ASTM E2509-2014 Standard Test Method for Temperature Calibration of Rheometers in Isothermal Mode《等温流变仪温度校准的标准试验方法》.pdf
《ASTM E2509-2014 Standard Test Method for Temperature Calibration of Rheometers in Isothermal Mode《等温流变仪温度校准的标准试验方法》.pdf》由会员分享,可在线阅读,更多相关《ASTM E2509-2014 Standard Test Method for Temperature Calibration of Rheometers in Isothermal Mode《等温流变仪温度校准的标准试验方法》.pdf(5页珍藏版)》请在麦多课文档分享上搜索。
1、Designation: E2509 08E2509 14Standard Test Method forTemperature Calibration of Rheometers in Isothermal Mode1This standard is issued under the fixed designation E2509; the number immediately following the designation indicates the year oforiginal adoption or, in the case of revision, the year of la
2、st revision. A number in parentheses indicates the year of last reapproval. Asuperscript epsilon () indicates an editorial change since the last revision or reapproval.1. Scope1.1 This test method describes the temperature calibration or conformance of rheometers. The applicable temperature range is
3、0 to 80C however other ranges may be selected for the purpose at hand.1.2 The values stated in SI units are the standard. to be regarded as standard. No other units of measurement are included inthis standard.1.3 There are no ISO equivalents to this standard.1.4 This standard does not purport to add
4、ress all of the safety concerns, if any, associated with its use. It is the responsibilityof the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatorylimitations prior to use.2. Referenced Documents2.1 ASTM Standards:2E473 Terminolog
5、y Relating to Thermal Analysis and RheologyE1142 Terminology Relating to Thermophysical Properties3. Terminology3.1 DefinitionsSpecific technical terms found in this standard are defined in Terminologies E473 and E1142, includingrheometer and rheometryrheometry.4. Summary of Test Method4.1 An electr
6、onic thermometer of known characteristics is placed in the center of a dummy test specimen in contact with thetorque applying instrument plates of a rheometer at constant (isothermal) temperature. The difference between the rheometer settemperature and that indicated by the thermometer is used to ca
7、librate the rheometer temperature signal.5. Significance and Use5.1 Rheological properties such as viscosity and storage and loss modulus change rapidly with temperature. High qualitydeterminations of these properties depend upon a stable and well-known temperature of the measuring apparatus.6. Inte
8、rferences6.1 In many rheological experiments, temperature is varied with time. The calibration in this test method is made under stableand isothermal temperature conditions. Thus the effects of changes in temperature with time are not addressed. This isothermalcalibration does not provide any inform
9、ation about the specimen under temperature scanning conditions.7. Apparatus7.1 An Electronic Thermometerelectronic thermometer that includes:7.1.1 Temperature Sensor,sensor, (such as a thermocouple, platinum resistance thermometer, thermistor, etc.) with an accuracy(traceable to a known absolute sta
10、ndard) and resolution of 60.1C and a range of 0 to 80C.1 This test method is under the jurisdiction of ASTM Committee E37 on Thermal Measurements and is the direct responsibility of Subcommittee E37.08 on Rheology.Current edition approved Feb. 1, 2008Aug. 1, 2014. Published April 2008August 2014. Or
11、iginally approved in 2008. Last previous edition approved in 2008 as E2509 08.DOI: 10.1520/E2509-08.10.1520/E2509-14.2 For referencedASTM standards, visit theASTM website, www.astm.org, or contactASTM Customer Service at serviceastm.org. For Annual Book of ASTM Standardsvolume information, refer to
12、the standards Document Summary page on the ASTM website.This document is not an ASTM standard and is intended only to provide the user of an ASTM standard an indication of what changes have been made to the previous version. Becauseit may not be technically possible to adequately depict all changes
13、accurately, ASTM recommends that users consult prior editions as appropriate. In all cases only the current versionof the standard as published by ASTM is to be considered the official document.Copyright ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959. United
14、 States1NOTE 1Sensors with other temperature ranges may be used at the operators convenience.NOTE 2Some sensors are available already affixed with dummy test specimens from section 7.27.1.2 Temperature indicator, to convert the signal presented by the temperature sensor into a digital electronic tem
15、peraturedisplay with the accuracy and precision indicated in section 7.1.1.7.2 Dummy Test Specimen,test specimen, Two two polymer sheets each 1mm 1 mm in thickness of such a diameter to fill thespace (i.e. (that is, gap) between the instrument plates.NOTE 3The dummy test specimen may be composed of
16、the material to be tested or some other representative polymer material. Polydimethylsiloxane(PDMS) (e.g. (for example, “Silly Putty”3) may be used for this purpose.NOTE 4Polydimethylsiloxane may leave a residue of silicone oil on the surfaces of the instrument plates. This oil should be removed pri
17、or tosubsequent use.7.3 Rheometer, Thethe essential instrumentation required providing the minimum rheological analytical capabilities for this testmethod include:7.3.1 A drive actuator, to apply torque or displacement to the specimen in a periodic manner capable of frequencies ofoscillation from 0.
18、01 to 10 rad/s (0.0016 to 1.6 Hz). This actuator may also be capable of providing static force or transient stepor displacement of the test specimen.7.3.2 A coupling shaft, or other means to transmit the torque or displacement from the actuator to the specimen.7.3.3 A geometry, tools or plates, to f
19、ix the specimen between the coupling shaft and a stationary position. For the purposes ofthis test, parallel plates are the preferred configuration7.3.4 Either a torque sensor, to measure force developed by the specimen or a position sensor to measure the angulardisplacement , either one being capab
20、le of measuring within limits appropriate to the specimen and test being performed.7.3.5 A temperature sensor, to provide an indication of the specimen temperature readable to within 6 0.1 C.60.1C.7.3.6 A furnace or heating/cooling element, to provide controlled heating or cooling of a specimen to a
21、 constant temperatureconstant to within +/- 0.1 C 60.1C over the temperature range of interest.7.3.7 Atemperature controller, capable of executing a specific temperature program by operating the furnace or heating/coolingelement between selected temperature limits constant to within 60.1C.7.3.8 A st
22、ress or stain controller, capable of executing a specific unidirectional or oscillatory stress or strain program betweenselected stress or strain limits capable of controlling within limits appropriate to the specimen and test being performed.7.3.9 A recording data collecting device, capable of reco
23、rding all signals and displaying on the Y-axis any subset of themeasured signal (such as to provide a means of acquiring, storing, and displaying measured or calculated signals, or both. Theminimum output signals required include applied force, position or frequency)frequency or calculated signal (s
24、uch as viscosity,storage modulus, loss modulus, or tangent delta) using a linear or logarithmic scale as a function of any subset of and theindependent experimental parameterparameters (such as temperature, time, stress, strain, or frequency of oscillation.) on theX-axis. oscillation).7.3.10 Auxilia
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
5000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ASTME25092014STANDARDTESTMETHODFORTEMPERATURECALIBRATIONOFRHEOMETERSINISOTHERMALMODE 等温 流变 温度 校准 标准 试验

链接地址:http://www.mydoc123.com/p-531279.html