《ASTM E1942-1998(2010) Standard Guide for Evaluating Data Acquisition Systems Used in Cyclic Fatigue and Fracture Mechanics Testing《周期性疲劳和断裂力学试验中采用的数据采集系统的评定标准指南》.pdf》由会员分享,可在线阅读,更多相关《ASTM E1942-1998(2010) Standard Guide for Evaluating Data Acquisition Systems Used in Cyclic Fatigue and Fracture Mechanics Testing《周期性疲劳和断裂力学试验中采用的数据采集系统的评定标准指南》.pdf(12页珍藏版)》请在麦多课文档分享上搜索。
1、Designation: E1942 98 (Reapproved 2010)Standard Guide forEvaluating Data Acquisition Systems Used in Cyclic Fatigueand Fracture Mechanics Testing1This standard is issued under the fixed designation E1942; the number immediately following the designation indicates the year oforiginal adoption or, in
2、the case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. Asuperscript epsilon () indicates an editorial change since the last revision or reapproval.1. Scope1.1 This guide covers how to understand and minimize theerrors associated with data acqu
3、isition in fatigue and fracturemechanics testing equipment. This guide is not intended to beused instead of certified traceable calibration or verification ofdata acquisition systems when such certification is required. Itdoes not cover static load verification, for which the user isreferred to the
4、current revision of Practices E4, or staticextensometer verification, for which the user is referred to thecurrent revision of Practice E83. The user is also referred toPractice E467.1.2 The output of the fatigue and fracture mechanics dataacquisition systems described in this guide is essentially a
5、stream of digital data. Such digital data may be considered tobe divided into two types Basic Data, which are a sequence ofdigital samples of an equivalent analog waveform representingthe output of transducers connected to the specimen under test,and Derived Data, which are digital values obtained f
6、rom theBasic Data by application of appropriate computational algo-rithms. The purpose of this guide is to provide methods thatgive confidence that such Basic and Derived Data describe theproperties of the material adequately. It does this by settingminimum or maximum targets for key system paramete
7、rs,suggesting how to measure these parameters if their actualvalues are not known.2. Referenced Documents2.1 ASTM Standards:2E4 Practices for Force Verification of Testing MachinesE83 Practice for Verification and Classification of Exten-someter SystemsE467 Practice for Verification of Constant Ampl
8、itude Dy-namic Forces in an Axial Fatigue Testing SystemE1823 Terminology Relating to Fatigue and Fracture Test-ing3. Terminology3.1 Definitions:3.1.1 bandwidth T1the frequency at which the ampli-tude response of the channel has fallen to 1/=2 of its valueat low frequency.3.1.1.1 DiscussionThis defi
9、nition assumes the sensorchannel response is low-pass, as in most materials testing. Anillustration of bandwidth is shown in Fig. 1.3.1.2 Basic Data samplethe sampled value of a sensorwaveform taken at fixed time intervals. Each sample representsthe actual sensor value at that instant of time.3.1.2.
10、1 DiscussionFig. 2 shows examples of Basic Datasamples.3.1.3 data rate T1the date rate is1td Hertz where thetime intervals between samples is tdin seconds.3.1.3.1 DiscussionThe data rate is the number of datasamples per second made available to the user, assuming therate is constant.3.1.4 Derived Da
11、taany waveform parameter which isderived from one or several of the Basic Data samples.3.1.4.1 DiscussionFig. 2 illustrates examples of DerivedData.3.1.5 noise levelthe standard deviation of the datasamples of noise in the transducer channel, expressed in theunits appropriate to that channel.3.1.6 p
12、eakthe point of maximum load in constant ampli-tude loading (see Terminology E1823).3.1.7 phase difference the angle in degrees separatingcorresponding parts of two waveforms (such as peaks), whereone complete cycle represents 360.3.1.7.1 DiscussionThe phase difference of a cyclic wave-form only has
13、 meaning in reference to a second cyclicwaveform of the same frequency.3.1.8 sampling rate T1the rate at which the analog-to-digital converter samples a waveform. This rate may not bevisible to the user of the data acquisition system.1This guide is under the jurisdiction of ASTM Committee E08 on Fat
14、igue andFracture and is the direct responsibility of SubcommitteeE08.03 on AdvancedApparatus and Techniques.Current edition approved Nov. 1, 2010. Published January 2011. Originallyapproved in 1998. Last previous edition approved in 2004 as E1942 - 98(2004).DOI: 10.1520/E1942-98R10.2For referenced A
15、STM standards, visit the ASTM website, www.astm.org, orcontact ASTM Customer Service at serviceastm.org. For Annual Book of ASTMStandards volume information, refer to the standards Document Summary page onthe ASTM website.1Copyright ASTM International, 100 Barr Harbor Drive, PO Box C700, West Consho
16、hocken, PA 19428-2959, United States.3.1.8.1 DiscussionA distinction is made here betweensampling rate and data rate, because in some data acquisitionsystems, the analog waveform may be sampled at a muchhigher rate than the rate at which data are made available to theuser. (Such a technique is commo
17、nly known as over-sampling).3.1.9 word sizethe number of significant bits in a singledata sample.3.1.9.1 DiscussionThe word size is one parameter whichdetermines the system resolution. Usually it will be determinedby the analog-digital converter used, and typically may be 12or 16 bits. If the word s
18、ize is w, then the smallest step changein the data that can be seen is 1 part in 2w, that is thequantization step is d =2w.3.1.10 valleyThe point of minimum load in constantamplitude loading (see Terminology E1823).4. Description of a Basic Data Acquisition System4.1 In its most basic form, a mechan
19、ical testing systemconsists of a test frame with grips which attach to a testspecimen, a method of applying forces to the specimen, and anumber of transducers which measure the forces and displace-ments applied to the specimen (see Fig. 3). The output fromthese transducers may be in digital or analo
20、g form, but if theyare analog, they are first amplified and filtered and thenconverted to digital form using analog-to-digital converters(ADCs). The resulting stream of digital data may be digitallyfiltered and manipulated to result in a stream of output BasicData which is presented to the user in t
21、he form of a displayedor printed output, or as a data file in a computer. Variousalgorithms may be applied to the Basic Data to deriveparameters representing, for example, the peaks and valleys ofthe forces and displacements applied to the specimen, or thestresses and strains applied to the specimen
22、 and so forth. Suchparameters are the Derived Data.4.1.1 The whole measurement system may be divided intothree sections for the purpose of verification: the mechanicaltest frame and its components, the electrical measurementsystem, and the computer processing of data. This guide isspecifically conce
23、rned only with the electrical measurementsystem commencing at the output of the transducers. Beforethe mechanical system is investigated for dynamic errors by themethods given in Practice E467, this guide can be used toascertain that the electrical measurement system has adequateperformance for the
24、measurements required for Practice E467.If the requirements of Practice E467 for the mechanical systemand the recommendations of this guide are met, then the userhas confidence that the Basic Data produced by the testingsystem are adequate for processing by subsequent computeralgorithms to produce f
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
5000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ASTME194219982010STANDARDGUIDEFOREVALUATINGDATAACQUISITIONSYSTEMSUSEDINCYCLICFATIGUEANDFRACTUREMECHANICSTESTING

链接地址:http://www.mydoc123.com/p-529872.html