ASTM E2108-2005 Standard Practice for Calibration of the Electron Binding-Energy Scale of an X-Ray Photoelectron Spectrometer《X-射线光电分光仪的电子结合能刻度表校准的标准实施规程》.pdf
《ASTM E2108-2005 Standard Practice for Calibration of the Electron Binding-Energy Scale of an X-Ray Photoelectron Spectrometer《X-射线光电分光仪的电子结合能刻度表校准的标准实施规程》.pdf》由会员分享,可在线阅读,更多相关《ASTM E2108-2005 Standard Practice for Calibration of the Electron Binding-Energy Scale of an X-Ray Photoelectron Spectrometer《X-射线光电分光仪的电子结合能刻度表校准的标准实施规程》.pdf(16页珍藏版)》请在麦多课文档分享上搜索。
1、Designation: E 2108 05Standard Practice forCalibration of the Electron Binding-Energy Scale of anX-Ray Photoelectron Spectrometer1This standard is issued under the fixed designation E 2108; the number immediately following the designation indicates the year oforiginal adoption or, in the case of rev
2、ision, the year of last revision. A number in parentheses indicates the year of last reapproval. Asuperscript epsilon (e) indicates an editorial change since the last revision or reapproval.1. Scope1.1 This practice describes a procedure for calibrating theelectron binding-energy (BE) scale of an X-
3、ray photoelectronspectrometer that is to be used for surface analysis withunmonochromated aluminum or magnesium Ka X-rays ormonochromated aluminum Ka X rays.1.2 It is recommended that the BE scale be calibrated afterthe instrument is installed or modified in any substantive way.Also, it is recommend
4、ed that the instrumental BE scale bechecked, and if necessary, recalibrated at intervals chosen toensure that BE measurements are statistically unlikely to bemade with greater uncertainty than a tolerance limit, specifiedby the analyst, based on the instrumental stability and theanalysts needs. Info
5、rmation is provided by which an analystcan select an appropriate tolerance limit for the BE measure-ments and the frequency of calibration checks.1.3 This practice is based on the assumption that the BEscale of the spectrometer is sufficiently close to linear that theBE scale can be calibrated by me
6、asurements of referencephotoelectron lines made near the extremes of the working BEscale. In most commercial instruments, X-ray sources withaluminum or magnesium anodes are employed and BEs aretypically measured over the 01000 eV range. This practicecan be used for the BE range from 0 eV to 1040 eV.
7、1.4 The assumption that the BE scale is linear is checked bya measurement made with a reference photoelectron line orAuger-electron line that appears at an intermediate position. Asingle check is a necessary but not sufficient condition forestablishing linearity of the BE scale.Additional checks can
8、 bemade with specified reference lines on instruments equippedwith magnesium or unmonochromated aluminum X-raysources, with secondary BE standards, or by following theprocedures of the instrument manufacturer. Deviations fromBE-scale linearity can occur because of mechanical misalign-ments, excessiv
9、e magnetic fields in the region of the analyzer,or imperfections or malfunctions in the power supplies. Thispractice does not check for, nor identify, problems of this type.1.5 After an initial check of the BE-scale linearity andmeasurements of the repeatability standard deviation for themain calibr
10、ation lines for a particular instrument, a simplifiedprocedure is given for routine checks of the calibration atsubsequent times.1.6 This practice is recommended for use with X-rayphotoelectron spectrometers operated in the constant-pass-energy or fixed-analyzer-transmission mode and for which thepa
11、ss energy is less than 200 eV; otherwise, depending on theconfiguration of the instrument, a relativistic equation could beneeded for the calibration equation. The practice should not beused for instruments operated in the constant-retardation-ratiomode at retardation ratios less than 10, for instru
12、ments with anenergy resolution worse than 1.5 eV, or in applications forwhich BE measurements are desired with tolerance limits of 60.03 eV or less.1.7 On instruments equipped with a monochromated alumi-num Ka X-ray source, a measurement of the position of aspecified Auger-electron line can be used,
13、 if desired, todetermine the average energy of the X rays incident on thespecimen. This information is needed for the determination ofmodified Auger parameters.1.8 This standard does not purport to address all of thesafety concerns, if any, associated with its use. It is theresponsibility of the use
14、r of this standard to establish appro-priate safety and health practices and determine the applica-bility of regulatory limitations prior to use.2. Referenced Documents2.1 ASTM Standards:2E 456 Terminology Relating to Quality and StatisticsE 673 Terminology Relating to Surface AnalysisE 902 Practice
15、 for Checking the Operating Characteristicsof X-Ray Photoelectron SpectrometersE 1016 Guide for Literature Describing Properties of Elec-trostatic Electron SpectrometersE 1078 Guide for Specimen Preparation and Mounting inSurface Analysis1This practice is under the jurisdiction of ASTM Committee E42
16、 on SurfaceAnalysis and is the direct responsibility of Subcommittee E42.03 onAuger ElectronSpectroscopy and X-Ray Photoelectron Spectroscopy.Current edition approved Nov. 1, 2005. Published December 2005. Originallyapproved in 2000. Last previous edition approved in 2000 as E 2108 00.2For reference
17、d ASTM standards, visit the ASTM website, www.astm.org, orcontact ASTM Customer Service at serviceastm.org. For Annual Book of ASTMStandards volume information, refer to the standards Document Summary page onthe ASTM website.1Copyright ASTM International, 100 Barr Harbor Drive, PO Box C700, West Con
18、shohocken, PA 19428-2959, United States.E 1523 Guide to Charge Control and Charge ReferencingTechniques in X-Ray Photoelectron Spectroscopy2.2 ISO Standards:3ISO 9001:2000, Quality Management SystemsRequirementsISO 15472:2001, Surface Chemical AnalysisX-Ray pho-toelectron SpectrometersCalibration of
19、 Energy ScalesISO 18115:2001, Surface Chemical AnalysisVocabulary3. Terminology3.1 DefinitionsFor definitions of terms used in X-rayphotoelectron spectroscopy and surface analysis, see Terminol-ogy E 673 and ISO 18115. For definitions of terms used instatistics, see Terminology E 456.3.2 Symbols and
20、 AbbreviationsTable 1 shows definitionsof the symbols and abbreviations used in this practice.4. Summary of Practice4.1 A procedure is given for calibrating the BE scale of anX-ray photoelectron spectrometer equipped with one or moreof the following sources of characteristic Ka X rays: magne-sium (M
21、g) source; unmonochromated aluminum (Al) source;or monochromated Al source. This procedure is based on ISO15472. In a first calibration for particular operating conditionsof the instrument, or after the instrument has been modified,measurements are made of the binding energies of specifiedcore level
22、s of copper and gold, and these values are then3Available from American National Standards Institute, 25 W. 43rd St., 4thFloor, New York, NY 10036.TABLE 1 Definitions of Symbols and Abbreviationsa measured energy scaling errorBE binding energy, in eVb measured zero offset error, in eVcinumber of cou
23、nts in the i-th channeleV electron voltsEcorrcorrected result for the binding energy corresponding to a given Emeas,ineVEelembinding energy of a frequently measured element at which the indicated binding energy scale is set, aftercalibration, to read correctly, in eVEKkinetic energy of a peak, with
24、reference to the Fermi level, in eVEmeasa measured binding energy, in eVEmeas naverage of the measured binding energies for the peak, n,inTable 3,ineVEmeas nione of a set of measurements of binding energy for the peak, n,inTable 3,ineVEref nreference values for the position of peak, n,inTable 3, on
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
5000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ASTME21082005STANDARDPRACTICEFORCALIBRATIONOFTHEELECTRONBINDINGENERGYSCALEOFANXRAYPHOTOELECTRONSPECTROMETERX

链接地址:http://www.mydoc123.com/p-530295.html