ASTM E499-1995(2006) Standard Test Methods for Leaks Using the Mass Spectrometer Leak Detector in the Detector Probe Mode《探测器探针方式用质谱仪检漏器检漏的试验方法》.pdf
《ASTM E499-1995(2006) Standard Test Methods for Leaks Using the Mass Spectrometer Leak Detector in the Detector Probe Mode《探测器探针方式用质谱仪检漏器检漏的试验方法》.pdf》由会员分享,可在线阅读,更多相关《ASTM E499-1995(2006) Standard Test Methods for Leaks Using the Mass Spectrometer Leak Detector in the Detector Probe Mode《探测器探针方式用质谱仪检漏器检漏的试验方法》.pdf(6页珍藏版)》请在麦多课文档分享上搜索。
1、Designation: E 499 95 (Reapproved 2006)Standard Test Methods forLeaks Using the Mass Spectrometer Leak Detector in theDetector Probe Mode1, 2This standard is issued under the fixed designation E 499; the number immediately following the designation indicates the year oforiginal adoption or, in the c
2、ase of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. Asuperscript epsilon (e) indicates an editorial change since the last revision or reapproval.This specification has been approved for use by agencies of the Department of Defense.1. Scope1.1 Th
3、ese test methods cover procedures for testing andlocating the sources of gas leaking at the rate of 4.5 3 1013mol/s (1 3 108Std cm3/s)3or greater. The test may beconducted on any device or component across which a pressuredifferential of helium or other suitable tracer gas may becreated, and on whic
4、h the effluent side of the leak to be testedis accessible for probing with the mass spectrometer samplingprobe.1.2 Two test methods are described:1.2.1 Test Method ADirect probing, and1.2.2 Test Method BAccumulation.1.3 This standard does not purport to address all of thesafety concerns, if any, ass
5、ociated with its use. It is theresponsibility of the user of this standard to establish appro-priate safety and health practices and determine the applica-bility of regulatory limitations prior to use.2. Referenced Documents2.1 ASTM Standards:4E 1316 Terminology for Nondestructive Examinations2.2 Ot
6、her Documents:SNT-TC-1A Recommended Practice for Personnel Qualifi-cation and Certification in Nondestructive Testing5ANSI/ASNT CP-189 ASNT Standard for Qualification andCertification of Nondestructive Testing Personnel53. Terminology3.1 DefinitionsFor definitions of terms used in this stan-dard, se
7、e Terminology E 1316, Section E.4. Summary of Test Methods4.1 Section 1.8 of the Leakage Testing Handbook6will be ofvalue to some users in determining which leak test method touse.4.2 These test methods require a leak detector with afull-scale readout of at least 4.5 3 1012mol/s (1 3 107Stdcm3/s)3on
8、 the most sensitive range, a maximum 1-min drift ofzero and sensitivity of 65 % of full scale on this range, and62 % or less on others (see 7.1). The above sensitivities arethose obtained by probing an actual standard leak in atmo-sphere with the detector, or sampling, probe, and not thesensitivity
9、of the detector to a standard leak attached directly tothe vacuum system.4.3 Test Method A, Direct Probing (see Fig. 1), is thesimplest test, and may be used in parts of any size, requiringonly that a tracer gas pressure be created across the area to betested, and the searching of the atmospheric si
10、de of the area bewith the detector probe. This test method detects leakage andits source or sources. Experience has shown that leak testingdown to 4.5 3 1011mol/s (1 3 106Std cm3/s)3in factoryenvironments will usually be satisfactory if reasonable precau-tions against releasing gas like the tracer g
11、as in the test area areobserved, and the effects of other interferences (Section 6) areconsidered.4.4 Test Method B, Accumulation Testing (see Fig. 2),provides for the testing of parts up to several cubic metres involume as in Fig. 2(a) or in portions of larger devices as in Fig.2(b). This is accomp
12、lished by allowing the leakage to accu-mulate in the chamber for a fixed period, while keeping it wellmixed with a fan, and then testing the internal atmosphere foran increase in tracer gas content with the detector probe. The1These test methods are under the jurisdiction of ASTM Committee E07 onNon
13、destructive Testing and are the direct responsibility of Subcommittee E07.08 onLeak Testing Method.Current edition approved Dec. 1, 2006. Published January 2007. Originallyapproved in 1973. Last previous edition approved in 2000 as E 499 - 95(2000).2(Atmospheric pressure external, pressure above atm
14、ospheric internal). Thisdocument covers the Detector Probe Mode described in Guide E 432.3The gas temperature is referenced to 0C. To convert to another gas referencetemperature, Tref, multiply the leak rate by (Tref+ 273)/273.4For referenced ASTM standards, visit the ASTM website, www.astm.org, orc
15、ontact ASTM Customer Service at serviceastm.org. For Annual Book of ASTMStandards volume information, refer to the standards Document Summary page onthe ASTM website.5Available fromAmerican Society for Nondestructive Testing (ASNT), P.O. Box28518, 1711 Arlingate Ln., Columbus, OH 43228-0518, http:/w
16、ww.asnt.org.6Marr, J. William, “Leakage Testing Handbook,” prepared for Liquid PropulsionSection, Jet Propulsion Laboratory, NationalAeronautics and SpaceAdministration,Pasadena, CA, Contract NAS 7-396, June 1961.1Copyright ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, P
17、A 19428-2959, United States.practical sensitivity attainable with this method depends pri-marily on two things: first, on the volume between the chamberand the object; and second, on the amount of outgassing oftracer gas produced by the object. Thus, a part having consid-erable exposed rubber, plast
18、ic, blind cavities or threads cannotbe tested with the sensitivity of a smooth metallic part. Thetime in which a leak can be detected is directly proportional tothe leak rate and inversely proportional to the volume betweenthe chamber and the part. In theory, extremely small leaks canbe detected by
19、this test method; however, the time required andthe effects of other interferences limit the practical sensitivityof this test method to about 4.5 3 1013mol/s (1 3 108Stdcm3/s)3for small parts.5. Personnel Qualification5.1 It is recommended that personnel performing leak test-ing attend a dedicated
20、training course on the subject and passa written examination. The training course should be appropri-ate for NDT level II qualification according to RecommendedPractice No. SNT-TC-1A of the American Society for Nonde-structive Testing or ANSI/ASNT Standard CP-189.6. Significance and Use6.1 Test Meth
21、od A is frequently used to test large systemsand complex piping installations that can be filled with a tracegas. Helium is normally used. The test method is used to locateleaks but cannot be used to quantify except for approximation.Care must be taken to provide sufficient ventilation to preventinc
22、reasing the helium background at the test site. Results arelimited by the helium background and the percentage of theleaking trace gas captured by the probe.6.2 Test Method B is used to increase the concentration oftrace gas coming through the leak by capturing it within anenclosure until the signal
23、 above the helium background can bedetected. By introducing a calibrated leak into the samevolume for a recorded time interval, leak rates can be mea-sured.FIG. 1 Method AE 499 95 (2006)27. Interferences7.1 Atmospheric HeliumThe atmosphere contains aboutfive parts per million (ppm) of helium, which
24、is beingcontinuously drawn in by the detector probe. This backgroundmust be “zeroed out” before leak testing using helium canproceed. Successful leak testing is contingent on the ability ofthe detector to discriminate between normal atmospherichelium, which is very constant, and an increase in heliu
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
5000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ASTME49919952006STANDARDTESTMETHODSFORLEAKSUSINGTHEMASSSPECTROMETERLEAKDETECTORINTHEDETECTORPROBEMODE

链接地址:http://www.mydoc123.com/p-532894.html