ASTM E2206-2006 Standard Test Method for Force Calibration Of Thermomechnical Analyzers《热机械分析仪的力校验的标准方法》.pdf
《ASTM E2206-2006 Standard Test Method for Force Calibration Of Thermomechnical Analyzers《热机械分析仪的力校验的标准方法》.pdf》由会员分享,可在线阅读,更多相关《ASTM E2206-2006 Standard Test Method for Force Calibration Of Thermomechnical Analyzers《热机械分析仪的力校验的标准方法》.pdf(3页珍藏版)》请在麦多课文档分享上搜索。
1、Designation: E 2206 06Standard Test Method forForce Calibration Of Thermomechnical Analyzers1This standard is issued under the fixed designation E 2206; the number immediately following the designation indicates the year oforiginal adoption or, in the case of revision, the year of last revision. A n
2、umber in parentheses indicates the year of last reapproval. Asuperscript epsilon (e) indicates an editorial change since the last revision or reapproval.1. Scope1.1 This test method describes the calibration or perfor-mance confirmation of the electronically applied force signalfor thermomechanical
3、analyzers over the range of 0 to 1 N.1.2 SI units are the standard.1.3 There is no ISO method equivalent to this standard.1.4 This standard does not purport to address all of thesafety concerns, if any, associated with its use. It is theresponsibility of the user of this standard to establish appro-
4、priate safety and health practices and determine the applica-bility of regulatory limitations prior to use.2. Referenced Documents2.1 ASTM Standards:2E4 Practices for Force Verification of Testing MachinesE 473 Terminology Relating to Thermal Analysis and Rhe-ologyE 617 Specification for Laboratory
5、Weights and PrecisionMass StandardsE 831 Test Method for Linear Thermal Expansion of SolidMaterials by Thermomechanical AnalysisE 1142 Terminology Relating to Thermophysical PropertiesE 1363 Test Method for Temperature Calibration of Ther-momechanical AnalyzersE2113 Test Method for Length Change Cal
6、ibration ofThermomechanical Analyzers3. Terminology3.1 The technical terms used in this standard are defined inTerminologies E 473 and E 1142.4. Summary of Test Method4.1 The electronic force signal generated by a thermome-chanical analyzer is compared to that exerted by gravity on aknown mass. The
7、thermomechanical analyzer may be said tobe in conformance if the performance is within establishedlimits, typically 1 %. Alternatively, the force signal may becalibrated using a two-point calibration method.5. Significance and Use5.1 Most thermomechanical analysis experiments are car-ried out with s
8、ome force applied to the test specimen. Thisforce is often created electronically. It may be constant orchanged during the experiment.5.2 This method demonstrates conformance or calibrates theelectronically applied force signal.5.3 This method may be used for research and development,quality control
9、, manufacturing or regulatory applications.5.4 Other thermomechanical analyzer calibration functionsinclude temperature by Test Method E 1363 and length changeby Test Method E2113.6. Apparatus6.1 Thermomechanical AnalyzerThe essential instrumen-tation required to provide a minimum thermomechanical a
10、naly-sis or thermodilatometric capability for this method includes:6.1.1 Rigid Specimen Holder, inert, low expansivity mate-rial typically 0.6 m/(m K) to center the specimen in thefurnace and to fix the specimen to mechanical ground.NOTE 1Materials of construction with greater expansivity may beused
11、 but shall be reported.6.1.2 Rigid (Expansion or Compression) Probe, inert, lowexpansivity material typically 0.6 m/(m K) which con-tacts the specimen with an applied compressive force (see Note1).6.1.3 Sensing Element, linear over a minimum range of2 mm to measure the displacement of the rigid prob
12、e to 6 1mresulting from changes in length of the specimen.6.1.4 Programmable Force Transducer, to generate a con-stant force (6 1.0 %) of up to 1.0 N that is applied through therigid probe to the specimen.NOTE 2Other force ranges may be used but shall be reported.6.1.5 Furnace, capable of providing
13、uniform controlledheating (cooling) of the specimen to a constant temperature orat a constant rate within the temperature range of 100 to 600C.1This test method is under the jurisdiction ofASTM Committee E37 on ThermalMeasurements and is the direct responsibility of Subcommittee E37.01 on ThermalTes
14、t Methods and Practices.Current edition approved March 1, 2006. Published April 2006. Originallyapproved in 2002. Last previous edition approved in 2002 as E 2206 02.2For referenced ASTM standards, visit the ASTM website, www.astm.org, orcontact ASTM Customer Service at serviceastm.org. For Annual B
15、ook of ASTMStandards volume information, refer to the standards Document Summary page onthe ASTM website.1Copyright ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959, United States.NOTE 3Other temperature ranges may be used but shall be reported.6.1.6 Temperatu
16、re Controller, capable of executing a spe-cific temperature program by operating the furnace betweenselected temperature limits at a rate of change of up to 10C/min constant to 0.1 C/min or at an isothermal temperatureconstant to 6 0.5 C.NOTE 4Other heating rates may be used but shall be reported.6.
17、1.7 Temperature Sensor, that can be attached to, in contactwith, or reproducibly positioned in close proximity to thespecimen to provide an indication of the specimen temperatureto 6 0.1 C.6.1.8 A means of sustaining an environment around thespecimen of inert purge gas with a purge gas rate of 10 to
18、 1006 5 mL/min.NOTE 5Typically, 99.9+ % pure nitrogen, argon, or helium is em-ployed when oxidation in air is a concern. Unless effects of moisture areto be studied, use of dry purge gas is recommended and is essential foroperation at subambient temperatures.6.1.9 Recording Device, capable of record
19、ing and display-ing any fraction of the specimen dimension (TMA curve) orforce, including signal noise, on the Y-axis as a function oftemperature or time, including signal noise, on the X-axis.6.2 50 to 100 g 6 0.002 % Class 4 or better mass (traceableto a national reference laboratory) in complianc
20、e with Speci-fication E 617.7. Calibration7.1 Prepare the thermomechanical analyzer for operationaccording to procedures recommended by the manufacturer ofthe thermomechanical analyzer as described in the OperationsManual.7.2 Other calibration procedures which may be used, butwhich are not required
21、in this standard include Test MethodsE 1363, E 831 and E2113.8. Procedure8.1 With no specimen present, lower the probe so that itcontacts the specimen holder. Zero the device so that no force(load) is applied by the probe to the specimen holder.NOTE 6The means for determining “no load” condition is
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
5000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ASTME22062006STANDARDTESTMETHODFORFORCECALIBRATIONOFTHERMOMECHNICALANALYZERS 机械 分析 校验 标准 方法 PDF

链接地址:http://www.mydoc123.com/p-530554.html