Size or ScreenAnalysis at No. 4 (4.75-mm) Sieve and Finer for Metal-Bearing Ores and RelatedMaterialsE663 Practice for Flame Ato
6、mic Absorption Analysis (Withdrawn 1997)3E882 Guide for Accountability and Quality Control in the Chemical Analysis Laboratory3. Terminology3.1 DefinitionsFor definitions of terms used in this test method, refer to Terminology E135.4. Summary of Test Method4.1 The sample is dissolved in HCl with the
7、 addition of a small amount of HNO3. After evaporation and dehydration, anyinsoluble residue is filtered, treated for the recovery of aluminum, and added to the main solution. The solution is aspirated intothe flame of a nitrous oxide-acetylene burner. The absorbed energy at the resonance line of th
8、e spectrum of aluminum is measuredby atomic absorption spectrometry and compared with that of the calibration solutions.5. Significance and Use5.1 This test method is intended as a referee method for compliance with compositional specifications for impurity content. Itis assumed that all who use thi
9、s procedure will be trained analysts capable of performing common laboratory practices skillfullyand safely. It is expected that work will be performed in a properly equipped laboratory and that proper waste disposal procedureswill be followed. Follow appropriate quality control practices such as th
10、ose described in Guide E882.6. Interferences6.1 None of the elements normally found in iron ores interfere with this test method.1 This test method is under the jurisdiction of ASTM Committee E01 on Analytical Chemistry for Metals, Ores, and Related Materials and is the direct responsibility ofSubco
11、mmittee E01.02 on Ores, Concentrates, and Related Metallurgical Materials.Current edition approved Oct. 1, 2009Oct. 1, 2013. Published April 2010November 2013. Originally approved in 1973. Last previous edition approved in 20082009 asE507 08.E507 09. DOI: 10.1520/E0507-09.10.1520/E0507-13.2 For refe
12、rencedASTM standards, visit theASTM website, www.astm.org, or contactASTM Customer Service at serviceastm.org. For Annual Book of ASTM Standardsvolume information, refer to the standards Document Summary page on the ASTM website.3 The last approved version of this historical standard is referenced o
13、n www.astm.org.This document is not an ASTM standard and is intended only to provide the user of an ASTM standard an indication of what changes have been made to the previous version. Becauseit may not be technically possible to adequately depict all changes accurately, ASTM recommends that users co
14、nsult prior editions as appropriate. In all cases only the current versionof the standard as published by ASTM is to be considered the official document.Copyright ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959. United States17. Apparatus7.1 Atomic Absorption
15、 Spectrometer, meeting the following criteria:7.1.1 Minimum SensitivityThe absorbance of the highest calibration solution (see 8.4) must be at least 0.3.7.1.2 Curve-LinearityThe difference between the readings of the two highest calibration solutions must be more than 1.4 timesthe difference between
16、 the readings for the zero solution and the lowest calibration solution.7.1.3 Minimum StabilityThe coefficient of variation of a number of measurements of the highest calibration solution and ofthe zero calibration solution must be less than 1.5 % and 0.5 %, respectively.NOTE 1Use of digital readout
17、 devices or printers, or both, is advisable.NOTE 2Instrument parameters will vary with each instrument. The following parameters were successfully used in several laboratories and they canbe used as guidelines. Solutions were aspirated into a nitrous oxide-acetylene flame of a premix burner.Hollow c
18、athode lamp, mA 25Wavelength, nm 396.15Nitrous oxide flow rate, L/min 13.8Acetylene flow rate, L/min 6.68. Reagents and Materials8.1 Purity of ReagentsReagent grade chemicals shall be used in all tests. Unless otherwise indicated, it is intended that allreagents conform to the specifications of the
19、Committee on Analytical Reagents of the American Chemical Society where suchspecifications are available.4 Other grades may be used, provided it is first ascertained that the reagent is of sufficiently high purityto permit its use without lessening the accuracy of the determination.8.2 Purity of Wat
20、erUnless otherwise indicated, references to water shall be understood to mean reagent water as defined byType I of Specification D1193.8.3 Aluminum, Standard Solution (1 mL = 0.5 mg Al)Dissolve 0.5000 g of high-purity aluminum in 25 mL of HCl, cool,transfer to a 1-L volumetric flask, dilute to volum
21、e, and mix.8.4 Aluminum Calibration SolutionsTransfer (5.0, 10.0, 20.0, 40.0, and 50.0)-mL portions of the standard aluminum solutionto 200-mL volumetric flasks. Dilute to about 100 mL, add 6 mL of HCl and 60 mL of the background solution, dilute to volume,and mix. These calibration solutions contai
22、n (12.5, 25, 50, 100, and 125) g/mL Al, respectively.8.5 Iron, Background Solution (10 g/L)Dissolve 10 g of high-purity iron in 50 mL of HCl and oxidize by the dropwiseaddition of HNO3. Evaporate to a syrupy consistency. Add 20 mL of HCl and dilute to 200 mL with water. Add 20 g of NaCldissolved in
23、200 mL water and dilute to 1 L with water.8.6 Zero Calibration SolutionTransfer 60 mL of the background solution to a 200-mL volumetric flask, add 6 mL of HCl,dilute to volume, and mix.9. Hazards9.1 For precautions to be observed in this method, refer to Practice E50.10. Sample Preparation10.1 The a
24、nalytical sample shall be pulverized to pass a 150-m sieve (see Test Method E276).NOTE 3To facilitate decomposition, some ores, such as specular hematite, require grinding to pass a 75-m sieve.11. Procedure11.1 Transfer approximately 1.0 g of the sample to a small weighing bottle previously dried at
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
5000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ASTME5072013RED0195STANDARDTESTMETHODFORDETERMINATIONOFALUMINUMINIRONORESBYFLAMEATOMICABSORPTIONSPECTROMETRY

链接地址:http://www.mydoc123.com/p-532911.html