ASTM E2108-2010 Standard Practice for Calibration of the Electron Binding-Energy Scale of an X-Ray Photoelectron Spectrometer《用X射线光电子分光计测定电子能量捆绑刻度的标准实施规程》.pdf
《ASTM E2108-2010 Standard Practice for Calibration of the Electron Binding-Energy Scale of an X-Ray Photoelectron Spectrometer《用X射线光电子分光计测定电子能量捆绑刻度的标准实施规程》.pdf》由会员分享,可在线阅读,更多相关《ASTM E2108-2010 Standard Practice for Calibration of the Electron Binding-Energy Scale of an X-Ray Photoelectron Spectrometer《用X射线光电子分光计测定电子能量捆绑刻度的标准实施规程》.pdf(16页珍藏版)》请在麦多课文档分享上搜索。
1、Designation: E2108 10Standard Practice forCalibration of the Electron Binding-Energy Scale of anX-Ray Photoelectron Spectrometer1This standard is issued under the fixed designation E2108; the number immediately following the designation indicates the year oforiginal adoption or, in the case of revis
2、ion, the year of last revision. A number in parentheses indicates the year of last reapproval. Asuperscript epsilon () indicates an editorial change since the last revision or reapproval.1. Scope1.1 This practice describes a procedure for calibrating theelectron binding-energy (BE) scale of an X-ray
3、 photoelectronspectrometer that is to be used for performing spectroscopicanalysis of photoelectrons excited by unmonochromated alu-minum or magnesium Ka X-rays or by monochromatedaluminum Ka X-rays.1.2 The calibration of the BE scale is recommended afterthe instrument is installed or modified in an
4、y substantive way.Additional checks and, if necessary, recalibrations are recom-mended at intervals chosen to ensure that BE measurements arestatistically unlikely to be made with an uncertainty greaterthan a tolerance limit, specified by the analyst, based on theinstrumental stability and the analy
5、sts needs. Information isprovided by which the analyst can select an appropriatetolerance limit for the BE measurements and the frequency ofcalibration checks.1.3 This practice is based on the assumption that the BEscale of the spectrometer is sufficiently close to linear to allowfor calibration by
6、measurements of reference photoelectronlines having BEs near the extremes of the working BE scale. Inmost commercial instruments, X-ray sources with aluminum ormagnesium anodes are employed and BEs are typically mea-sured over the 01000 eV range. This practice can be used forthe BE range from 0 eV t
7、o 1040 eV.1.4 The assumption that the BE scale is linear is checked bya measurement made with a reference photoelectron line orAuger-electron line that appears at an intermediate position. Asingle check is a necessary but not sufficient condition forestablishing linearity of the BE scale.Additional
8、checks can bemade with specified reference lines on instruments equippedwith magnesium or unmonochromated aluminum X-raysources, with secondary BE standards, or by following theprocedures of the instrument manufacturer. Deviations fromBE-scale linearity can occur because of mechanical misalign-ments
9、, excessive magnetic fields in the region of the analyzer,or imperfections or malfunctions in the power supplies. Thispractice does not check for, nor identify, problems of this type.1.5 After an initial check of the BE-scale linearity andmeasurements of the repeatability standard deviation for them
10、ain calibration lines for a particular instrument, a simplifiedprocedure is given for routine checks of the calibration atsubsequent times.1.6 This practice is recommended for use with X-rayphotoelectron spectrometers operated in the constant-pass-energy or fixed-analyzer-transmission mode and for w
11、hich thepass energy is less than 200 eV; otherwise, depending on theconfiguration of the instrument, a relativistic equation could beneeded for the calibration. The practice should not be used forinstruments operated in the constant-retardation-ratio mode atretardation ratios less than 10, for instr
12、uments with an energyresolution above 1.5 eV, or in applications for which BEmeasurements are desired with tolerance limits of 60.03 eV orless.1.7 On instruments equipped with a monochromated alumi-num Ka X-ray source, a measurement of the position of aspecified Auger-electron line can be used, if d
13、esired, todetermine the average energy of the X-rays incident on thespecimen. This information is needed for the determination ofmodified Auger parameters.1.8 The values stated in SI units are to be regarded asstandard. No other units of measurement are included in thisstandard.1.9 This standard doe
14、s not purport to address all of thesafety concerns, if any, associated with its use. It is theresponsibility of the user of this standard to establish appro-priate safety and health practices and determine the applica-bility of regulatory limitations prior to use.2. Referenced Documents2.1 ASTM Stan
15、dards:2E456 Terminology Relating to Quality and StatisticsE673 Terminology Relating to Surface AnalysisE902 Practice for Checking the Operating Characteristics of1This practice is under the jurisdiction of ASTM Committee E42 on SurfaceAnalysis and is the direct responsibility of Subcommittee E42.03
16、onAuger ElectronSpectroscopy and X-Ray Photoelectron Spectroscopy.Current edition approved Nov. 1, 2010. Published December 2010. Originallyapproved in 2000. Last previous edition approved in 2005 as E2108 05. DOI:10.1520/E2108-10.2For referenced ASTM standards, visit the ASTM website, www.astm.org,
17、 orcontact ASTM Customer Service at serviceastm.org. For Annual Book of ASTMStandards volume information, refer to the standards Document Summary page onthe ASTM website.1Copyright ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959, United States.X-Ray Photoelec
18、tron SpectrometersE1016 Guide for Literature Describing Properties of Elec-trostatic Electron SpectrometersE1078 Guide for Specimen Preparation and Mounting inSurface AnalysisE1523 Guide to Charge Control and Charge ReferencingTechniques in X-Ray Photoelectron Spectroscopy2.2 ISO Standards:3ISO 9001
19、:2000 Quality Management SystemsRequirementsISO 15472:2001 Surface Chemical AnalysisX-Ray Pho-toelectron SpectrometersCalibration of Energy ScalesISO 18115:2001 Surface Chemical AnalysisVocabulary3. Terminology3.1 DefinitionsFor definitions of terms used in X-rayphotoelectron spectroscopy and surfac
20、e analysis, see Terminol-ogy E673 and ISO 18115:2001. For definitions of terms used instatistics, see Terminology E456.3.2 Symbols and AbbreviationsTable 1 shows definitionsof the symbols and abbreviations used in this practice.4. Summary of Practice4.1 A procedure is given for calibrating the BE sc
21、ale of anX-ray photoelectron spectrometer equipped with one or moreof the following sources of characteristic Ka X-rays: magne-sium (Mg) source; unmonochromated aluminum (Al) source;3Available from American National Standards Institute (ANSI), 25 W. 43rd St.,4th Floor, New York, NY 10036, http:/www.
22、ansi.org.TABLE 1 Definitions of Symbols and Abbreviationsa measured energy scaling errorBE binding energy, in eVb measured zero offset error, in eVcinumber of counts in the i-th channeleV electron voltsEcorrcorrected result for the binding energy corresponding to a given Emeas,ineVEelembinding energ
23、y of a frequently measured element at which the indicated binding energy scale is set, aftercalibration, to read correctly, in eVEKkinetic energy of a peak, with reference to the Fermi level, in eVEmeasa measured binding energy, in eVEmeas naverage of the measured binding energies for the peak, n,in
24、Table 3,ineVEmeas nione of a set of measurements of binding energy for the peak, n,inTable 3,ineVEref nreference values for the position of peak, n,inTable 3, on the binding energy scale, in eVEppeak binding energy, in eVE0binding energy for first data channel at lower binding energy than the channe
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
5000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ASTME21082010STANDARDPRACTICEFORCALIBRATIONOFTHEELECTRONBINDINGENERGYSCALEOFANXRAYPHOTOELECTRONSPECTROMETER

链接地址:http://www.mydoc123.com/p-530296.html