ASTM E2108-2016 Standard Practice for Calibration of the Electron Binding-Energy Scale of an X-Ray Photoelectron Spectrometer《X-射线光电分光仪电子结合能刻度表的校准标准实施规程》.pdf
《ASTM E2108-2016 Standard Practice for Calibration of the Electron Binding-Energy Scale of an X-Ray Photoelectron Spectrometer《X-射线光电分光仪电子结合能刻度表的校准标准实施规程》.pdf》由会员分享,可在线阅读,更多相关《ASTM E2108-2016 Standard Practice for Calibration of the Electron Binding-Energy Scale of an X-Ray Photoelectron Spectrometer《X-射线光电分光仪电子结合能刻度表的校准标准实施规程》.pdf(22页珍藏版)》请在麦多课文档分享上搜索。
1、Designation: E2108 10E2108 16Standard Practice forCalibration of the Electron Binding-Energy Scale of anX-Ray Photoelectron Spectrometer1This standard is issued under the fixed designation E2108; the number immediately following the designation indicates the year oforiginal adoption or, in the case
2、of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. Asuperscript epsilon () indicates an editorial change since the last revision or reapproval.1. Scope1.1 This practice describes a procedure for calibrating the electron binding-energy (BE) scale of
3、 an X-ray photoelectronspectrometer that is to be used for performing spectroscopic analysis of photoelectrons excited by unmonochromated aluminumor magnesium K X-rays or by monochromated aluminum K X-rays.1.2 The calibration of the BE scale is recommended after the instrument is installed or modifi
4、ed in any substantive way.Additional checks and, if necessary, recalibrations are recommended at intervals chosen to ensure that BE measurements arestatistically unlikely to be made with an uncertainty greater than a tolerance limit, specified by the analyst, based on theinstrumental stability and t
5、he analystsanalysts needs. Information is provided by which the analyst can select an appropriatetolerance limit for the BE measurements and the frequency of calibration checks.1.3 This practice is based on the assumption that the BE scale of the spectrometer is sufficiently close to linear to allow
6、 forcalibration by measurements of reference photoelectron lines having BEs near the extremes of the working BE scale. In mostcommercial instruments, X-ray sources with aluminum or magnesium anodes are employed and BEs are typically measured at leastover the 0100001200 eV range. This practice can be
7、 used for the BE range from 0 eV to 1040 eV.1.4 The assumption that the BE scale is linear is checked by a measurement made with a reference photoelectron line orAuger-electron line that appears at an intermediate position. A single check is a necessary but not sufficient condition forestablishing l
8、inearity of the BE scale. Additional checks can be made with specified reference lines on instruments equipped withmagnesium or unmonochromated aluminum X-ray sources, with secondary BE standards, or by following the procedures of theinstrument manufacturer. Deviations from BE-scale linearity can oc
9、cur because of mechanical misalignments, excessive magneticfields in the region of the analyzer, or imperfections or malfunctions in the power supplies. This practice does not check for, noridentify, problems of this type.type but simply verifies the linearity of the BE scale.1.5 After an initial ch
10、eck of the BE-scale linearity and measurements of the repeatability standard deviation for the maincalibration lines for a particular instrument, a simplified procedure is given for routine checks of the calibration at subsequenttimes.1.6 This practice is recommended for use with X-ray photoelectron
11、 spectrometers operated in the constant-pass-energy orfixed-analyzer-transmission mode and for which the pass energy is less than 200 eV; otherwise, depending on the configurationof the instrument, a relativistic equation could be needed for the calibration. The practice should not be used for instr
12、umentsoperated in the constant-retardation-ratio mode at retardation ratios less than 10, for instruments with an energy resolution above1.5 eV, or in applications for which BE measurements are desired with tolerance limits of 60.03 eV or less.1.7 On instruments equipped with a monochromated aluminu
13、m K X-ray source, a measurement of the position of a specifiedAuger-electron line can be used, if desired, to determine the average energy of the X-rays incident on the specimen. Thisinformation is needed for the determination of modified Auger parameters.1.8 The values stated in SI units are to be
14、regarded as standard. No other units of measurement are included in this standard.1.9 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibilityof the user of this standard to establish appropriate safety and health practices and d
15、etermine the applicability of regulatorylimitations prior to use.1 This practice is under the jurisdiction of ASTM Committee E42 on Surface Analysis and is the direct responsibility of Subcommittee E42.03 on Auger ElectronSpectroscopy and X-Ray Photoelectron Spectroscopy.Current edition approved Nov
16、. 1, 2010Nov. 1, 2016. Published December 2010December 2016. Originally approved in 2000. Last previous edition approved in 20052010as E2108 05.E2108 10. DOI: 10.1520/E2108-10.10.1520/E2108-16.This document is not an ASTM standard and is intended only to provide the user of an ASTM standard an indic
17、ation of what changes have been made to the previous version. Becauseit may not be technically possible to adequately depict all changes accurately, ASTM recommends that users consult prior editions as appropriate. In all cases only the current versionof the standard as published by ASTM is to be co
18、nsidered the official document.Copyright ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959. United States12. Referenced Documents2.1 ASTM Standards:2E456 Terminology Relating to Quality and StatisticsE673 Terminology Relating to Surface Analysis (Withdrawn 2012
19、)3E902 Practice for Checking the Operating Characteristics of X-Ray Photoelectron Spectrometers (Withdrawn 2011)3E1016 Guide for Literature Describing Properties of Electrostatic Electron SpectrometersE1078 Guide for Specimen Preparation and Mounting in Surface AnalysisE1523 Guide to Charge Control
20、and Charge Referencing Techniques in X-Ray Photoelectron Spectroscopy2.2 ISO Standards:4ISO 9001:20009001:2015 Quality Management SystemsRequirements management systemsRequirementsISO 15472:200115472:2010 Surface Chemical AnalysisX-Ray Photoelectron SpectrometersCalibration of Energy Scale-schemical
21、 analysisX-ray photoelectron spectrometersCalibration of energy scalesISO 18115:200118115-1:2013 Surface Chemical AnalysisVocabulary chemical analysisVocabularyPart 1: General termsand terms used in spectroscopy3. Terminology3.1 DefinitionsFor definitions ofSince Terminology E673 terms used in X-ray
22、 photoelectron spectroscopy and surfaceanalysis, see Terminologywas withdrawn in 2012, definitions of terms used in Auger and E673 and ISO 18115:2001.X-rayphotoelectron spectroscopy are now based on ISO 18115-1:2013.5 For definitions of terms used in statistics, see TerminologyE456.3.2 Symbols and A
23、bbreviationsTable 1 shows definitions of the symbols and abbreviations used in this practice.4. Summary of Practice4.1 A procedure is given for calibrating the BE scale of an X-ray photoelectron spectrometer equipped with one or more of thefollowing sources of characteristic K X-rays: magnesium (Mg)
24、 source; unmonochromated aluminum (Al) source; ormonochromated Al source. This procedure is based on ISO 15472:2001.15472:2010. In a first calibration for particular operatingconditions of the instrument, or after the instrument has been modified, measurements are made of the BEs of specified core l
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
5000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ASTME21082016STANDARDPRACTICEFORCALIBRATIONOFTHEELECTRONBINDINGENERGYSCALEOFANXRAYPHOTOELECTRONSPECTROMETERX

链接地址:http://www.mydoc123.com/p-530297.html