《ASTM E2465-2006 Standard Test Method for Analysis of Ni-Base Alloys by X-ray Fluorescence Spectrometry《用X射线荧光光谱测定法分析不锈钢和合金钢的标准试验方法》.pdf》由会员分享,可在线阅读,更多相关《ASTM E2465-2006 Standard Test Method for Analysis of Ni-Base Alloys by X-ray Fluorescence Spectrometry《用X射线荧光光谱测定法分析不锈钢和合金钢的标准试验方法》.pdf(10页珍藏版)》请在麦多课文档分享上搜索。
1、Designation: E 2465 06Standard Test Method forAnalysis of Ni-Base Alloys by X-ray FluorescenceSpectrometry1This standard is issued under the fixed designation E 2465; the number immediately following the designation indicates the year oforiginal adoption or, in the case of revision, the year of last
2、 revision. A number in parentheses indicates the year of last reapproval. Asuperscript epsilon (e) indicates an editorial change since the last revision or reapproval.1. Scope1.1 This test method covers the analysis of Ni-base alloysby wavelength dispersive X-ray Fluorescence Spectrometry forthe det
3、ermination of the following elements:Element Concentration RangeManganese 0.17 to 1.6 %Phosphorus 0.005 to 0.015 %Silicon 0.02 to 0.6 %Chromium 11 to 22 %Nickel 31 to 77 %Aluminum 0.12 to 1.3 %Molybdenum 0.045 to 10 %Copper 0.014 to 2.5 %Titanium 0.20 to 3.0 %Niobium 1.43 to 5.3 %Iron 2 to 46 %Tungs
4、ten 0.016 to 0.50 %Cobalt 0.014 to 0.35 %NOTE 1Unless exceptions are noted, concentration ranges can beextended by the use of suitable reference materials. Once these elementranges are extended they must be verified by some experimental means.This could include but not limited to Gage Repeatability
5、and Reproduc-ibility studies and/or Inter-laboratory Round Robin studies. Once thesestudies are completed, they will satisfy the ISO 17025 requirements forcapability.1.2 This standard does not purport to address all of thesafety concerns, if any, associated with its use. It is theresponsibility of t
6、he user of this standard to establish appro-priate safety and health practices and to determine theapplicability of regulatory limitations prior to use.2. Referenced Documents2.1 ASTM Standards:2E 135 Terminology Relating to Analytical Chemistry forMetals, Ores, and Related MaterialsE 305 Practice f
7、or Establishing and Controlling Spectro-chemical Analytical Curves3E 1361 Guide for Correction of Interelement Effects inX-Ray Spectrometric AnalysisE 1601 Practice for Conducting an Interlaboratory Study toEvaluate the Performance of an Analytical MethodE 1622 Practice for Correction of Spectral Li
8、ne Overlap inWavelength-Dispersive X-Ray Spectrometry32.2 Other Documents:ISO 17025 General requirements for the competence oftesting and calibration laboratories3. Terminology3.1 Definitions: For definitions of terms used in this testmethod, refer to Terminology E 135.4. Summary of Test Method4.1 T
9、he test specimen is finished to a clean, uniform surface,then irradiated with an X-ray beam of high energy. Thesecondary X-rays produced are dispersed by means of crystalsand the intensities are measured by suitable detectors atselected wavelengths. The outputs of the detectors in voltagepulses are
10、counted. Radiation measurements are made based onthe time required to reach a fixed number of counts, or on thetotal counts obtained for a fixed time (generally expressed incounts per unit time). Concentrations of the elements aredetermined by relating the measured radiation of unknownspecimens to a
11、nalytical curves prepared with suitable referencematerials. A fixed-channel, polychromator system or a sequen-tial, monochromator can be used for measurement of theelements.5. Significance and Use5.1 This procedure is suitable for manufacturing control andfor verifying that the product meets specifi
12、cations. It providesrapid, multi-element determinations with sufficient accuracy toassure product quality. The analytical performance data in-cluded may be used as a benchmark to determine if similar1This test method is under the jurisdiction of ASTM Committee E01 onAnalytical Chemistry for Metals,
13、Ores, and Related Materials and is the directresponsibility of Subcommittee E01.08 on Ni and Co and High Temperature Alloys.Current edition approved May 15, 2006. Published June 2006.2For referenced ASTM standards, visit the ASTM website, www.astm.org, orcontact ASTM Customer Service at serviceastm.
14、org. For Annual Book of ASTMStandards volume information, refer to the standards Document Summary page onthe ASTM website.3Withdrawn.1Copyright ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959, United States.X-ray spectrometers provide equivalent precision and
15、 accu-racy, or if the performance of a particular spectrometer haschanged.6. Interferences6.1 Interelement effects, or matrix effects, exist for some ofthe elements listed. Mathematical correction may be used tosolve for these elements. Various mathematical correctionprocedures are commonly utilized
16、. See Guide E 1361 andPractice E 1622. Any of these procedures that achieves ana-lytical accuracy equivalent to that provided by this test methodis acceptable.7. Apparatus7.1 Specimen Preparation Equipment:7.1.1 Surface Grinder or Sander With Abrasive Belts orDisks, or Lathe, capable of providing a
17、flat, uniform surface onthe reference materials and test specimens. Aluminum oxideand zirconium oxide belts and discs with a grit size of between60 and 180 have been found suitable.7.2 Excitation Source:7.2.1 Tube Power Supply, providing a constant potential orrectified power of sufficient energy to
18、 produce secondaryradiation of the specimen for the elements specified. Thegenerator may be equipped with a line voltage regulator andcurrent stabilizer.7.2.2 X-ray Tubes, with targets of various high-purity ele-ments that are capable of continuous operation at requiredpotentials and currents and th
19、at will excite the elements to bedetermined.7.3 Spectrometer, designed for X-ray emission analysis andequipped with specimen holders and a specimen chamber. Thechamber shall contain a specimen spinner, and must beequipped for vacuum or helium-flushed operation for thedetermination of elements of ato
20、mic number 20 (calcium) orlower.7.3.1 Analyzing Crystals, flat or curved crystals with opti-mized capability for the diffraction of the wavelengths ofinterest. The use of synthetic multilayer structures can also befound in some state-of-the-art-equipment.7.3.2 Collimators or Slits, for controlling t
21、he divergence ofthe characteristic X-rays. Use per the equipment manufactur-ers recommendations.7.3.3 Detectors, sealed or gas-flow proportional type, scin-tillation counters or equivalent.7.3.4 Vacuum System, providing for the determination ofelements whose radiation is absorbed by air (for example
22、,silicon, phosphorus, and sulfur). The system shall consist of avacuum pump, gage, and electrical controls to provide auto-matic pump down of the optical path, and maintain a controlledpressure, usually 13Pa (100 mm Hg) or less, controlled to 63Pa (20 mm Hg). A helium-flushed system is an alternativ
23、e to avacuum system.7.4 Measuring System, consisting of electronic circuits ca-pable of amplifying and integrating pulses received from thedetectors. For some measurements, a pulse height selector inconjunction with the detectors may be used to remove highorder lines and background. The system shall
24、 be equipped withan appropriate device.8. Reagents and Materials8.1 Detector Gas (P-10), consisting of a mixture of 90 %argon and 10 % methane, for use with gas-flow proportionalcounters only.9. Reference Materials9.1 Certified Reference Materials are available from na-tional metrology institutes, i
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
5000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ASTME24652006STANDARDTESTMETHODFORANALYSISOFNIBASEALLOYSBYXRAYFLUORESCENCESPECTROMETRY 射线 荧光 光谱 测定法 分析

链接地址:http://www.mydoc123.com/p-531181.html