ASTM E1944-1998(2002) Standard Practice for Describing and Measuring Performance of Laboratory Fourier Transform Near-Infrared (FT-NIR) Spectrometers Level Zero and Level One Tests.pdf
《ASTM E1944-1998(2002) Standard Practice for Describing and Measuring Performance of Laboratory Fourier Transform Near-Infrared (FT-NIR) Spectrometers Level Zero and Level One Tests.pdf》由会员分享,可在线阅读,更多相关《ASTM E1944-1998(2002) Standard Practice for Describing and Measuring Performance of Laboratory Fourier Transform Near-Infrared (FT-NIR) Spectrometers Level Zero and Level One Tests.pdf(5页珍藏版)》请在麦多课文档分享上搜索。
1、Designation: E 1944 98 (Reapproved 2002)Standard Practice forDescribing and Measuring Performance of LaboratoryFourier Transform Near-Infrared (FT-NIR) Spectrometers:Level Zero and Level One Tests1This standard is issued under the fixed designation E 1944; the number immediately following the design
2、ation indicates the year oforiginal adoption or, in the case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. Asuperscript epsilon (e) indicates an editorial change since the last revision or reapproval.1. Scope1.1 This practice covers two levels
3、 of tests to measure theperformance of laboratory Fourier transform near infrared(FT-NIR) spectrometers. This practice applies to the short-wave near infrared region, approximately 800 nm (12 500cm-1) to 1100 nm (9090.91 cm-1); and the long-wavelengthnear infrared region, approximately 1100 nm (9090
4、.91 cm-1)to2500 nm (4000 cm-1). This practice is intended mainly fortransmittance measurements of gases and liquids, although it isbroadly applicable for reflectance measurements.1.2 The values stated in SI units are to be regarded as thestandard.1.3 This standard does not purport to address all of
5、thesafety concerns, if any, associated with its use. It is theresponsibility of the user of this standard to establish appro-priate safety and health practices and determine the applica-bility of regulatory limitations prior to use.2. Referenced Documents2.1 ASTM Standards:E 131 Terminology Relating
6、 to Molecular Spectroscopy2E 168 Practices for General Techniques of Infrared Quanti-tative Analysis2E 932 Practices for Describing and Measuring Performanceof Dispersive Infrared Spectrometers2E 1252 Practice for General Techniques for QualitativeAnalysis2E 1421 Practice for Describing and Measurin
7、g Performanceof Fourier Transform Infrared (FT-IR) Spectrometers:Level Zero and Level One Tests23. Terminology3.1 For definitions of terms used in this practice, refer toTerminology E 131. All identifications of spectral regions andabsorbance band positions are given in nanometers (nm), andwavenumbe
8、rs (cm-1); and spectral energy, transmittance, re-flectance, and absorbance are signified by the letters E, T, Rand A respectively. A subscripted number signifies a spectralposition in nanometers, with wavenumbers in parenthesis (forexample,A1940(5154.64), denotes the absorbance at 1940 nm or 5154.6
9、4cm-1).4. Significance and Use4.1 This practice permits an analyst to compare the generalperformance of a laboratory instrument on any given day withthe prior performance of that instrument. This practice is notintended for comparison of different instruments with eachother, nor is it directly appli
10、cable to dedicated process FT-NIRanalyzers. This practice requires the use of a check samplecompatible with the instrument under test as described in 5.3.5. Test Conditions5.1 Operating ConditionsIn obtaining spectrophotometricdata for the check sample, the analyst must select the properinstrumental
11、 operating conditions in order to realize satisfac-tory instrument performance. Operating conditions for indi-vidual instruments are best obtained from the manufacturersinstructional literature due to the variations with instrumentdesign. It should be noted that many FT-NIR instruments aredesigned t
12、o work best if left in standby mode when they are notin use. A record should be kept to document the operatingconditions selected during a test so that they can be duplicatedfor future tests. Note that spectrometers are to be tested onlywithin their respective recommended measurement wavelength(wave
13、number) ranges.5.2 Instrumental characteristics can influence these mea-surements in several ways. Vignetting of the beam (that is, theaperture of the sample cell is smaller than the diameter of thenear infrared beam at the focus) reduces the transmittancevalue measured in nonabsorbing regions, and
14、on most instru-ments can change the apparent wavelength (or wavenumber)scale by a small amount, usually less than 0.01 nm (0.1 cm-1).Focus changes can also change transmittance values, so thesample should be positioned in the same location in the samplecompartment for each measurement. The angle of
15、acceptance1This practice is under the jurisdiction of ASTM Committee E-13 on MolecularSpectroscopy and is the direct responsibility of Subcommittee E 13.03 on InfraredSpectroscopy.Current edition approved March 10, 1998. Published August 1998.2Annual Book of ASTM Standards, Vol 03.06.1Copyright ASTM
16、 International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959, United States.(established by the f number) of the optics between the sampleand detector significantly affects apparent transmittance. Heat-ing of the sample by the beam or by the higher temperatureswhich exist ins
17、ide most spectrometers changes absorbancessomewhat, and even changes band ratios and locations slightly.Allow the sample to come to thermal equilibrium prior tomeasurement.5.3 The recommended check sample should meet the fol-lowing requirements: the check sample should be fully com-patible with the
18、requirements for repeatable sample presenta-tion to the measuring spectrophotometer. The check sampleshould consist of a single pure compound or precisely knownmixture of compounds which is spectroscopically stable overmonths or years. The spectra obtained from such a checksample should be known to
19、indicate changes in the spectro-photometer, not the check sample itself. It is recommended thatindependent verification of the integrity of the check sample beused prior to test measurement. The check sample should bemeasured under precisely the sample measurement conditionsof temperature, humidity,
20、 and instrument set up configuration.Suggested check samples may include, but are not limited tothe following: for gases, water vapor at 5.89 Torr and 1atmosphere ina2mgascell, or methane at 18 psig pressure ina 10 cm gas cell; for liquids, pure spectroscopic gradehydrocarbon compounds (for example,
21、 toluene, decane, isooc-tane, etc.), or precise mixtures of these pure compounds; forreflectance measurements of solids, rare earth oxides mixedwith white halon powder, or Spectralon3-based rare earth oxidereflectance standards. Reference reflectance standards yieldinga featureless, near 100 % refle
22、ctance spectrum are purepowdered sulfur, halon, or Spectralon.6. Level Zero Tests6.1 Nature of TestsRoutine checks of instrument perfor-mance can be performed within a few minutes. They aredesigned to uncover malfunctions or other changes in instru-ment operation but not to specifically diagnose or
23、quantita-tively assess any malfunction. For Level Zero tests, a resolu-tion of 4 cm-1and a nominal measurement time of 30 s isrecommended. Resolution and measurement times can bespecified to match conditions used for routine measurementapplications. The exact measurement time, along with the date,ti
24、me, sample identification, number of scans, and operatorsname, should always be recorded.6.2 PhilosophyThe philosophy of the tests is to usepreviously stored test results as bases for comparison and thevisual display screen or plotter to overlay the current testresults with the reference results (kn
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
5000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ASTME194419982002STANDARDPRACTICEFORDESCRIBINGANDMEASURINGPERFORMANCEOFLABORATORYFOURIERTRANSFORMNEARINFRAREDFTNIRSPECTROMETERSLEVELZEROANDLEVELONETESTSPDF

链接地址:http://www.mydoc123.com/p-529878.html